Acceleration of Gravity
This scale is equipped with an acceleration of gravity function which means that it can be
calibrated in one location and then adjusted to match the acceleration of gravity at the location
where it will used.
With the display showing A9CC= press the PRINT key to show the current setting. If the value
displayed is acceptable, press the PRINT key to save it. Otherwise use the UNITS key to
select the new setting and then press the PRINT key to save it.
0
Use Default Acceleration of Gravity
1
Enter Acceleration of Gravity Values
NOTE: If you select 1 (Enter Acceleration of Gravity Values) the following additional prompts
will be displayed:
Acceleration of Gravity (Calibration Location)
The display will change to show CAL9C=. Press the PRINT key to show the current
setting. This is the acceleration of gravity value of the location where the scale was
calibrated. If the value displayed is acceptable, press the PRINT key to save it. Otherwise
press the UNITS key to enter the proper digit at the blinking location. Press the ZERO
key to step to the left and the next digit location. Repeat the process until all digits have
been entered. After all digits have been correctly entered, press the PRINT key to store
the value and advance to the next step. Consult the factory for the Acceleration of Gravity
value for your location.
Acceleration of Gravity (Operation Location)
The display will change to show OP9C=. Press the PRINT key to show the current
setting. This is the acceleration of gravity value for the location where the scale will be
operated. If the value displayed is acceptable, press the PRINT key to save it. Otherwise
press the UNITS key to enter the proper digit at the blinking location. Press the ZERO
key to step to the left and the next digit location. Repeat the process until all digits have
been entered. After all digits have been correctly entered, press the PRINT key to store
the value and advance to the next step. Consult the factory for the Acceleration of Gravity
value for your location.
Zero Tracking Range
With the display showing trA= press the PRINT key to show the value assigned to the
automatic Zero Tracking Range. This is the value in scale divisions that will be automatically
zeroed off. Use the UNITS key to step through the values. Once the proper value is shown
press the PRINT key to store the value. Allowable values are:
0
Disable Zero Tracking
1
.5d (0.5 division) Zero Tracking
2
1d (1 division) Zero Tracking
12