SDS2000X HD Series Digital Oscilloscope User Manual Contents INTRODUCTION ....................... 9 IMPORTANT SAFETY INFORMATION ................10 ................... 10 ENERAL AFETY UMMARY ..................13 AFETY ERMS AND YMBOLS ....................14 ORKING NVIRONMENT ....................15 OOLING EQUIREMENTS ................ 16 OWER AND ROUNDING EQUIREMENTS ........................
Page 4
SDS2000X HD Series Digital Oscilloscope User Manual ............... 35 ONNECTING TO XTERNAL EVICES YSTEMS 6.3.1 Power Supply ......................35 6.3.2 LAN ......................... 35 6.3.3 USB Peripherals ..................... 35 6.3.4 Auxiliary Output ...................... 35 6.3.5 Waveform Generator ....................36 6.3.6 Probes ........................36 6.3.7 Logic Probe ......................
Page 5
SDS2000X HD Series Digital Oscilloscope User Manual 12 DIGITAL CHANNELS ..................... 63 ........................63 VERVIEW ............... 64 NABLE ISABLE THE IGITAL HANNELS ....................66 IGITAL HANNEL ETUP 13 HORIZONTAL AND ACQUISITION SETUP ..............70 ......................70 IMEBASE ETUP ......................71 CQUISITION ETUP 13.2.1 Overview .........................
Page 6
SDS2000X HD Series Digital Oscilloscope User Manual ...................... 107 RIGGER OUPLING ....................... 108 OISE EJECT ......................109 RIGGER 16 SERIAL TRIGGER AND DECODE ................114 ........................114 VERVIEW I2C T ................116 RIGGER AND ERIAL ECODE 16.2.1 I2C Signal Settings ....................116 16.2.2 I2C Trigger ......................
Page 7
SDS2000X HD Series Digital Oscilloscope User Manual 16.10.1 MIL-STD-1553B Signal Settings ................142 16.10.2 MIL-STD-1553B Serial Decode ................142 SENT T ................144 RIGGER AND ERIAL ECODE 16.11.1 SENT Signal Settings ................... 144 16.11.2 SENT Trigger ......................144 16.11.3 SENT Serial Decode .................... 148 ..................
Page 8
SDS2000X HD Series Digital Oscilloscope User Manual ........................185 LGEBRA 19.3.1 Differential ......................185 19.3.2 Integral ........................186 19.3.3 Square Root ......................187 19.3.4 Absolute ........................ 187 19.3.5 Sign........................188 19.3.6 exp10 ........................189 19.3.7 lg ........................... 190 19.3.8 Interpolate ......................190 ....................
Page 9
SDS2000X HD Series Digital Oscilloscope User Manual ......................235 NRUSH URRENT ......................236 WITCHING ........................239 ......................... 240 ODULATION ......................240 UTPUT IPPLE ...................... 241 ....................243 RANSIENT ESPONSE PSRR ........................... 244 ....................... 245 OWER FFICIENCY SOA ..........................245 28 BODE PLOT ......................... 248 ........................
Page 10
SDS2000X HD Series Digital Oscilloscope User Manual 32.2.6 Reference Position Setting ................... 279 ....................... 282 UTTON ....................... 282 UICK CTION I/O S ........................283 ETTING 32.5.1 LXI ........................283 32.5.2 LAN ........................283 32.5.3 Network Storage ....................284 32.5.4 Web Server ......................285 32.5.5 Emulation ......................
SDS2000X HD Series Digital Oscilloscope User Manual Important Safety Information This manual contains information and warnings that must be followed by the user for safe operation and to keep the product in a safe condition. General Safety Summary Carefully read the following safety precautions to avoid personal injury and prevent damage to the instrument and any products connected to it.
Page 13
SDS2000X HD Series Digital Oscilloscope User Manual high voltage damage. The internal devices of the equipment are sensitive to static electricity, and direct contact can easily cause irreparable damage to the equipment. It is necessary to return to the factory or the company's designated maintenance organization for maintenance. Be sure to pull out the power supply when repairing the equipment.
Page 14
SDS2000X HD Series Digital Oscilloscope User Manual Do not use the equipment for measurements on mains circuits. Do not use the equipment for measurements on voltages exceeding the voltage ranges described in the manual. The maximum additional transient voltage cannot exceed 1300 V. The responsible body or operator should refer to the instruction manual to preserve the protection afforded by the equipment.
SDS2000X HD Series Digital Oscilloscope User Manual Safety Terms and Symbols When the following symbols or terms appear on the front or rear panel of the instrument or in this manual, they indicate special care in terms of safety. This symbol is used where caution is required. Refer to the accompanying information or documents to protect against personal injury or damage to the instrument.
SDS2000X HD Series Digital Oscilloscope User Manual Working Environment The design of the instrument has been verified to conform to EN 61010-1 safety standard per the following limits: Environment The instrument is used indoors and should be operated in a clean and dry environment with an ambient temperature range.
SDS2000X HD Series Digital Oscilloscope User Manual Note: Installation (overvoltage) category I refers to situations where equipment measurement terminals are connected to the source circuit. In these terminals, precautions are done to limit the transient voltage to a correspondingly low level. Installation (overvoltage) category II refers to the local power distribution level which applies to equipment connected to the AC line (AC power).
SDS2000X HD Series Digital Oscilloscope User Manual Power and Grounding Requirements The instrument operates with a single-phase, 100 to 240 Vrms (+/-10%) AC power at 50/60 Hz (+/- 5%). No manual voltage selection is required because the instrument automatically adapts to line voltage. Depending on the type and number of options and accessories (probes, PC port plug-in, etc.), the instrument can consume up to 193 W of power.
SDS2000X HD Series Digital Oscilloscope User Manual Cleaning Clean only the exterior of the instrument, using a damp, soft cloth. Do not use chemicals or abrasive elements. Under no circumstances allow moisture to penetrate the instrument. To avoid electrical shock, unplug the power cord from the AC outlet before cleaning. Warning: Electrical Shock Hazard! No operator serviceable parts inside.
Page 20
SDS2000X HD Series Digital Oscilloscope User Manual UL 61010-2-030:2018. Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use – Part2-030: Particular requirements for testing and measuring circuits. Canadian certification CAN/CSA-C22.2 No. 61010-1:2012/A1:2018-11. Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use –...
SDS2000X HD Series Digital Oscilloscope User Manual Informations essentielles sur la sécurité Ce manuel contient des informations et des avertissements que les utilisateurs doivent suivre pour assurer la sécurité des opérations et maintenir les produits en sécurité. Exigence de Sécurité Lisez attentivement les précautions de sé...
Page 22
SDS2000X HD Series Digital Oscilloscope User Manual Entretien du matériel. En cas de dé faillance de l'équipement, ne pas dé monter et entretenir l'équipement sans autorisation. L'équipement contient des condensateurs, de l'alimentation é lectrique, des transformateurs et d'autres dispositifs de stockage d'é nergie, ce qui peut causer des blessures à haute tension. Les dispositifs internes de l'é...
Page 23
SDS2000X HD Series Digital Oscilloscope User Manual Le Circuit d 'alimentation électrique ne peut pas être mesuré à l' aide du dispositif, ni la tension qui dé passe la plage de tension dé crite dans le présent manuel. Seuls les ensembles de sondes conformes aux spé cifications du fabricant peuvent être utilisés.
SDS2000X HD Series Digital Oscilloscope User Manual Termes et symboles de sé curité Lorsque les symboles ou termes suivants apparaissent sur le panneau avant ou arrière de l'instrument ou dans ce manuel, ils indiquent un soin particulier en termes de sécurité. Ce symbole est utilisé...
SDS2000X HD Series Digital Oscilloscope User Manual Environnement de travail La conception de l'instrument a été certifiée conforme à la norme EN 61010-1, sur la base des valeurs limites suivantes: Environnement L'instrument doit être utilisé à l'intérieur dans un environnement propre et sec dans la plage de température ambiante.
Page 26
SDS2000X HD Series Digital Oscilloscope User Manual Installation (overvoltage) Category Definitions Définition de catégorie d 'installation (surtension) La caté gorie II d'installation (surtension) est un niveau de signal applicable aux terminaux de mesure d' équipement relié s au circuit source.Dans ces bornes, des mesures préventives sont prises pour limiter la tension transitoire à...
SDS2000X HD Series Digital Oscilloscope User Manual Exigences de refroidissement Cet instrument repose sur un refroidissement à air forcé avec des ventilateurs internes et des ouvertures de ventilation. Des pré cautions doivent être prises pour éviter de restreindre le flux d'air autour des ouvertures (trous de ventilateur) de chaque côté...
SDS2000X HD Series Digital Oscilloscope User Manual du cordon d'alimentation doit être insérée dans une prise secteur correspondante contenant un contact de sé curité avec la terre. Utilisez uniquement le cordon d'alimentation spé cifié pour cet instrument et certifié pour le pays d'utilisation. Avertissement: risque de choc électrique! Toute interruption du conducteur de terre de protection à...
SDS2000X HD Series Digital Oscilloscope User Manual Conditions anormales Utilisez l'instrument uniquement aux fins spécifiées par le fabricant. N'utilisez pas la lunette s'il y a des signes visibles de dommages ou si elle a été soumise à de fortes contraintes de transport. Si vous pensez que la protection de l'oscilloscope a été...
SIGLENT shall not be responsible for any defect, damage, or failure caused by any of the following: Attempted repairs or installations by personnel other than SIGLENT.
SDS2000X HD Series Digital Oscilloscope User Manual Document Conventions For convenience, text surrounded by a box border is used to represent the button of the front panel. For example, Default represents the "Default" button on the front panel. Italicized text with shading is used to represent the touchable or clickable menu/button/region on the touch screen.
