Figure 3-2 Sitrans F Us Clamp-On Communications Module A5E31949269 - Siemens SITRANS F US Hardware Installation Instructions

Ultrasonic flowmeters clamp-on communications module
Hide thumbs Also See for SITRANS F US:
Table of Contents

Advertisement

F US Clamp-on Communications Module Installation
Figure 3-2
7. Gently press the Communications Module down to connect it to the I/O Module.
8. Tighten the captive screws to secure the Communications Module and I/O Module.
9. If using Ethernet IP, connect the Ethernet connector.
Note
Ethernet Web Utility Discovery.zip
This Web utility is used to locate communications modules on the same network.
The web utility can be found at
http://www.fieldserver.com/techsupport/utility/downloads.php. (Discovery.zip)
The default IP address is: 192.168.0.2
10.If using RS485, connect RS-485 connector to J1.
11.Referring back to figure 3-1, after Communications Module installation is complete, place
the new Communications Module cover supplied in the mounting kit over the board and
tighten the captive screw
12.Close the flow meter cover and secure the side latch.
12
Ethernet Connector J5
SW1
SW2
J3
SITRANS F US Clamp-on Communications Module A5E31949269
RS-485 Connector J1
Captive Screw and Standoff
Standoff
.
Compact Hardware Installation Instructions, 2/2014, A5E32586525-AD
SITRANS F US Clamp-on Communications Module

Advertisement

Table of Contents
loading

Table of Contents