Description of the Upgrade
Installation Procedure for the Upgrade
Option Enable Procedure
1. Locate the email(s) from Keysight which contain license file attachments.
2. Copy the license file(s) from the email(s) to the root directory of the USB
A single license file may contain more than one feature.
3. Insert the USB flash drive to the PNA-X's USB drive slot,. Within 5
Attempting to re-install a license file that is already installed may generate
a "Corrupt Media" error message. Ignore this message.
4. Disconnect the USB flash drive from the PNA-X.
5. On the analyzer, click or press to open the KLM software from your
Step 39. Verify the PNA-X Analyzer Program is Running with the
Correct Options
Once the analyzer has restarted and the Network Analyzer program is again
running:
1. Start the Network Analyzer program.
2. Once the Network Analyzer program is running:
If the options have not been enabled or if the option 219 license has not
been removed, contact Keysight Technologies. Refer to
Assistance from Keysight" on page
50
These emails are a result of
Information" on page
flash drive. More than one license file may be copied to the USB flash
drive.
seconds, the PNA-X should display a small "New licenses installed"
message.
Else, load the license key file(s), manually move your license file(s) to
C:\Program Files\Agilent\licensing. It may take Keysight License
Manager an extra ~5 seconds to enable the licenses.
PNA-X's Windows taskbar by pressing Start > More Programs > Keysight
License Manager folder > Keysight License Manager and verify the
options are correct.
— Press Help > About NA and verify that Option 219 is listed in the
PNA-X application.