SDS2000X HD Series Digital Oscilloscope User Manual Adjust the Supporting Legs Adjust the supporting legs properly to use them as stands to tilt the oscilloscope upwards for stable placement as well as easier operation and observation of the instrument. Power on SDS2000X HD provides two ways for power on, which are: Auto Power-on When the “Auto Power-on”...
SDS2000X HD Series Digital Oscilloscope User Manual Power on by Manual When the " Auto Power-on” option is disabled, the power button on the front panel is the only control for the power state of the oscilloscope. Shut down Press the power button to turn off the oscilloscope. Or follow the steps below: Utility >...
SDS2000X HD Series Digital Oscilloscope User Manual Quick Start Front Panel Overview A. Touch Screen Display: The display and major functions area. See the "Touch Screen Display" chapter for more details B. Front Panel: Includes knobs and buttons. See the "Front Panel" chapter for more details C.
SDS2000X HD Series Digital Oscilloscope User Manual Rear Panel Overview A. Ext Trigger Input B. Auxiliary Out: Outputs the trigger indicator. When Mask Test is enabled, outputs the pass / fail signal C. WaveGen: Built-in waveform generator output D. 100M LAN Port: Connect the port to the network for remote control E.
SDS2000X HD Series Digital Oscilloscope User Manual Connecting to External Devices/Systems 6.3.1 Power Supply The standard power supply for the instrument is 100~240 V, 50/60 Hz. Please use the power cord provided with the instrument to connect it to AC power. 6.3.2 Connect the LAN port to the network with a network cable with an RJ45 connector for remote control.
6.3.6 Probes The SDS2000X HD series oscilloscope supports active probes and passive probes. The specifications and probe documents can be obtained at int.siglent.com, www.siglentna.com, or www.siglenteu.com. Probe Compensation When a passive probe is used for the first time, you should compensate it to match the input channel of the oscilloscope.
Use a non-metallic driver to adjust the low-frequency compensation adjustment hole on the probe until the waveform displayed is as the “Perfectly compensated” in the figure above. It’s not necessary to compensate active probes like the SIGLENT SAP series. 6.3.7 Logic Probe To connect the logic probe: Insert the probe, with the correct side facing up, until you hear a “click”.
SDS2000X HD Series Digital Oscilloscope User Manual Touch Screen Display Overview The entire SDS2000X HD display is a capacitive touch screen. Use your fingers to touch, drag, pinch, spread, or draw a selection box. Many controls that display information also work as “buttons” to access other functions.
SDS2000X HD Series Digital Oscilloscope User Manual Trigger Level Line (Vertical) and Trigger Delay Indicator (Horizontal) show the trigger position of the waveform. Cursors show where measurement points have been set. Move the cursors to quickly reposition the measurement point. Channel Descriptor boxes include analog channels (Cx), digital channels (D), math (Fx), and reference (Ref).
SDS2000X HD Series Digital Oscilloscope User Manual Grid Area The grid area displays the waveform traces. Traces can be moved by dragging and re-scaled by pinch and spread. The area is divided into 8 (vertical) * 10 (horizontal) grids. The best display effect can be obtained by adjusting the waveform intensity and graticule.
SDS2000X HD Series Digital Oscilloscope User Manual Channel Descriptor Box Channel Descriptor boxes include analog channels (Cx), digital channels (D), math (Fx), and reference (Ref). They are located under the grid area, showing the parameters of the corresponding traces. Touching the boxes creates a dialog box. See chapter “Vertical Setup” for more details. Below is an example for analog channel 1: A.
Page 44
SDS2000X HD Series Digital Oscilloscope User Manual :AC coupling, 50 Ω impedance :Ground Vertical Scale: The scale of each grid in the vertical direction. For example, when the vertical scale is 1.00 V/div, the full scale of the oscilloscope is 1.00 V/div * 8 div=8 V. Vertical Offset: The offset of the channel in the vertical direction.
SDS2000X HD Series Digital Oscilloscope User Manual Timebase and Trigger Descriptor Boxes The Timebase Descriptor box shows the parameters of the timebase. See chapter “Horizontal and Acquisition Setup” for more information. Trigger delay Horizontal scale (timebase) # Samples Sample Rate Trigger delay: The time offset of trigger position.
Page 46
SDS2000X HD Series Digital Oscilloscope User Manual EXT/5:5x attenuation of external trigger channel AC Line: AC mains supply D0~D15: Digital channels Trigger coupling: Coupling mode of the current trigger source. It is only valid when the trigger source is C1~C4, EXT, or EXT/5.
SDS2000X HD Series Digital Oscilloscope User Manual Dialog Box The dialog box on the right side of the screen is the main area for setting the parameters of the selected function. A. Title bar. Touching the bar can hide the dialog box, and touching again can open the dialog box.
Page 48
SDS2000X HD Series Digital Oscilloscope User Manual Let’s use the operation of setting the “deskew” of a channel as an example: If the expected value is 65 ns, input “65” on the virtual keypad, and then choose the units n to complete the operation. On the virtual keypad, touching the button Max , Min , and Default quickly sets the parameter to the maximum,...
SDS2000X HD Series Digital Oscilloscope User Manual Touch Gestures Waveforms, cursors, and trigger levels can be adjusted and a rectangular zone can be drawn by touch gestures in the grid area. Drag the waveform left and right to move it on Pinch and spread the waveform horizontally the horizontal axis to re-scale the timebase...
SDS2000X HD Series Digital Oscilloscope User Manual Touch and drag the cursor to move it Touch and drag the cursor information region to move the pair of cursors simultaneously Draw a rectangular box to create a zone or a histogram region. At the beginning of the gesture keep the angle close to 45°...
SDS2000X HD Series Digital Oscilloscope User Manual Front Panel Overview The front panel is designed to operate the basic functions without having to open the software menu. Most of the front panel controls duplicate functionality available through the touch screen display but the operation is more quickly achieved.
SDS2000X HD Series Digital Oscilloscope User Manual Vertical Control A. When a channel is disabled, push its channel button to turn it on. When a channel is turned on but not activated, push the button to activate it. When the channel is turned on and activated, push the button to disable it B.
SDS2000X HD Series Digital Oscilloscope User Manual Horizontal Control A. Rotate to adjust the horizontal scale (time/div); push to enable Zoom; push again to switch between the main window and zoom window. B. Push to enable Zoom. Push again to exit Zoom mode. C.
SDS2000X HD Series Digital Oscilloscope User Manual Run/Stop Button Press the button to switch the acquisition state between Run and Stop. When the state is Run, the button is illuminated in green. When the state is Stop, the button is illuminated in red. Auto Setup Button Auto Setup forces the oscilloscope to automatically set the vertical scale, horizontal scale, and trigger level according to the input signal...
SDS2000X HD Series Digital Oscilloscope User Manual Other Buttons Enables/Disables measurements and recalls the MEASURE dialog box. Performs a quick screenshot save or saves the currently save type file to the storage path. See the section "Save Button" for details. Enables/Disables the touch screen.
SDS2000X HD Series Digital Oscilloscope User Manual Multiple Approaches to Recall Functions The oscilloscope can recall functions through different approaches. Menu Bar If you are familiar with common computer programs, you may first choose to access a function by the drop-down menu from the menu bar at the top of the display. For example, to open the trigger setup dialog box, you can follow the steps below: Trigger >...
SDS2000X HD Series Digital Oscilloscope User Manual Descriptor Box For the setup of channels, math, ref, timebase, and trigger, there are dialog boxes at the bottom of the display. For the introduction of the descriptor box, see sections "Channel Descriptor Box" and "Timebase and Trigger Descriptor Box".
SDS2000X HD Series Digital Oscilloscope User Manual 10 Quickly Capture the Signal This is an example of how to acquire a signal quickly. In this example, we assume the signal is connected to channel 1 and channel 1 is turned off. First, press the channel 1 button to turn on channel 1.
SDS2000X HD Series Digital Oscilloscope User Manual 11 Vertical Setup Turn on/off a Channel From the Front Panel When a channel is disabled, push its channel button to turn it on. When a channel is turned on but not activated, push the button to activate it. When the channel is turned on and activated, push the button to disable it From the Touch Screen Touch the + button and then select the expected channel to turn it on, and its channel descriptor...
SDS2000X HD Series Digital Oscilloscope User Manual Channel Setup Touch the channel descriptor box, a quick dialog will pop up. Vertical scale and offset can also be set from this dialog box. A. Touch the region to set the vertical scale with the universal knob or virtual keypad B.
Page 61
SDS2000X HD Series Digital Oscilloscope User Manual Activating a channel recalls the channel dialog box, displaying more parameters: A. Turn channel on/off B. Coupling (DC, AC, or GND) C. Bandwidth limit (Full, 200 MHz, or 20 MHz) D. Probe settings including attenuation (1X, 10X, 100X, or custom) and probe check.
Page 62
SDS2000X HD Series Digital Oscilloscope User Manual Bandwidth Limit Full bandwidth passes through signals with high-frequency components, but it also means that noise with high-frequency components can pass through. When the frequency component of the signal is very low, better signal-to-noise ratios (SNR) can be obtained by turning on a bandwidth limit. The SDS2000X HD provides two hardware bandwidth limit options: 20 MHz and 200 MHz.
Page 63
SDS2000X HD Series Digital Oscilloscope User Manual Apply to With this setting, some common functions such as Trigger, FFT, Simple Measure, Cursor, Search, DVM, Histogram, Mask Test, Counter, and AWG can be quickly applied to the selected channel. Once a function is specified, it will switch directly to the function menu and automatically set that channel as the source.
