2 Operation

2.1 Functional test in accordance with CISPR 16-1-2

The R&S ENY-FTS in connection with a network analyzer allows the functional testing
of the impedance stabilization networks (ISN) R&S ENY21, R&S ENY41, R&S ENY81
and R&S ENY81-CA6. This functional testing includes the verification of the
asymmetrical impedance and phase, voltage division factor, longitudinal conversion
loss and decoupling attenuation.
Requirements and examples for the measurement of the parameters of an ISN are
defined in Annex E of the basic standard CISPR 16-1-2. The ITU-T recommendations
G.117 and O.9 are offering further background information for measurements on
symmetrical telecommunication lines. For this purpose the R&S ENY-FTS contains a
LCL balance measuring bridge with a balanced output of 100 C (1 mm banana socket)
and various adapters for the reference calibration of the bridge and the connection of
the measuring bridge to the ISN.
A-8-0-P1 to P4
A-801
Jig TRA U0-150
Jig LCL-Probe
Jig CAL U100B
Fig. 1
Operating Manual 1309.8949.14 - 02
A-2-0-RJ45
A-TERMINATION
A-2-0-RJ11
SAR T801
Parts of the R&S ENY-FTS
BMB-1000
2x Cable
A-CALLOAD
A-REFLOAD
2x SAR T802
SAR T602
SAR T2x0
SAR T4x0
Operation
A-8-CAT6-P1 to P4
A-8-CAT6-P1x to P4x
A-50
A-SHORT
A-OPEN
A-150
A-800
Jig IMA U0
SAR M116
SAR T8x0
7

Advertisement

Chapters

Table of Contents
loading

This manual is also suitable for:

Eny41Eny81-ca6

Table of Contents