Voltage Division Factor - R&S ENY21 Operating Manual

Eny-fts adapter set for immunity testing with
Table of Contents

Advertisement

Available languages

Available languages

R&S ENY-FTS

2.3 Voltage division factor

The test set-up is calibrated with the measurement cables connected as shown in
Figure 7. See FULL-TWO-PORT calibration (TOSM) of the network analyzers.
Fig. 7
The voltage division factor is measured between the Receiver port and EUT port of the
ISN. The measurement is performed by using the Calibration Adapter "Jig CAL
U100B". The AE port of the network is terminated with a 150 C resistor ("A-150" or "Jig
TRA U0-150" for R&S ENY81-CA6 respectively). The measuring result is a voltage
division factor of around 9.5 dB. The test set-up for measuring the voltage division
factor is shown in the Figure 8. For R&S ENY81-CA6 it is shown in Figure 9.
Fig. 8
Fig. 9
for R&S ENY81-CA6
Operating Manual 1309.8949.14 - 02
Calibration set-up for Voltage Division Factor
Test set-up for Voltage Division Factor in accordance with CISPR 16-1-2
Test set-up for Voltage Division Factor in accordance with CISPR 16-1-2