ANNUNCIATORS

The annunciators are turned on to indicate that the display is in the mode corresponding to the
annunciator label or that the status indicated by the label is active.
0 (Center-of-Zero)
The Center-of-Zero annunciator is located to the lower right of the ounce weight display and is
turned on to indicate that the weight is within +/- 1/4 division of the center of zero.
 (Stable)
The (Stable) annunciator is located to the far lower right of the ounce weight display and is
turned on when the weight display is stable. When off, it means that the change in successive
weight samples is greater than the motion limits selected during setup and calibration.
POUND
The POUND annunciator is located on the left of the weight display and is turned on when the
pounds & ounces mode is selected to show that the displayed weight is pounds.
OUNCE
The OUNCE annunciator is located slightly to the right center of the weight display and is
turned on when the pounds & ounces mode is selected to show that the displayed weight is
ounces.
kg
The kg annunciator is located on the right of the weight display and is used to indicate that the
displayed units of weight measurement is kilograms.
LOW BATTERY
The low battery annunciator is used with the battery operation. It will turn ON to indicate that
the batteries will soon need to be replaced (if using Alkaline) or recharged (if using NiCad or
NiMH). No change in operation will occur until just before the battery voltage drops to a level
where operation is affected. At this level, the indicator will automatically turn itself off.
WEIGHT LOCK
The WEIGHT LOCK annunciator is turned on to show that the scale is locked onto the weight.
bAbY
This will be displayed for three to five seconds after a ten to fifteen second time has elapsed
during which the scale was unable to lock on to a valid weight. This indicates to the user that
the baby is too active to obtain a valid weight. The display message will be repeated until
either a valid weight is obtained or the baby is removed from the scale tray.