HP 8559A Operation And Service Manual page 55

Spectrum analyzer
Hide thumbs Also See for 8559A:
Table of Contents

Advertisement

MODEL 8559A
RESOLUTION BAN DWlDTH ACCURACY (Cont'd)
19. Set RESOLUTION BW to 3 kHz, leaving FREQ SPAN/DIV set to 0. Repeat steps 14 through 17.
20. Set RESOLUTION BW to 1 kHz, leaving FREQ SPAN/DIV set to 0. Repeat steps 14 through 17.
21. Leave signal generator connected to A16J3 if continuing on with next performance test.
PERFORMANCE TESTS
PERFORMANCE TESTS
Min.
Actual
2.5 kHz
Min.
Actual
0.5 kHz
Max.
3.45 kHz
Max.
1.15 kHz