Technical Data - Air Techniques F3700T Instruction Manual

Digital imaging systems
Table of Contents

Advertisement

Electrical Requirements:
Supply Voltage:
Supply Current:
Line Cord:
Power Supply:
Physical Properties:
Dimensions:
Weight:
Environmental Conditions:
Unit in Operation
Temperature:
Humidity:
Storage and Transport
Temperature:
Humidity:
Storage and Transport
Temperature:
Humidity:
Note: Resolution of the device is dependent on operating mode and specific imaging
plate type used.
Resolution (LP/mm)
Classic
Intraoral
Compliance Data:
Laser Classification:
Laser Product Report
Accession Number:
Classification:
Class 1, No Applied Parts, Transportable, Continuous Operation,
Equipment not suitable for use in the presence of flammable anaesthetic mixture(s) .
Electromagnetic Interference:
Electromagnetic interference between the equipment and other devices can occur . Do not use the
equipment in close conjunction with sensitive devices, or devices creating high electromagnetic
disturbances .
Air Techniques, Inc.
Universal: 100V to 240V (90V to 264V) 50/60 Hz
1 .2 A Maximum
Country specific 3-meter power cord
24 Volt Power Supply, P/N B7495 provided
ScanX Intraoral, P/N F3600
Length
Width
39 .4 cm
39 .4 cm
20 Kg
10° to 40°C
10 to 80% (Non-condensing)
(Scanner and accessories)
-29° to 54°C
5 to 95% (Non-condensing)
(PSP Plate)
Maximum 33°C
Maximum 80% (Non-condensing)
Vertical
3 to 18
10 to 18
Class I Laser Product Compliance with
FDA HHS 21 CFR 1040 .10 and IEC 60825-1
0212282-00
ScanX Classic, P/N F3700 &
ScanX Ortho, P/N F3750
Height
Length
40 .6 cm
39 .4 cm
Horizontal
3 to 12+
10 to 12+
Classic & Intraoral

TECHNICAL DATA

Width
Height
39 .4 cm
62 .2 cm
23 .1 Kg