Using the Device

This SmartClass Fiber device allows users to inspect, test, certify,
and save results quickly and easily by driving the user's behavior
and incrementally stepping them through each application as it
should be used in a proper testing workflow as follows:
Starting from the Home Screen
1. Certify the patch cord/test lead endfaces:
a.
Press
b.
Inspect patch cord/test lead end A using the PCM.
c.
Press the [TEST] button on the PCM.
d.
Press
e.
Move end A over to the OPM port.
f.
Inspect patch cord/test lead end B using the PCM.
g.
Press the [TEST] button on the PCM.
h.
Press
i.
Leave end B in the PCM.
2. Certify the bulkhead connector endface:
a.
Press
b.
Inspect the bulkhead endface using the Probe Microscope.
c.
Press the [TEST] button on the Probe.
d.
Press
e.
Plug patch cord/test lead end B into the bulkhead port.
3. Measure the optical power:
a.
Press
b.
Select desired wavelength (OPM value will be displayed on
the screen).
c.
Press
d.
Repeat as necessary for other wavelengths.
focenter.com
• 508-992-6464 | (800) 473-4237 •
23 Centre Street • New Bedford, MA 02740 USA
to activate the PCM.
to save result (if necessary).
to save result (if necessary).
to activate the Probe Microscope.
to save result.
to switch to the OPM.
to save result.