HP 8559A Operation And Service Manual page 107

Spectrum analyzer
Hide thumbs Also See for 8559A:
Table of Contents

Advertisement

MODEL 8559A
Adjustment
Name
LOG GAIN
1 dB (offset)
-12D (-12 dB)
1 B
2B
3B
4B
5B
6B
6C
1 A
2A
3A
4A
5A
6A
v3+
v 3
-
v 2
-
v2+
SYM
TABLE 5-1. ADJUSTABLE COMPONENTS (3 OF 3)
Reference
Designator
Paragraph
Description
Adjusts dc offset circuitry at output of Log Amplifier
Assembly A14 for 10 dB steps in log mode.
Adjusts LOG 10 dB/DIV translation.
Adjusts variable gain amplifier for 0 dB with REF
LEVEL FINE control set to 0 dB.
Adjusts variable gain amplifier for -12 dB with REF
LEVEL FINE control set to -1 2 dB.
Adjusts slope of Band 1 response.
Adjusts slope of Band 2 response.
Adjusts slope of Band 3 response.
Adjusts slope of Band 4 response.
Adjusts slope of Band 5 response.
Adjusts high-end breakpoint in slope of Band 5
response.
Adjusts slope of Band 6 response.
Adjusts high-end breakpoint in slope of Band 6
response.
Adjusts gain of Band 1.
Adjusts gain of Band 2.
Adjusts gain of Band 3.
Adjusts gain of Band 4.
Adjusts gain of Band 5.
Adjusts gain of Band 6.
Adjusts diode bias for Band 6.
Adjusts diode bias for Band 5.
Adjusts diode bias for Bands 1 and 2.
Adjusts diode bias for Band 3.
Adjusts diode bias for Band 4.
Adjusts symmetry of first crystal bandwidth fdter
stage.
ADJUSTMENTS

Advertisement

Table of Contents

Troubleshooting