Operating manual (Translation) Operating manual ..........page 1 - 60 Temperature calibrators Series TP37 / TP3M Types: TP 37 200 E • TP 37 165 E • TP 37 166 E • TP 37 700 E • TP 3M 165 E • TP 3M 255 E. Please keep this operating manual for future reference.
Further information on the series, the models of the individual types that can be supplied, and their fittings can be found in our catalogues at "catalogues.sika.net". Type plate: You find the type plate on the rear of the device. It includes the type designation, the serial number and the key electric specifications (...
The calibrator is under guarantee for 12 months as from the date of delivery for construction errors or material defects. The guarantee is limited to repair or replacing the calibrator. SIKA also provides an extra 5-year guarantee for calibrators which are calibrated and in- spected annually by the SIKA-DAkkS laboratory.
In order to guarantee that the device operates safely, the operator must act competently and be conscious of safety issues. SIKA provides support for the use of its products either personally or via relevant literature. The customer verifies that our product is fit for purpose based on our technical information.
Page 9
The calibrator cannot then be started any more. After it has cooled, send the calibrator for inspection to SIKA. Risk of injury from hazardous gases! When liquids are heated, the evaporation can result in dangerous gases being formed.
Mineral oil: SIKA supplies the calibrators only with silicone oil. Using mineral oil is possible, but must be done on your own responsibility. The danger and the risk must be borne by the customer and our warranty will be rendered null and void.
TP37 / TP3M Construction, Function and Test Task Construction, Function and Test Task 3.1 Construction The calibrator consists of a robust, black-and-red painted steel housing and has a carry- ing handle at the top The front part of the housing holds the electronics of the TP37 / TP3M for controlling the ref- ...
TP37 / TP3M Construction, Function and Test Task 3.3 Test tasks Test tasks are containers for defined test conditions. They are helpful for recurring testing processes, for standardising test sequences and for generating measurement logs. All the required settings and configurations for the calibration of a test specimen are com- piled in a test task.
IMPORTANT! Mains connecting cable! The mains connecting cable may only be replaced by an equivalent cable. Use only original cables from SIKA or approved cables of the same type with the cor- rect design as replacements ( "Electrical characteristics").
TP37 / TP3M Commissioning and operation 4.3.3 Calibrator for surface temperature sensors Type TP 37 200 E TP 37 165 E TP 37 166 E TP 37 700 E TP 3M 165 E TP 3M 255 E Surface ...
Page 21
TP37 / TP3M Commissioning and operation Draining pump and magnetic lifter: The draining pump is used for pumping out the calibration liquid from the tank of the micro bath. The magnetic stirrer is removed with the help of the magnetic lifter. Both the activities are required before another adapter sleeve or calibration insert is inserted in the calibrator.
Page 23
TP37 / TP3M Commissioning and operation 4.3.4.3 Filling the micro bath * IMPORTANT! Pay attention to § 4.3.4.2 "Information on filling amounts"! When filling, leave enough room for thermal expansion, displacement by sensors and level rise due to stirring. ...
Put on the main switch. The fan of the device starts and the SIKA logo appears for ~ 40 s on the screen. The type designation and the current software version are displayed for ~ 4 s.
Page 25
TP37 / TP3M Commissioning and operation Switching off: CAUTION! Before switching off, pay attention to the block temperature! The calibrator should first reach a safe block temperature before switching off. Upon switching off outside the safe temperature range, the calibrator can get damaged. ...
TP37 / TP3M Operation 5.1 Main window The individual icons and the most important ranges of the main window are presented in brief in the following. Calibrator Set Up: By using the icon, you can reach the "Calibrator Set Up" window. There, you can change the settings of the calibrator (...
TP37 / TP3M Operation 5.3.1 Operating elements The following operating elements can be found in the parameter range of the window. Information box: Different kinds of information is displayed in the information box. This information can be explana- tions for the window, information on the selected element of the selection list, measurement values or measurement value curves.
TP37 / TP3M Operation 5.4 Functionality of the calibrator Here, you can select and manage the function of the calibrator for the relevant measuring in- sert. You can create new functions or tailor unprotected functions to your requirements. IMPORTANT! When changing functions, please consider! The individual functions of the calibrator are independent.
TP37 / TP3M Operation Setting time Input the new value in the input window and confirm the input. The new value is adopted and you will be returned to the previous window. Stirring speed You can change the stirring speed here. The parameter field is only displayed with "Select Function"...