Page 64
SDS2000X HD Series Digital Oscilloscope User Manual Before deskew After deskew Note: When the channel with deskew compensation is the trigger source, the trigger delay indicator does not change as the deskew value changes. Invert When invert is enabled, the waveform is rotated 180 degrees around earth potential (0 Volts). Before invert After invert Trace...
The equipment shall be used only for the purposes specified by the manufacturer. The SPL2016 probe is used only for SIGLENT's special series of oscilloscopes. Protection mechanisms can be compromised if the way the devices connected by the SPL2016 are not used for their intended purpose.
SDS2000X HD Series Digital Oscilloscope User Manual damaged. Maintenance work shall be performed by maintenance personnel with appropriate qualifications. Keep product surface clean and dry. SDS2000HD-16LA Option The software option adds the following functions to the oscilloscope: Digital channel acquisition and analysis - Acquire and analyze the signals connected to the digital logic probe, including waveform display, save, parameter measurement, etc.
Page 67
SDS2000X HD Series Digital Oscilloscope User Manual A. Digital channel waveform display, which shares the same grid area with the analog channels. B. Digital channel descriptor box C. Dialog box D. Digital channel indicators. Up to 16 digital channels are organized in two groups with different thresholds: D15~D8 and D7~D0.
SDS2000X HD Series Digital Oscilloscope User Manual Digital Channel Setup Touch the digital descriptor box, then the quick menu of digital channel settings pops up above the descriptor box. In the menu height and position of digital channels can be set: A.
Page 69
SDS2000X HD Series Digital Oscilloscope User Manual A. Turn on/off the digital channels B. Labels can be set to data, address, or custom characters. C. Logic threshold of D7~D0. The oscilloscope will automatically set the threshold according to the specified logic family, or the user can set the threshold manually using the Custom option.
Page 70
SDS2000X HD Series Digital Oscilloscope User Manual The configurable logical level includes TTL, CMOS, LVCMOS 3.3 V, LVCMOS 2.5 V, and Custom. The setting range of the custom threshold is -10.0 V to + 10.0 V. Bus Setting Press the Digital button on the front panel or touch the Digital descriptor box on the bottom to open the Digital dialog box, touch Bus to open the Digital Bus dialog box.
Page 71
SDS2000X HD Series Digital Oscilloscope User Manual Bus Data Setting The bus data setting supports setting the digital channel data by bit. Touch Data to open the bus data setting dialog. In this dialog, the mapping relationship between the bus bits and the digital channels is shown as Bit.x[Dy].
SDS2000X HD Series Digital Oscilloscope User Manual 13 Horizontal and Acquisition Setup Timebase Setup The timebase setup is used to adjust the scale and offset of the X (horizontal) axis. This setting applies to all analog, digital channels, and all math traces except FFT. Touch the timebase descriptor box and the quick menu of the timebase settings will pop up.
SDS2000X HD Series Digital Oscilloscope User Manual Acquisition Setup 13.2.1 Overview Touch Acquire Menu on the quick menu of the timebase settings, or touch the menu bar Acquire > Menu to recall the Acquire dialog box on the right side. A.
Page 74
SDS2000X HD Series Digital Oscilloscope User Manual X Interpolation Sinc Interpolation Acq Mode: "Fast" is the default setting. The SDS2000X HD provides a very high waveform update rate in fast mode. "Slow" mode will slow down the waveform update on purpose. Memory Depth: The memory depth that can be supported.
SDS2000X HD Series Digital Oscilloscope User Manual 13.2.2 Acquisition The acquisition mode is used to determine how to acquire and process the signal. Normal: The oscilloscope samples the signal with equal time intervals. For most waveforms, the best display effect can be obtained using this mode. Peak: Peak detect mode.
Page 76
SDS2000X HD Series Digital Oscilloscope User Manual Average mode(128) Normal mode Note: Average acquisition is only valid for periodic signals, and it is important to ensure that the waveform is triggering stably when using average mode. ERES: In this mode, the oscilloscope digitally filters the sampling points, which can reduce the broadband random noise on the input signal and improve the signal-to-noise ratio, to improve the effective resolution (ENOB) of the oscilloscope.
SDS2000X HD Series Digital Oscilloscope User Manual 13.2.3 Memory Management Memory Management controls how the instrument stores the acquired samples. Auto: The default acquisition setting. After setting the maximum memory in the Auto mode, the oscilloscope automatically adjusts the sample rate and memory depth according to the time base. The adjustment principle is to follow the formula: Sample rate = number of samples/time of acquisition Where the time of acquisition is the time corresponding to the full display (i.e.
SDS2000X HD Series Digital Oscilloscope User Manual If you would like to stop the display in Roll mode, press the Run/Stop button. To clear the display and restart an acquisition in Roll mode, press the Run/Stop button again. Note:Zoom is only supported after Stop in Roll mode. 13.2.5 Sequence Touch Acquire >...
Page 79
SDS2000X HD Series Digital Oscilloscope User Manual Sequence N+1 Sequence N Sequence N-1 Stored Waveform Example: Input a pulse sequence with a period of 50 ms to C1. The rise time of the pulse is 2 ns, while fall time is 100 ns;...
Page 80
SDS2000X HD Series Digital Oscilloscope User Manual the segments to maximum (74403 in this example, up to 80,000 depending on the number of samples at the current time base). Wait patiently until the acquisition completes, then all the waveforms satisfying the trigger conditions are displayed on the screen. In Sequence mode, there is no waveform displayed on the screen until the acquisition is completed.
SDS2000X HD Series Digital Oscilloscope User Manual History Touch Analysis > History to recall the history dialog box. A. Turn on or off history mode B. Specify the frame index C. Set the playing mode D. Play backward, pause, and play forward E.
Page 82
SDS2000X HD Series Digital Oscilloscope User Manual Observing the 5412th frame captured by Sequence in history mode Touch the List area, turn on the list, in which the time label corresponding to the 5412th waveform is displayed. The time resolution is microseconds. Time label types include AcqTime or Delta T , AcqTime corresponds to the absolute time of the frame, synchronized with the real-time clock of the oscilloscope;...
Page 83
SDS2000X HD Series Digital Oscilloscope User Manual In addition to manually specifying a frame, history mode supports autoplay: Press the softkey to replay the waveform from the current frame to the first. Press the softkey to stop replay. Press the softkey to replay the waveform from the current frame to the last.
SDS2000X HD Series Digital Oscilloscope User Manual 14 Zoom The SDS2000X HD supports waveform zoom in the horizontal and vertical directions. Press down the horizontal knob or the button on the front panel to turn on the zoom function. When the Zoom function is on, the waveform area is divided into upper and lower parts. The area of about 1/3 height above is the main window, and the area of about 2/3 height below is the zoom window.
Page 85
SDS2000X HD Series Digital Oscilloscope User Manual Vertical setting Horizontal setting When the zoom window is activated, the zoom area can be expanded or compressed by rotating the horizontal and vertical scale knob. Rotating the horizontal and vertical position knob to move the position of the region.
Page 86
SDS2000X HD Series Digital Oscilloscope User Manual Double click a location in the main window to quickly set the center of the zoom area. Adjust the horizontal position of the waveform by dragging left and right in the gray area of the main window Adjust the vertical position of the waveform by dragging up and down in the gray area of the...
SDS2000X HD Series Digital Oscilloscope User Manual 15 Trigger Overview The oscilloscope only acquires waveforms of interest (i.e. the ones that satisfy the trigger condition) and aligns all trigger events at the trigger position to form a stable waveform display. The trigger is one of the most important features of an oscilloscope since we can only analyze a signal that we can trigger reliably and stably.
SDS2000X HD Series Digital Oscilloscope User Manual Trigger Setup Touch the trigger descriptor box to display the quick menu of trigger settings. The trigger setup dialog box is displayed on the right side of the screen. A. Touch the level region and rotate the Level knob on the front panel to adjust the trigger level;...
SDS2000X HD Series Digital Oscilloscope User Manual Trigger Related Label Trigger level Indicator Horizontal 0 position Horizontal 0 position Indicator (out of screen) Indicator Trigger Level Both analog and digital triggers must have a correct trigger level value. The oscilloscope judges whether a waveform satisfies the trigger condition when it crosses the trigger level.
SDS2000X HD Series Digital Oscilloscope User Manual Trigger Mode The trigger mode determines how the oscilloscope acquires waveforms. Auto:An internal timer triggers the sweep after a preset timeout period if no trigger has been found so that the oscilloscope continuously updates the display whether a trigger happens or not. Auto mode is suitable for unknown signals or DC signals.
SDS2000X HD Series Digital Oscilloscope User Manual Trigger Type 15.5.1 Overview The trigger modes of the SDS2000X HD are digital designs. Compared with analog trigger circuits, digital triggers can not only greatly optimize trigger precision and trigger jitter, but also support multiple trigger types and complex trigger conditions.
SDS2000X HD Series Digital Oscilloscope User Manual 15.5.2 Edge Trigger Edge trigger distinguishes the trigger points by seeking the specified edge (rising, falling, alternating) and trigger level. The trigger source and slope can be set in the trigger dialog box. Touch the Source area to select trigger source, and touch the Slope area to select rising, falling, or alternating.
Page 93
SDS2000X HD Series Digital Oscilloscope User Manual Rising -- Only trigger on the positive slope Falling -- Only trigger on the negative slope Adjust Upper/Lower Level The slope trigger requires upper and lower trigger levels. When the trigger type is slope trigger, touch the trigger descriptor box, and the pop-up quick menu will show two levels.
