Electromagnetic immunity
RadiForce MX317W has been tested at the following compliance levels (C) according to the
testing requirements (T) for Professional healthcare facility environments and Home
healthcare environments specified in IEC60601-1-2.
The customer or the user of RadiForce MX317W should assure that it is used in such an
environment.
Immunity test
Electrostatic
discharge (ESD)
IEC61000-4-2
Electrical fast
transients / bursts
IEC61000-4-4
Surges
IEC61000-4-5
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC61000-4-11
Power frequency
magnetic fields
IEC61000-4-8
38
Test level (T)
±8 kV contact
±8 kV contact
discharge
discharge
±15 kV air discharge
±15 kV air discharge
±2 kV power lines
±2 kV power lines
±1 kV input / output
±1 kV input / output
lines
lines
±1 kV line to line
±1 kV line to line
±2 kV line to ground
±2 kV line to ground
0 % U
(100 % dip in
0 % U
T
U
) 0.5 cycles and 1
U
T
cycle
cycle
70 % U
(30 % dip in
70 % U
T
U
) 25 cycles / 50 Hz
U
T
0 % U
(100 % dip in
0 % U
T
U
) 250 cycles / 50
U
T
Hz
Hz
30 A/m
30 A/m
(50 / 60 Hz)
Compliance level
Electromagnetic environment -
(C)
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30 %.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment.
(100 % dip in
Mains power quality should be
T
) 0.5 cycles and 1
that of a typical commercial or
T
hospital environment. If the user
of RadiForce MX317W requires
(30 % dip in
T
continued operation during power
) 25 cycles / 50 Hz
T
mains interruptions, it is
(100 % dip in
recommended that RadiForce
T
) 250 cycles / 50
MX317W be powered from an
T
uninterruptible power supply or a
battery.
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.
The product should be kept at
least 15 cm away from the
source of power frequency
magnetic fields during use.
Guidance