Magstim Alpha Coil Range
Operating Manual
Pregnancy: few medical papers on the effects of magnetic stimulation on the unborn child/foetus
or the expectant mother are currently available. As a result, The Magstim Company Limited does
not recommend the use of magnetic stimulation on or near anyone who is pregnant, and by
anyone who is pregnant.
Environmental Conditions: The system must not be used or stored under environmental
conditions that fall outside those specified in section 6.3 of this operating manual.
Liquids: The Alpha coils must not be immersed in water, placed in an ice bucket, or refrigerated;
not even if first placed in a plastic bag. The coil does not have any specialised protection against
the ingress of liquids; conditions where ingress of liquid, or the forming of condensation within the
coil can occur must be avoided, as the electrical insulation will be compromised. Cooling must
only be performed using a flow of cool air from a fan or air conditioning unit.