SDS2000X HD Series Digital Oscilloscope User Manual input signal is greater than the specified lower limit of time and lower than the specified upper limit of time value. Outside a range of time value(--][--)-- Trigger when the positive or negative slope time of the input signal is greater than the specified upper limit of time and lower than the specified lower limit of time value.
SDS2000X HD Series Digital Oscilloscope User Manual Within a range of time values ([--,--])-- Trigger when the positive or negative pulse time of the input signal is greater than the specified lower limit of time and lower than the specified upper limit of the time value.
Page 96
SDS2000X HD Series Digital Oscilloscope User Manual HDTV 720P/60 Progressive Tri-level HDTV 1080P/50 Progressive Tri-level HDTV 1080P/60 Progressive Tri-level HDTV 1080i/50 Interlaced Tri-level HDTV 1080i/60 Interlaced Tri-level Custom The table below shows the parameters of the custom video trigger: Frame Rate 25Hz, 30 Hz, 50 Hz,60 Hz Of Lines 300 ~ 2000...
Page 97
SDS2000X HD Series Digital Oscilloscope User Manual Set the video trigger for a video signal Touch Sync for trigger mode selection, video trigger mode has "Any" and "Select" options. In "Any" mode, the video signal can be triggered on any line that meets the conditions. In the "Select" mode, the signal can be triggered by a specified field and a specified line.
Page 98
SDS2000X HD Series Digital Oscilloscope User Manual Touch the Standard and select the "NTSC". Touch the Sync and select the "Select" to make the Field and Line optional, then select "1" in the "Field", and set the "Line" to "22" by using the universal knob or the virtual keypad. Trigger on a Specific Line of Video (NTSC) Use "Custom"...
SDS2000X HD Series Digital Oscilloscope User Manual Select the "Select" mode, then set the specified line and the specified field to trigger the signal. Assuming that the "Field" is set to 8, you can select any field from 1 to 8, and each field can choose any line from 1 to 100.
SDS2000X HD Series Digital Oscilloscope User Manual When the window trigger type is set to "Relative", touch the trigger descriptor box. The pop-up menu will show two user-defined parameters: “Level +/-Delta" and "Level Center". The above two parameters can be set in the following two ways: Select the parameter in the Level +/-Delta...
SDS2000X HD Series Digital Oscilloscope User Manual Trigger source, slope (rising, falling), limit range, and time value can be set in the trigger dialog box. Holdoff, coupling, and noise reject can be set in interval trigger, see the sections "Holdoff", "Trigger Coupling"...
SDS2000X HD Series Digital Oscilloscope User Manual 15.5.9 Runt Trigger Runt trigger looks for pulses that cross one threshold but not another as shown in the figure below: A positive runt pulse across through the low level but not the high level. ...
Page 103
SDS2000X HD Series Digital Oscilloscope User Manual Logic (AND, OR, NAND, NOR), source, limit range, and time value can be set in the trigger dialog box. Source Setting Touch the Source Setting area to recall the following dialog box and set up for each channel separately.
SDS2000X HD Series Digital Oscilloscope User Manual The logical setting of analog channels The logical setting of digital channels Limit Range This setting is particularly useful to filter the hazard signals of combinational logic. Holdoff can be set in pattern trigger, see the section "Holdoff" for details. 15.5.11 Qualified Trigger The qualified trigger is an edge trigger after a user-defined qualified condition is satisfied.
SDS2000X HD Series Digital Oscilloscope User Manual When the type is “Edge”, the oscilloscope triggers at the first edge after the specified edge (Rising or Falling) of the qualified source; when the type is “Edge with Delay”, a time limit condition is available. Touch the Qualified Setting region to set the qualified source and threshold;...
SDS2000X HD Series Digital Oscilloscope User Manual 15.5.13 Delay Trigger A delay trigger is true when the edge of source B occurs after meeting the set conditions of source A and a user-defined delay time. The setting of source A is similar to that of the pattern trigger and can be used for logical "and"...
SDS2000X HD Series Digital Oscilloscope User Manual Touch the CLK Setting region to set the clock source, threshold, and slope; Touch the Data Setting region to set the data source, threshold, and state. 15.5.15 Serial Trigger Refer to chapter "Serial Trigger and Decode" for detailed information. Trigger Source The trigger sources supported by each trigger type are different.
SDS2000X HD Series Digital Oscilloscope User Manual Holdoff Holdoff is an additional condition for triggers and can be used to stabilize the triggering of complex waveforms (such as a pulse series). It can be set to a time or number of events. Holdoff by Time Holdoff time is the amount of time that the oscilloscope waits before re-arming the trigger circuitry.
SDS2000X HD Series Digital Oscilloscope User Manual Acq Start-- The initial position of holdoff is the first time point satisfying the trigger condition. In the example above, each holdoff starts from the first rising edge of the pulse sequence. Last Trig Time--The initial position of holdoff is the time of the last trigger. In the example above, the last trigger position is at the second rising edge of the pulse sequence and the second holdoff starts from that point.
SDS2000X HD Series Digital Oscilloscope User Manual Noise Reject Noise Reject adds additional hysteresis to the trigger circuitry. By increasing the trigger hysteresis, the noise immunity becomes better but the trigger sensitivity degrades. Noise Reject = Off Noise Reject = On 1 0 8 i n t .
SDS2000X HD Series Digital Oscilloscope User Manual Zone Trigger The SDS2000X HD includes a zone trigger to help isolate elusive glitches. There are two user-defined areas: Zone1 and Zone2. Users can set the property of each zone as “intersect” or “not intersect” as an additional condition to further isolate the interesting event quickly.
Page 112
SDS2000X HD Series Digital Oscilloscope User Manual When the finger moves out of the screen, a menu pops up for selecting the zone and setting the zone properties: Once a zone is created, it can be moved by dragging. Just touch and hold the zone box and use a dragging gesture.
Page 113
SDS2000X HD Series Digital Oscilloscope User Manual Select C1 as the source, turn on zone1, Select C1 as the source, turn on zone1, and set the property as "Intersect" and set the property as "Not Intersect" Dialog Box Touch Zone > Zone Setting to recall the dialog box. A.
Page 114
SDS2000X HD Series Digital Oscilloscope User Manual With a simple edge trigger, it is unlikely to trigger consistently on this anomaly. Thanks to the high waveform update rate of the SDS2000X HD, we can confirm there is bus contention happening by enabling the persistence display, as shown in the figure below: In this case zone trigger is a quick and simple way to capture the waveform.
Page 115
SDS2000X HD Series Digital Oscilloscope User Manual Now, we can accurately capture the exact bus contention waveform: i n t . s i g l e n t . c o m 11 3...
SDS2000X HD Series Digital Oscilloscope User Manual 16 Serial Trigger and Decode Overview The SDS2000X HD supports serial bus trigger and decoding on the following serial bus protocols: I2C, SPI, UART, CAN, LIN, FlexRay, CAN FD, I2S, MIL-STD-1553B, SENT, and Manchester. Press the Setup button on the front panel or touch the trigger descriptor box, and then select the Type as Serial in the trigger dialog box to set the serial trigger: A.
Page 117
SDS2000X HD Series Digital Oscilloscope User Manual A. Set the list of decode result B. Select the bus to set, Bus1 and Bus2 C. Turn on/off the bus D. Set the bus display E. Select the serial bus protocol Touch to set the signal, including the mapping relation between channels and bus signals, and the thresholds.
SDS2000X HD Series Digital Oscilloscope User Manual I2C Trigger and Serial Decode This section covers triggering and decoding I2C signals. Please read the following for more details: "I2C Signal Settings", "I2C Trigger" and "I2C Serial Decode". 16.2.1 I2C Signal Settings Connect the serial data signal (SDA) and the serial clock signal (SCL) to the oscilloscope, set the mapping relation between channels and signals, and then set the threshold level of each signal.
SDS2000X HD Series Digital Oscilloscope User Manual Copy Setting Touch the Protocol Copy in the decode dialog box to synchronize the settings between the trigger and decode. A. Copy the decode settings to trigger B. Copy the trigger settings to decode C.
Page 120
SDS2000X HD Series Digital Oscilloscope User Manual Restart — The oscilloscope will be triggered when another “Start” occurs before a “Stop". No Ack — The oscilloscope will be triggered when the SDA line is high during any SCL’s ACK bit. EEPROM —...
Page 121
SDS2000X HD Series Digital Oscilloscope User Manual Frame (Start: 7-bit address: R/W: Ack: Data: Ack: Data2)—If all bits match, then trigger on the Ack bit followed by the Data2. 10 Address & Data — If all bits match, then trigger on the Ack bit followed by the Data. Frame (Start: Address 1st byte: R/W: Ack: Address 2nd byte: Ack: Data) i n t .
Page 122
SDS2000X HD Series Digital Oscilloscope User Manual If you set the trigger condition to 7 address & data or 10 address & data: Address can be selected in the hexadecimal range of 0x00 to 0x7F (7-bit) or 0x3FF (10-bit). If the address is selected as "0xXX (7-bit address)"...
SDS2000X HD Series Digital Oscilloscope User Manual 16.2.3 I2C Serial Decode The layout of the touch screen display when I2C decode is enabled is as follows: A. The waveform display area shows the original waveforms of the bus signals B. The bus display shows the decoding result of the bus. At most two buses can be decoded at the same time.
Page 124
SDS2000X HD Series Digital Oscilloscope User Manual The data value is displayed in white. A red point at the end of a segment indicates there is not enough space on the display to show the complete content of a frame, and some content is hidden, such as List TIME —...
SDS2000X HD Series Digital Oscilloscope User Manual SPI Trigger and Serial Decode This section covers triggering and decoding SPI signals. Please read the following for more details: "SPI Signal Settings", "SPI Trigger" and "SPI Serial Decode". 16.3.1 SPI Signal Settings Connect the CLK, MOSI, MISO, and CS signals to the oscilloscope and set the mapping relation between channels and signals.
