Battery Type
With the display showing niCAd= press the PRINT key to show current setting. If the value
displayed is acceptable, press the PRINT key again to save it. Otherwise use the UNITS key to
select the new setting and then press the PRINT key to save it. NOTE: This setting may be
revised without having to enter the calibration mode. Allowable values are:
0
Alkaline batteries - battery charging is DISABLED
1
NiCad or NiMH batteries - battery charging is ENABLED
2
Battery charging is ENABLED and FORCED ON. This selection
forces battery charging for NiCad or NiMH batteries that are
discharged. NOTE: After 15 hours of charging, the indicator will
automatically change the niCAd= setting back to a 1.
IMPORTANT: Selecting 1 or 2, enables battery charging. DO NOT select 1 or 2,
when using alkaline batteries.
Threshold
NOTE: This prompt will only be displayed if you selected USA = 1 (Domestic).
The display will change to show tHrSh=, the threshold lock value. This value is the weight in
ounces that the load on the scale must rise above for the Weight Lock feature to function.
Press the PRINT key to show the current setting. If the value displayed is acceptable, press
the PRINT key again to save it. Otherwise, use the UNITS and ZERO keys to enter a new
value and then press the PRINT key to save it. Allowable values are: 0 through 99. Note, that a
setting of 0 will disable the threshold lock.
Power Up Zero
With the display showing PUO= press the PRINT key to show current setting. If the value
displayed is acceptable, press the PRINT key again to save it. Otherwise use the UNITS key to
select the new setting and then press the PRINT key to save it. NOTE: This setting may be
revised without having to enter the calibration mode.
0
Power Up Zero is DISABLED
1
Power Up Zero is ENABLED. The weight display will be reset to
zero automatically when the scale is turned on.
Automatic Shutoff
The Automatic Shutoff feature will automatically turn the scale off after a predetermined period
of inactivity to prolong battery life. To turn the scale back on you must press the ON key.
With the display showing ASH=, press the PRINT key to show the current setting. If the value
displayed is acceptable, press the PRINT key again to save it. Otherwise, use the UNITS key
to select the number of minutes (time approximate) of inactivity before turning the scale off and
then press the PRINT key to save it. NOTE: This setting may be revised without having to
enter the calibration mode. Allowable values are 0 through 9. Note that 0 disables the
Automatic Shutoff feature.
14
SETUP AND CALIBRATION, CONT.
8535-M074-O1 Rev C • 6745 Baby Scale Operation