MODEL 8559A
4-16.
RESOLUTION BANDWIDTH SELECTIVITY (Cont'd)
PROCEDURE:
1. Set equipment controls as follows:
Spectrum Analyzer:
TUNING
RESOLUTIONBW
INPUTATTEN
REFERENCELEVEL
REFLEVELFINE
SWEEP TIME/DIV
SWEEPTRIGGER
ALTIF
SIGIDENT
BLCLIP
VIDEO FILTER
Signal Generator:
COUNTERMODE
AM
FM
FREQUENCYTUNE
RF
OUTPUTLEVEL
2. Connect equipment as shown in Figure 4-1 1.
For early instruments that do not feature A16J3, a 321.4 MHz, - 25 dBm
signal can be injected directly into the input of Third Converter Assembly
A10 at blue cable A1OW1. Set signal generator COUNTER MODE to EXPAND
3. Adjust signal generator frequency until spectrum analyzer trace is at peak. Put signal generator OUTPUT
LEVEL to position trace at top graticule line.
4. Tune signal generator until trace drops to 2 divisions above graticule baseline. Record signal generator
frequency.
PERFORMANCE TESTS
NOTE
PERFORMANCE TESTS
AUTO
OFF
OFF
OFF
12 o'clock
OFF
OFF
ON
MHz
4-21