Page 126
SDS2000X HD Series Digital Oscilloscope User Manual The CS signal should be set to correct CS Type , including CS, ~CS, and Clock Timeout. CS – Active high. The CS signal needs a complete rising edge in the screen to be regarded as active.
Page 127
SDS2000X HD Series Digital Oscilloscope User Manual When the CS type is set to Clock Timeout, turn on Cursor, measure the clock idle time between frames as 150 us, and measure the interval between clock pulses as 1.28 us, then set the timeout to a value between 1.28 us and 150 us.
SDS2000X HD Series Digital Oscilloscope User Manual 16.3.2 SPI Trigger The trigger condition for the SPI trigger is mainly about data. Touch Trigger Setting in the dialog box to set data: A. Trigger Type: MISO or MOSI B. Set the trigger position type. When set to Any, it will trigger on the specified data.
SDS2000X HD Series Digital Oscilloscope User Manual UART Trigger and Serial Decode This section covers triggering and decoding UART signals. Please read the following for more details: "UART Signal Settings", "UART Trigger" and "UART Serial Decode". 16.4.1 UART Signal Settings Connect the RX and TX signals to the oscilloscope, set the mapping relation between channels and signals, and then set the threshold level of each signal.
SDS2000X HD Series Digital Oscilloscope User Manual 16.4.2 UART Trigger Touch Trigger Setting in the dialog box to set the trigger condition: A. Source Type: RX or TX B. Trigger Condition: Start, Stop, Data or Error C. When the "trigger condition" is Data, set the compare type to =, >, <...
SDS2000X HD Series Digital Oscilloscope User Manual CAN Trigger and Serial Decode This section covers triggering and decoding CAN signals. Please read the following for more details: "CAN Signal Settings", "CAN Trigger" and "CAN Serial Decode". 16.5.1 CAN Signal Settings Connect the CAN_H and CAN_L signals to the oscilloscope set the mapping relation between channels and signals, and then set the threshold level of each signal.
SDS2000X HD Series Digital Oscilloscope User Manual 16.5.3 CAN Serial Decode The configuration of CAN decoding is similar to that of I2C decoding. On the bus: ID is displayed in green. LEN (data length) is displayed in light yellow. ...
SDS2000X HD Series Digital Oscilloscope User Manual LIN Trigger and Serial Decode This section covers triggering and decoding LIN signals. Please read the following for more details: “LIN Signal Settings”, “LIN Trigger” and “LIN Serial Decode”. 16.6.1 LIN Signal Settings Connect the LIN signal to the oscilloscope, and then set the threshold level of the signal.
SDS2000X HD Series Digital Oscilloscope User Manual 16.6.3 LIN Serial Decode The configuration of LIN decoding is similar to that of I2C decoding. On the bus: ID is displayed in green LEN (data length) and CHK are displayed in blue ...
SDS2000X HD Series Digital Oscilloscope User Manual FlexRay Trigger and Serial Decode This section covers triggering and decoding FlexRay signals. Please read the following for more details: "FlexRay Signal Settings", "FlexRay Trigger" and "FlexRay Serial Decode". 16.7.1 FlexRay Signal Settings Connect the FlexRay signal to the oscilloscope, and then set the threshold level of the signal.
SDS2000X HD Series Digital Oscilloscope User Manual Symbol— The oscilloscope triggers on CAS/MTS (Conflict Avoidance Symbol/ Media access Test Symbol) or WUS (Wake-up Symbol) Errors— The oscilloscope will trigger when a data error happens, including errors on FSS, BSS, FES, Header CRC, and Frame CRC.
Page 137
SDS2000X HD Series Digital Oscilloscope User Manual In the list view: Time— The horizontal offset of the current data frame head relative to the trigger position. FID— Frame ID, the symbol occupies a single line of the list. ...
SDS2000X HD Series Digital Oscilloscope User Manual CAN FD Trigger and Serial Decode This section covers triggering and decoding CAN FD signals. Please read the following for more details: "CAN FD Signal Settings", "CAN FD Trigger" and "CAN FD Serial Decode". 16.8.1 CAN FD Signal Settings Connect the CAN FD signal to the oscilloscope, and then set the threshold level of the signal.
SDS2000X HD Series Digital Oscilloscope User Manual Trigger Condition Start— The oscilloscope triggers at the beginning of the frame. Remote— The oscilloscope triggers on a remote frame with a specified ID. ID, ID Bits (11-bit or 29-bit), and Curr ID Byte (1st, 2nd, 3rd, or 4th byte) can be set. Curr ID Byte is used to specify the byte to be adjusted when using the universal knob.
Page 140
SDS2000X HD Series Digital Oscilloscope User Manual The red point at the end of a segment indicates there is not enough space on the display to show the complete content of a frame and some content is hidden. In the list view: ...
SDS2000X HD Series Digital Oscilloscope User Manual I2S Trigger and Serial Decode This section covers triggering and decoding I2S signals. Please read the following for more details: "I2S Signal Settings", "I2S Trigger" and "I2S Serial Decode". 16.9.1 I2S Signal Settings Connect the WS, BCLK, and Data signals to the oscilloscope, set the mapping relation between channels and signals, and then set the threshold level of each signal.
SDS2000X HD Series Digital Oscilloscope User Manual 16.9.2 I2S Trigger Touch Trigger Setting in the I2S trigger dialog box to set the trigger conditions: A. Audio Variant: Audio-I2S, Audio-LJ, Audio-RJ B. Trigger Condition: Data, Mute, Clip, Glitch, Rising Edge, Falling Edge C.
SDS2000X HD Series Digital Oscilloscope User Manual Clip — Trigger on the clipped signal. Clip signal: Volume is greater than the set value and duration reaches the set value. Touch Clip Level to set the clip threshold by the universal knob or virtual keypad. The range of values is related to the number of Data Bits .
SDS2000X HD Series Digital Oscilloscope User Manual MIL-STD-1553B Trigger and Serial Decode This section covers triggering and decoding MIL-Standard 1553B signals. Please read the following for more details: "MIL-STD-1553B Signal Settings" and "MIL-STD-1553B Serial Decode". 16.10.1 MIL-STD-1553B Signal Settings Connect the MIL-STD-1553B signal to the oscilloscope, set the mapping relation between channels and signals, and then set the threshold level of each signal.
Page 145
SDS2000X HD Series Digital Oscilloscope User Manual In the decoding list view: Time— The horizontal offset of the current data frame head relative to the trigger position. RTA— The RT address Type— Type of the word Data—...
SDS2000X HD Series Digital Oscilloscope User Manual SENT Trigger and Serial Decode This section covers triggering and decoding SENT signals. Please read the following for more details: "SENT Signal Settings", "SENT Trigger" and "SENT Serial Decode". 16.11.1 SENT Signal Settings Connect the SENT signal to the oscilloscope, set the mapping relation between channels and signals, and then set the threshold level of each signal.
Page 147
SDS2000X HD Series Digital Oscilloscope User Manual Touch Trigger Setting in the SENT trigger dialog box to select the trigger condition: Start — The oscilloscope will be triggered at the start of the message (after 56 synchronization ticks). You can select the type of message: Fast Channel Message, Slow Channel Message, or any. Fast Channel —...
Page 148
SDS2000X HD Series Digital Oscilloscope User Manual The serial message frame contains 21 bits of payload data. Two different configurations can be chosen determined by the configuration bit (serial data bit #3, serial communication nibble No. 8): Frame (enhanced serial message with 4bits ID) —16-bit data and 4-bit message ID, the configuration bit is 1.
Page 149
SDS2000X HD Series Digital Oscilloscope User Manual If you set the trigger condition to Slow Channel: ID can be selected in the hexadecimal range of 0x0 to 0xF (short serial / enhanced serial with 4 bits ID)or 0x00 to 0xFF(enhanced serial with 8 bits ID).
SDS2000X HD Series Digital Oscilloscope User Manual 16.11.3 SENT Serial Decode The configuration of SENT decoding is similar to that of I2C decoding. On the bus: For Fast Channel: SYNC is displayed in pink STATE is displayed in green ...
Page 151
SDS2000X HD Series Digital Oscilloscope User Manual In the list view: Time — The horizontal offset of the current data frame head relative to the trigger position. Sync — Sync pulse width (only fast channel) State — Status & Communication nibble (only fast channel) ...
SDS2000X HD Series Digital Oscilloscope User Manual Manchester Serial Decode This section covers decoding Manchester signals. Please read the following for more details: "Manchester Signal Settings" and "Manchester Serial Decode". 16.12.1 Manchester Signal Settings Connect the Manchester signal to the oscilloscope, set the mapping relation between channels and signals, and then set the threshold level of the signal.
SDS2000X HD Series Digital Oscilloscope User Manual 16.12.2 Manchester Serial Decode The configuration of Manchester decoding is similar to that of I2C decoding. On the bus: SYNC is displayed in pink The header is displayed in green DATA is displayed in white ...
SDS2000X HD Series Digital Oscilloscope User Manual 17 Cursors Overview Cursors are important tools when measuring signals. Rapid measurements can be performed using cursors in both horizontal and vertical directions. The cursor types include X1, X2, X1-X2, Y1, Y2, and Y1-Y2, used to indicate X-axis values (time or frequency) and Y-axis values (amplitude) on a selected waveform (CH1/CH2/CH3/CH4/Func/REF/Histogram).
