ENGLISH 1. DESCRIPTION The A-LEGIO SENSE panel has been designed to control the proliferation of legionella and other hydrodiffused pathogens through an ON / OFF or proportional dosage of sodium hypochlorite or chlorine dioxide at the same time as a dosage of an anticorrosive conditioner.
A-POOL TOP EVO PH-CL ENGLISH OVERALL DIMENSIONS NB: although devices may be arranged differently on the panel depending on the model selected, the panel’s overall dimensions do not change. NB: although devices may be arranged differently on the panel depending on the model selected, panel’s...
A-POOL TOP EVO PH-CL ENGLISH 2. PROGRAMMING THE PANEL After installing and connecting all the required probes and devices for the purchased panel, follow the programming instructions set out in the pump and instrument manuals. A list summarising the operations to follow for each panel is provided below. HC151+ PUMP MENUS AND PROGRAMMING Please refer to the HC151+ pump manual.
Page 7
A-POOL TOP EVO PH-CL ENGLISH 3. Immerse the probe in the cap, being careful to prevent any excess electrolyte from spilling over (if so, rinse). 4. Unscrew the top part of the probe and connect the bipolar cable to the probe and the instrument’s CL terminal box: 5.
A-POOL TOP EVO PH-CL ENGLISH CL(J) PROBE with mV signal To fit the CL(J-mV) probe to the modular probe holder, follow the instructions provided for the CL(J) 4-20mA probe; to prepare the probe for use, proceed as follows: Remove the protective cap and unscrew the membrane cap from the cell body. Open the electrolyte bottle, screw the nozzle in place and allow any excess air to come out.
A-POOL TOP EVO PH-CL ENGLISH CLEANING THE PROBE HOLDER FILTER 1) Close the valve. 2) Remove the fork. 3) Extract the filter. 4) Pull the filter out and clean it with water. 5) Unscrew the breather plug and the cap. 6) Allow water to run through the probe holder until free of debris and dirt.