TP37 / TP3M Operation 5.6 Test Task Here, you can select and manage existing test tasks. You can create new test tasks or tailor unprotected test tasks to your requirements. From the parameter field "Test Task" of the main window, you get to the window "Select Test Task".
Page 37
TP37 / TP3M Operation Behaviour at test end / Safety temperature range Using the parameter fields of the new window, you can set the safety temperature range and define the behaviour at test end. The two parameter fields "Safety range from" and "Safety range to" are only visible if "Safety temperature range"...
TP37 / TP3M Operation Defining the number of steps: Note in this context that the number of test points is greater than the number of steps by 1. Tap on the parameter field "Number of Steps" to change the number of steps. Input the desired number and confirm the input.
TP37 / TP3M Operation 5.7 Calibrator settings The most important windows for the setting of the calibrator are described in the following sections. 5.7.1 Calibrator Set Up In the "Calibrator Set Up" window, the settings of the calibrator are displayed and the param- eters are changed.
Page 43
System backup: When taking a system backup, all the settings of the calibrator, including the soft- ware, are saved to the connected medium. The required "SIKA" folder is created if it does not exist. Insert the USB stick in the calibrator and wait for 20-30 s so that the USB stick is recog- nised.
Page 44
Export user files: You can use this parameter field to save all the user settings of test tasks, test spec- imens and functions to the connected medium. The required "SIKA" folder is created if it does not exist. Insert the USB stick in the calibrator and wait for 20-30 s so that the USB stick is recog- nised.
TP37 / TP3M Operation 5.7.2 Measurement reports Here, you can select saved data of testing processes and have them displayed. The meas- urement log is saved under the name of the testing process. Several measurement logs can have the same name, if a testing process was carried out at different times. From the main window, you get to the window "Select measurement report".
Page 47
TP37 / TP3M Testing process / Calibration Display range: The following information is shown in the display range: Current temperature values of the reference used. Current test point. Probable remaining test duration for the current test point. ...
Contact your supplier or directly to SIKA. Please send the device for repair with a brief de- scription of the problem, the environmental conditions and the length of time the device was operational before the problem occurred.
TP37 / TP3M Maintenance and cleaning The following measures must be taken before you send the unit to SIKA for repair: Clean the device thoroughly. This is of extreme importance if the medium is hazardous to health, i.e. caustic, toxic, carcinogenic or radioactive etc.
Fit the fuse compartment back in place and connect the mains connecting cable ( § 4.2 "Electrical connection"). Should the fuses blow repeatedly, there is probably a fault in the calibrator. In this case, send the calibrator to SIKA for repairs ( § 7+§ 7.1). 8.2 Cleaning External cleaning: Clean the TP37 / TP3M with a dry or slightly damp lint-free cloth.
Page 51
IMPORTANT! Temperature fuse at over-temperature! An air flow that is too low can result in the temperature fuse getting triggered. The cali- brator is then rendered operationally disabled, and must be sent to SIKA. Ensure that the grille openings are always clear.
The calibrator should, depending on the application situation, be inspected at appropriate in- tervals on the basis of DIN ISO 10012. We recommend you to return the calibrator to SIKA at intervals of max. 12 months or approx. 500 operating hours for recalibration and readjust- ment.
The calibrator of the series TP37 / TP3M consist of various different materials. It must not be disposed of with household waste (EU Directive 2002/96/EC). Take the TP37 / TP3M to your local recycling plant send the TP37 / TP3M back to your supplier or to SIKA. Technical changes reserved - 53 -...
Page 55
TP37 / TP3M Technical data Type TP 37 TP 37 TP 37 TP 37 TP 3M TP 3M Characteristics 200 E 165 E 166 E 700 E 165 E 255 E Electrical characteristics Supply voltage 100…240 V • 50/60 Hz Power consumption ~ 600 VA ~ 400 VA...
TP37 / TP3M Guideline DAkkS-DKD-R 5-4 • Appendix B 11 Guideline DAkkS-DKD-R 5-4 • Appendix B Details of the DKD technical committee "Temperature and Humidity" for the operation of tempera- ture block calibrators. Having a calibrator certificate from a DAkkS calibration laboratory confirms that the temperature block calibrator meets the high demands of the calibrating ability of such a device, which are outlined in the guideline DAkkS-DKD-R 5-4.