Page 155
SDS2000X HD Series Digital Oscilloscope User Manual Cursors Mode Manual -- Manually sets the cursors’ position. Cursors type (horizontal, vertical, horizontal + vertical) is available in this mode. Track -- The cursor type is automatically set to "horizontal + vertical". In this mode, only horizontal cursors are adjustable, while the vertical cursors automatically attach to the cross-point of the cursor and the source waveform.
Page 156
SDS2000X HD Series Digital Oscilloscope User Manual Cursors Type X (horizontal) -- Vertical dotted lines that measure horizontal time (when the source is an FFT waveform, X cursors measure frequency). X cursors (time) X cursors (frequency) X1 — The left (default) vertical dotted line. It can be manually moved to any horizontal position on the screen.
Page 157
SDS2000X HD Series Digital Oscilloscope User Manual Y1 — The upper (default) horizontal dotted line. It can be manually moved to any vertical position on the screen. Y2 — The lower (default) horizontal dotted line. It can be manually moved to any vertical position on the screen.
Page 158
SDS2000X HD Series Digital Oscilloscope User Manual Cursors Reference X cursors reference: Fixed Delay– When the timebase is changed, the value of X cursors remains fixed. Fixed Position– When the timebase is changed, the X cursors remain fixed to the grid position on the display.
Page 159
SDS2000X HD Series Digital Oscilloscope User Manual Fixed position, timebase is changed to 100 ns/div, the grid number of X cursors (-1div, 2div) remains fixed. The value of X1 and X2 is changed to -100 ns, 200 ns. Fixed delay, timebase is changed to 100 ns/div, the value of X cursors (-50 ns, 100 ns) remains fixed.
SDS2000X HD Series Digital Oscilloscope User Manual Select and Move Cursors The cursors can be selected and moved directly by gestures and the universal knob on the front panel, in addition, they can be selected in the cursor’s value dialog box. Gestures Directly touch the cursor and drag it, as shown below: Touch the display area of △...
Page 161
SDS2000X HD Series Digital Oscilloscope User Manual Universal Knob Move the cursor position by turning the universal knob on the front panel. Press the knob to select different cursor lines. For example, if the current cursor is X1, press to select X2, and press again to select X1-X2.
SDS2000X HD Series Digital Oscilloscope User Manual 18 Measurement Overview The SDS2000X HD features a strong automatic measurement list. These parameters can be automatically measured without cursors and include common measurements such as rise time, fall time, peak-peak, and period. The SDS2000X HD can also measure multiple channels at the same time, showing up to 5 parameter measurements with statistics while in the M1 display mode and up to 12 parameters in the M2 mode.
Page 163
SDS2000X HD Series Digital Oscilloscope User Manual D. Measure dialog box Press the Measure button on the front panel or touch Measure > Menu to open the dialog box. A. Enable/disable measure B. Set the mode of measure: Simple or Advanced. "Simple" shows the specified basic measurement parameters of the selected channel.
SDS2000X HD Series Digital Oscilloscope User Manual Set Parameters Touch Type in the measure dialog box, or touch in the measurement parameters and statistics display area to open the parameter selection window: A. Set the source of the current setting. B.
Page 165
SDS2000X HD Series Digital Oscilloscope User Manual In the parameter selection area, the channel corresponding to Source A is specified first, and then the channel corresponding to Source B . Finally, the measurement parameter is selected. For example, to activate the skew between C1 and C2, you can follow the following steps: Source A >...
Page 166
SDS2000X HD Series Digital Oscilloscope User Manual To remove Period from the “Favorite” tab: Advanced > Type > Favorite > Period > Delete from favorite 1 6 4 i n t . s i g l e n t . c o m...
SDS2000X HD Series Digital Oscilloscope User Manual Type of Measurement 18.3.1 Vertical Measurement Vertical measurement includes 19 parameters: Max: Highest value in the input waveform Min: Lowest value in the input waveform Pk-Pk: Difference between maximum and minimum data values ...
Page 168
SDS2000X HD Series Digital Oscilloscope User Manual RMS: Root mean square of the data Cycle RMS: Root mean square of the data in the first cycle Median: Value at which 50% of the measurements are above and 50% are below ...
SDS2000X HD Series Digital Oscilloscope User Manual L@T: Level measured at trigger position 18.3.2 Horizontal Measurement Horizontal measurement includes 17 parameters: Period: Time between the middle threshold points of two consecutive like-polarity edges. Freq: Reciprocal of the period ...
SDS2000X HD Series Digital Oscilloscope User Manual Time@min: First time of minimum value +Width: Time difference between the middle threshold of a rising edge to the middle threshold of the next falling edge of the pulse -Width: Time difference between the middle threshold of a falling edge to the middle threshold of the next rising edge of the pulse ...
SDS2000X HD Series Digital Oscilloscope User Manual +Area@DC: Area of the waveform above zero -Area@DC: Area of the waveform below zero Area@DC: Area of the waveform AbsArea@DC: Absolute area of the waveform +Area@AC: Area of the waveform above average ...
Page 172
SDS2000X HD Series Digital Oscilloscope User Manual Phase: Phase difference between two edges Skew: Time of source A edge minus time of nearest source B edge FRFR: The time between the first rising edge of source A and the following first rising edge of source B at the middle threshold ...
SDS2000X HD Series Digital Oscilloscope User Manual Trend After adding a measurement parameter, a Trend can be used to observe the long-term change of the selected measurement value over time. A. Measurement parameter display area B. Trend plot display area C.
SDS2000X HD Series Digital Oscilloscope User Manual Track The measure values VS. time plot of a horizontal parameter (e.g. frequency, rise time) in one frame can be observed when the Track is enabled. The upper limit of the statistic number in a frame is set in Statistics > AIM Limit , which means the values exceeding the limit will not be shown in the track plot.
Page 175
SDS2000X HD Series Digital Oscilloscope User Manual In M2 mode, up to 12 parameter measurements are displayed at a time. When statistics are enabled, they are distributed on the right side of the measurement item. Touch a row to add or replace a measurement.
SDS2000X HD Series Digital Oscilloscope User Manual Measurement Statistics Measurement statistics are based on the total number of captured waveforms. In Roll mode, measurement statistics increase over time. Touch Statistics Setting in the measure dialog box to recall the Statistics Config dialog box: A.
Page 177
SDS2000X HD Series Digital Oscilloscope User Manual Press the Clear Sweeps button or touch Reset Statistics in the statistics setting dialog box, or touch the symbol in the statistics display area to clear and restart statistics. When the oscilloscope detects the trace under test is clipped, an extra overflow indicator will appear after the measurement value: Waveform clipped at the top Waveform clipped at the bottom...
SDS2000X HD Series Digital Oscilloscope User Manual Statistics Histogram After enabling statistics on a selected measurement, you can also activate a statistical histogram. The histogram appears at the bottom of the statistics area. This enables users to quickly view the probability distribution of the measured parameters.
SDS2000X HD Series Digital Oscilloscope User Manual Simple Measurements Enabling Simple Measurement displays all selected measurement parameters of the specified channel at the same time. The font color of the measurement parameters is consistent with the color of the specified source. Yellow for Channel 1, Purple for Channel 2, etc. Gate Sometimes the user may want to measure parameters for a certain specified time range of the signal and ignore signal parts that lie outside of that range.
SDS2000X HD Series Digital Oscilloscope User Manual Amplitude Strategy According to different types of input signals, users can choose the corresponding amplitude calculation strategy which can measure top and bottom values with more accuracy. Touch Config > Amplitude Strategy in the measure dialog box to recall the amplitude strategy dialog box: A.
SDS2000X HD Series Digital Oscilloscope User Manual C. Set the base value calculation strategy. When set to Histogram, the value at the lower half of the waveform will be counted, and the value with the maximum probability will be identified as the base value; when set to min, the minimum value of the waveform will be identified as the base value D.
SDS2000X HD Series Digital Oscilloscope User Manual Threshold Type Percent: Set according to the percentage of the waveform. The setting range of lower value and upper value is 1% ~ 99%, and the low value shall not be greater than the middle value and high value Absolute: Set according to the vertical range.
SDS2000X HD Series Digital Oscilloscope User Manual 19 Math Overview The SDS2000X HD supports 2 math traces and multiple operators. Arithmetic operators: Addition (+), subtraction (-), multiplication (x), division (/), average, ERES, identity, negation, maxhold, minhold; Algebra operators: differential (d/dt), integral (∫dt), square root (√), absolute (|y|), sign, exp, ln, interpolate;...
Page 184
SDS2000X HD Series Digital Oscilloscope User Manual Math Operation Unit Addition (+) or V, A, or U* Subtraction (-) *(used when the units of two sources are not consistent) ∧ ∧ Multiplication (x) 2, A 2, or W None, Ω (Resistance unit Ohms), S (conductance Division (/) unit Siemens) dBVrms, Vrms, dBArms, Arms, dBm...
SDS2000X HD Series Digital Oscilloscope User Manual Arithmetic The SDS2000X HD supports Addition, Subtraction, Multiplication, Division, Average, ERES, Identity, Negation, Max-hold, and Min-hold. 19.2.1 Addition / Subtraction / Multiplication / Division The SDS2000X HD can perform arithmetic operations including addition, subtraction, multiplication, or division on any two analog input channels, and the values of Source A and Source B are computed point-by-point.
SDS2000X HD Series Digital Oscilloscope User Manual 19.2.3 Average / ERES Average / ERES can be set in acquisition mode and arithmetic function as well, but with different compute methods. Average and ERES in acquisition mode is computed by hardware with a higher speed and sampling rate than computed by software in the arithmetic mode.
