Advertisement

ADJUSTMENTS
5-29.
CAL OUTPUT AND REF LEVEL CAL ADJUSTMENTS
These adjustments should be followed by frequency response adjustments,
since adiustment of A12R57 1A (offset) will shift the freauencv resDonse of
Band 1 (.bl
-
REFERENCE:
A10 and A12 Schematics
DESCRIPTION:
The 35 MHz CAL OUTPUT signal is adjusted for proper amplitude and frequency using a power meter and
frequency counter. Adjustment range of the front-panel REF LEVEL CAL control is set using the CAL OUT-
PUT signal as a reference.
D l SPLAY
FIGURE 5-20. CAL OUTPUT AND REF LEVEL CAL ADJUSTMENTS TEST SETUP
EQUIPMENT:
Power Meter
Power Sensor
Adapter, Type N (m) to BNC (f)
Extender Cable Assembly
ADJUSTMENTS
NOTE
3 GHz).
SPECTRUM
ANALYZER
OUTPUT
.......................................................
POWER
COUNTER
METER
0 0 0 0 0 0
0
ADAPTER
MODEL 8559A
.
0
H P 432A/435A/B
H P 8481A
H P 1250-0780
H P 5060-0303