FIOR & GENTZ NEURO TRONIC Instructions For Use Manual page 6

Hide thumbs Also See for NEURO TRONIC:
Table of Contents

Advertisement

Risk of Falling Due to Use of Unauthorised Accessories
Use only the accessories specified or supplied by the manufacturer in order to avoid increased electromag-
netic emissions and reduced electromagnetic immunity of the knee joint system.
Risk of Falling Due to Electromagnetic Interference
Do not use the knee joint system in close proximity to or stacked with other portable RF communication
devices in order to avoid impairing the function of the knee joint system. If such use is necessary, observe
the knee joint system and other portable RF communication devices in use to ensure that they function
normally.
Risk of Falling Due to Electromagnetic Interference
Use portable RF communication devices (including peripherals such as antenna cables and external anten-
nas) at a safety distance of at least 30cm from all components of the knee joint system to avoid impairing
the function of the knee joint system. If use at a distance of less than 30cm is necessary, observe the knee
joint system during use to ensure that it functions normally. Also note the safety distances for RF commu-
nication devices specified in these instructions for use (see paragraph 26.5).
Risk of Electric Shock Due to Improper Handling
Only use the supplied accessories to avoid electric shock and damage to the knee joint system.
Risk of Injury Due to Improper Handling of the Controller or Remote Control
Use the controller and the remote control as described in these instructions for use. The controller is a
sensitive electronic device with an integrated lithium-polymer battery. Pay particular attention to:
- not wearing the orthosis during the battery charging process,
- avoiding contact with strong heat or fire,
- not charging the controller under direct sunlight, and
- not opening the controller or the remote control.
Risk of Injury Due to Improper Handling of the System Joint
When using the system joint, an opening is formed between the joint's upper and lower part, in which
clothing or skin could get caught. Please inform the patient of this risk.