SDS2000X HD Series Digital Oscilloscope User Manual Algebra The SDS2000X HD can perform algebraic operations including differential (d/dt), integral (∫dt), square root (√), absolute(|x|), sign, exp ( ), exp10 (10 ), ln, lg and interpolate (Intrp). 19.3.1 Differential The differential (d/dt) operator is used to calculate the derivative of the selected source.
SDS2000X HD Series Digital Oscilloscope User Manual The range of “dx” in the d/dt menu is 4~ number of samples pts. Note: Differentiation is sensitive to noise. It is helpful to set the acquisition mode to “Average” or “ERES” to help minimize the visible effects of additional noise. 19.3.2 Integral Integral operation integrates the waveforms on the screen or within the specified gate.
SDS2000X HD Series Digital Oscilloscope User Manual 19.3.3 Square Root Square root (√) calculates the square root of the selected source. If the waveform value is negative (the waveform is below the ground level), the result is displayed as zero. 19.3.4 Absolute Absolute(|x|)calculates the absolute value of the selected trace.
SDS2000X HD Series Digital Oscilloscope User Manual 19.3.5 Sign In mathematics, the sign function or signum function (from signum, Latin for "sign") is an odd mathematical function that extracts the sign of a real number. The sign function of a real number x is defined as follows: Sign(x) = -1, if x <...
SDS2000X HD Series Digital Oscilloscope User Manual 19.3.6 exp10 The exponential operation includes the exponential operation based on constant e and the exponential operation 10 based on 10. For example: y(x) = i n t . s i g l e n t . c o m 1 8 9...
SDS2000X HD Series Digital Oscilloscope User Manual 19.3.7 Logarithmic operation includes logarithm base e (ln) and logarithm base 10 (lg). In logarithmic operation, if the waveform value is negative (the waveform is below the ground level), the result is displayed as zero. ...
SDS2000X HD Series Digital Oscilloscope User Manual Frequency Analysis The result of FFT (Fast Fourier Transform) calculations is the frequency spectrum of the source signal. The horizontal axis of the FFT display is labeled using frequency (Hz) units instead of time (seconds). In addition, the vertical axis provides the option of logarithmic scaling (dBVrms/dBArms or dBm).
Page 194
SDS2000X HD Series Digital Oscilloscope User Manual rate in the time domain. Assuming the max points value is set to 2 Mpts: When the number of points in the time domain, N, is less than 2 Mpts, the FFT takes the number which is an integer power of 2 closest to N.
Page 195
SDS2000X HD Series Digital Oscilloscope User Manual Windows Spectral leakage in FFTs can be considerably decreased when a window is used. SDS2000X HD provides five windows that have different characteristics and apply to different scenarios. For example, for a two-tone signal with a very close frequency interval, it is suitable to use a Rectangle window with the best frequency resolution.
Page 196
SDS2000X HD Series Digital Oscilloscope User Manual Split Mode, Zoom off Split Mode, Zoom on Full-Screen Mode Exclusive Mode FFT Mode Normal: Displays the FFT result of each frame directly. Max-Hold: Holds the maximum value in the historic frame on the display until cleared. This mode is suitable for detecting discontinuous waves, such as sporadic pulse signals, or frequency hopping signals.
Page 197
SDS2000X HD Series Digital Oscilloscope User Manual Touch Vertical Horizontal in the math dialog box to recall the FFT vertical or horizontal settings dialog box: A. Set the vertical scale and reference level B. Set the unit (dBVrms, Vrms, and dBm). When the unit is dBm, the oscilloscope will automatically calculate the dBm value based on the Ext Load value set in area...
Page 198
SDS2000X HD Series Digital Oscilloscope User Manual Touch Scale to set the vertical scale of the FFT waveform by the universal knob or the virtual keypad. It can also be set by the scale knob shared by Decode, Digital, Math, and Ref on the front panel.
Page 199
SDS2000X HD Series Digital Oscilloscope User Manual When the FFT tool Peaks is selected, the dialog box is as follows: A. Turn on or off the table. Turn on the table, the peaks searched with the limit of Search Threshold and Search Excursion will be displayed in a table B.
Page 200
SDS2000X HD Series Digital Oscilloscope User Manual Note: The FFT dialog box is longer than the display. Slide the dialog box area up and down by gestures, or scroll the mouse wheel to the view non-displayed areas. The following shows the peaks of the FFT waveform: Measure the FFT waveform Press the Cursors button on the front panel to turn on the cursor function, and then specify the source as "Math".
SDS2000X HD Series Digital Oscilloscope User Manual Formula Editor Touch Formula Editor in the operation setting page to recall the editor: A. Formula display text box B. Text box operation area, which can clear and modify the entered formula C. Special operators D.
Page 202
SDS2000X HD Series Digital Oscilloscope User Manual 2 0 0 i n t . s i g l e n t . c o m...
SDS2000X HD Series Digital Oscilloscope User Manual 20 Reference Data from analog channels math saved reference locations (REFA/REFB/REFC/REFD) in the built-in nonvolatile memory. The saved reference waveform can be recalled to be compared with the current waveform. A. Channel waveform B.
Page 204
SDS2000X HD Series Digital Oscilloscope User Manual A. Select the location of the reference (REFA/REFB/REFC/REFD) B. Select the source (C1~C4 and MATH) C. Enable/disable Display of the reference waveform D. Set the label text of the reference trace. E. Save the specified waveform in to the specified location in Adjust the Reference Waveform...
SDS2000X HD Series Digital Oscilloscope User Manual 21 Search The SDS2000X HD can search for the specified events in a frame. The location of the events is displayed with white triangle indicators. In YT mode or Roll mode with the acquisition stopped, up to 1000 events are supported.
Page 206
SDS2000X HD Series Digital Oscilloscope User Manual Touch the menu Analysis > Search to recall the search dialog box and turn on it. Setup Menu Select and set the search type in the Setup Menu . The SDS2000X HD provides five search conditions: Edge, Slope, Pulse, Interval, and Runt.
Page 207
SDS2000X HD Series Digital Oscilloscope User Manual Search setup is similar to the corresponding trigger type. See the sections “Edge Trigger”, “Slope Trigger”, ”Pulse Trigger”, ”Interval Trigger” and ”Runt Trigger” for details. Copy The SDS2000X HD supports replication between search settings and trigger settings. ...
SDS2000X HD Series Digital Oscilloscope User Manual 22 Navigate Touch the menu Analysis > Navigate to recall the navigate dialog box. The SDS2000X HD provides three navigate types: Search Event, Time, and History Frame. Navigate by Time The oscilloscope automatically adjusts the trigger delay according to the direction set by the user.
Page 209
SDS2000X HD Series Digital Oscilloscope User Manual Navigate by Search Event When the Search function is turned on and the acquisition is stopped, Navigate is usable to find search events (see the chapter "Search" for search function). Touch Type in the navigate dialog box to select the navigate type as “Search Event”.
Page 210
SDS2000X HD Series Digital Oscilloscope User Manual Navigate by History Frame When the history function is turned on, Navigate can be used to play history frames (see the chapter "History" for details of the history function). Touch Type in the navigate dialog box to select the navigate type as “History Frame”.
Page 211
SDS2000X HD Series Digital Oscilloscope User Manual First, set the trigger type to Runt to trigger on the dwarf pulse. See the section “Runt Trigger” for details. Then turn on the Search function and operate Copy from Trigger in the search dialog box so that the oscilloscope searches for the dwarf pulses according to the same setting as the trigger.
Page 212
SDS2000X HD Series Digital Oscilloscope User Manual Turn on the Zoom function to observe the full view of the frame and the detail of the third dwarf pulse at the same time: Press the Run/Stop button on the front panel to stop the acquisition, and then follow the steps Navigate >...
SDS2000X HD Series Digital Oscilloscope User Manual 23 Mask Test Overview Users can create masks and define the rule used to evaluate Pass/Fail operations. An event violating the rule is defined as a failure and a pulse can be generated from the "Aux Out" port on the back panel.
Page 214
SDS2000X HD Series Digital Oscilloscope User Manual Perform Analysis > Mask Test to open the Mask Test dialog box: A. Turn on/off the test B. Select the source(C1~C4) C. Select the rule(All In, All Out, Any In, and Any Out) D.
SDS2000X HD Series Digital Oscilloscope User Manual Mask Setup Touch Mask Setup in the Mask Test dialog box to set the mask. There are two methods to create a mask, one is by setting horizontal and vertical values, and the other is by drawing a polygon mask. A.
SDS2000X HD Series Digital Oscilloscope User Manual X = 0.2 div, Y = 0.2 div X = 1 div, Y = 1 div Saving and recalling mask files (*.msk) is similar to the operation of setup files, see the chapter "Save/Recall"...
Page 217
SDS2000X HD Series Digital Oscilloscope User Manual D. Toolbar E. Coordinate edit area. Set the X ordinate and Y ordinate by the virtual keypad and then touch the “Input” button the perform the ordinate update Display or hide the coordinates of the polygon vertices on the display G.
SDS2000X HD Series Digital Oscilloscope User Manual Select a vertex (point) Select a side (line segment) Select a polygon Pass/Fail Rule The Pass/Fail rule is specified at the Type region in the Mask Test dialog box. All In: All data points must be inside the mask to pass the test. Even a single point outside the mask will cause a failure.
SDS2000X HD Series Digital Oscilloscope User Manual 24 DVM Overview The DVM (Digital Voltage Meter) function can be used to measure parameters such as DC and AC amplitudes. The SDS2000X HD measures the specified parameter of the input signal and can display it in various formats, including Bar, Histogram, and Trend.
SDS2000X HD Series Digital Oscilloscope User Manual Touch the menu Analysis > DVM to open the DVM dialog box: Turn on or off the DVM Select the source (C1~C4) Select the Mode: DC Mean, DC RMS, AC RMS, Peak-Peak, and Amplitude Turn on or off the Auto Range Turn on or off the bar diagram Turn on or off the histogram diagram...
SDS2000X HD Series Digital Oscilloscope User Manual DC Mean: Average of the data DC RMS: Root mean square of the data at DC coupling AC RMS: Root mean square of the data at AC coupling Peak-Peak: Difference between maximum and minimum data value ...
Page 222
SDS2000X HD Series Digital Oscilloscope User Manual A. Mode B. Histogram display area C. Current value D. The bin includes the maximum value and probability with which values fall into it E. Statistics counts Reset statistics Trend The trend diagram indicates the trend of the measured values over time. Touch trend in the DVM dialog box to display it.
Page 223
SDS2000X HD Series Digital Oscilloscope User Manual D. Current value E. Maximum value Minimum value G. Average value H. Reset statistics Press the Clear Sweeps button on the front panel to restart counting and statistics. Touch the symbol in the histogram and trend display areas to restart statistics. When all the 3 diagrams are closed, there is a simple information bar on the top-left of the waveform display area to show the current value of the DVM: i n t .
SDS2000X HD Series Digital Oscilloscope User Manual 25 Counter Overview The counter is used to measure the frequency and period of a signal or count the events happening within it. The counter is asynchronous to the acquisition system of the oscilloscope. It can work well even if the acquisition of the oscilloscope is stopped (indicated by a red-colored Run/Stop button) Touch the menu Analysis >...
SDS2000X HD Series Digital Oscilloscope User Manual Value – The latest count Mean – The average of all historical counts Min – The minimum of all historical counts Max – The maximum of all historical counts Stdv – The standard deviation of all historical counts, used to judge the distribution of historical count parameters Count –...
Page 226
SDS2000X HD Series Digital Oscilloscope User Manual Frequency: Average frequency over a set period Period: The reciprocal of the average frequency over a set period Totalizer: Cumulative count When the mode is set to totalizer, the edge of the counting source needs to be set. When the counting gate is turned on, the counter can count only when the gate-source meets the specified conditions.
SDS2000X HD Series Digital Oscilloscope User Manual 26 Histogram Overview The SDS2000X HD supports waveform histograms for observing probability distributions of the waveform in the specified region. The statistics can be performed in both horizontal and vertical directions. The histogram continues to update as long as the acquisition is active. Touch Analysis >...
Page 228
SDS2000X HD Series Digital Oscilloscope User Manual Horizontal Vertical Both Region Setting See section “Region Setting” for details. Histogram Statistics When histogram statistics is turned on, the oscilloscope will show the statistical parameters of the histogram on the display. 2 2 6 i n t .
SDS2000X HD Series Digital Oscilloscope User Manual Sum – Total samples falling in the histogram region Peak – Samples in the highest bin Max – Maximum value of the samples Min – Minimum value of the samples Pk-Pk – Maximum - Minimum Mean –...
Page 230
SDS2000X HD Series Digital Oscilloscope User Manual When the finger moves out of the touch screen, a menu pops up. Select "Histogram" in the menu: After the region is created, it can be moved by dragging gesture. 2 2 8 i n t .
Page 231
SDS2000X HD Series Digital Oscilloscope User Manual Dialog Box Touch Histogram > Region Setting to recall the dialog box. A. Set the left border of the histogram region B. Set the right border of the histogram region C. Set the top border of the histogram region D.
Transient Response, PSRR, Power Efficiency, Output Ripple, etc. Full use of the Power analysis requires a differential voltage probe like the SIGLENT DPB series, a current probe like the SIGLENT CP series, the SIGLENT DF2001A deskew fixture, and the power analysis activation license (part number SDS2000HD-PA).
Page 233
SDS2000X HD Series Digital Oscilloscope User Manual −1 ∗ ∑ ∗ P: Active Power = =0 S: Apparent Power = Vrms * Irms Q: Reactive Power =√Apparent Power − Active Power φ: Power Phase Angle: Phase difference between voltage and current cos φ: Power Factor, which is the ratio of active power and apparent power.
Page 234
SDS2000X HD Series Digital Oscilloscope User Manual A. Set the input voltage source B. Set the input current source C. Set the periods displayed D. View the connection guide E. Return to the previous menu Connection Guide Touch Signal > Connection Guide to recall the connection guide of power quality, as shown in the figure below.
SDS2000X HD Series Digital Oscilloscope User Manual Current Harmonics Current harmonics are used to analyze the input current harmonics. An FFT of the selected channel is performed to get the harmonic components. The signal settings and connection guide are the same as the power quality test.
Page 236
SDS2000X HD Series Digital Oscilloscope User Manual Limit Value (RMS) -- Limits specified by the selected standard Margin (%) --The margin specified by selected standard parameters. The margin value is (standard value - measured value) / standard value * 100% Pass/Fail State -- Determine whether the measured value is passed or failed according to the selected standard.
SDS2000X HD Series Digital Oscilloscope User Manual Inrush Current A large current far greater than the stable current may flow through at the moment of switching on a power supply. The large current is called the inrush current. The current waveform when switching on is shown in the figure below: Signal Settings Touch Signal to recall signal settings dialog box:...
SDS2000X HD Series Digital Oscilloscope User Manual Switching Loss Switch loss analysis can be used to calculate the power dissipated in the switching period. Deskew Calibration A relatively small skew can cause a large measurement error of switching loss, especially during the on phase when the voltage is close to zero and the non-on phase when the current is close to zero.
Page 239
SDS2000X HD Series Digital Oscilloscope User Manual C. Connect the voltage and current probes to the oscilloscope input channel D. Touch Signal to enter the signal setting menu. Select the corresponding input channel, and then touch Auto Deskew to perform the calibration. When the process completes, a message will pop up indicating whether the deskew operation was successful or not.
Page 240
SDS2000X HD Series Digital Oscilloscope User Manual A. Set the voltage reference, i.e. the switch level at the edge of the input switch. This value is a percentage of the maximum switching voltage. Adjusting the value to ignore background noise. This value is used to determine the threshold hysteresis of the switch edge B.
SDS2000X HD Series Digital Oscilloscope User Manual Slew Rate The slew rate measures the change rate of voltage or current during switching. Connection Guide Touch Signal > Connection Guide to recall the connection guide of slew rate, as shown in the figure below.
SDS2000X HD Series Digital Oscilloscope User Manual Modulation Modulation analysis measures the control pulse signal of the switching device (MOSFET) and observes the pulse width, duty, period, frequency, and other trends of the control pulse signal in response to different events. Connection Guide Touch Signal >...
SDS2000X HD Series Digital Oscilloscope User Manual Turn On/Turn Off Turn-on analysis determines the time taken for the power supply to reach 90% of its steady-state output. Turn-off analysis determines the time taken for the power supply to fall to 10% of its maximum output voltage.
Page 244
SDS2000X HD Series Digital Oscilloscope User Manual Turn Off -- Determines the time it has taken for the power supply to be turned off to fall to a certain percentage of its maximum output voltage. The turning off time is between T2 and T1, where: T1 = When the input voltage finally falls to a certain percentage (usually 10%) of its maximum amplitude T2 = When the output DC voltage finally falls to a certain percentage (usually 10%) of its maximum...
SDS2000X HD Series Digital Oscilloscope User Manual Transient Response Transient response analysis can determine the response speed of the output voltage of the power supply to the change of the output load. This time starts from the first time that the output voltage exits the stable band and ends at the last time that the output voltage enters the stable band.
SDS2000X HD Series Digital Oscilloscope User Manual Note: The low and high current values are used to calculate the trigger level and adjust the oscilloscope vertical scale. After the load changes, the current value changes from low to high (or from high to low), and the oscilloscope triggers and performs to measure the transient response stability time.
SDS2000X HD Series Digital Oscilloscope User Manual Power Efficiency Power efficiency analysis can test the overall efficiency of the power supply by measuring the output power and input power. This analysis is only supported on 4-channel models because all of the input voltage, input current, output voltage, and output current are necessary for the measurement.
Page 248
SDS2000X HD Series Digital Oscilloscope User Manual Connection Guide Touch Signal > Connection Guide to recall the connection guide of power efficiency, as shown in the figure below. Please follow the instructions in this figure for connection. Touch the icon on the top right of the guide to close.
Page 249
SDS2000X HD Series Digital Oscilloscope User Manual Voltage and current waveforms on the MOSFET at power-up SOA, in which the mask is created according to the voltage limit, current limit, and power limit. The result shows “Pass” because all the stresses are within the mask. i n t .
At this time, either the SIGLENT SAG1021I arbitrary waveform generator or one of a SIGLENT SDG series arbitrary function generators are supported. During the sweep, the oscilloscope configures the generator output frequency and amplitude and then compares the input signal to the output of the DUT.
SDS2000X HD Series Digital Oscilloscope User Manual Configuration 28.2.1 Connection A. DUT input and output channels B. Channel gain. When it is set to Auto, the oscilloscope will automatically adapt the vertical scale according to the signal amplitude; when it is set to Hold, it will always keep the vertical scale before the test operation C.
Page 252
SDS2000X HD Series Digital Oscilloscope User Manual C. Set the sweep frequency. The frequency mode can be linear or decade. When it is set to linear, the corresponding center frequency and span frequency need to be set; when it is set to a decade, the corresponding start frequency and stop frequency need to be set D.
Page 253
SDS2000X HD Series Digital Oscilloscope User Manual Touch Nodes to set the signal node number through the universal knob, or touch ▲ to increase the nodes and ▼ to decrease. Touch the cell in the table area to set the frequency and amplitude of the corresponding node. Touch to activate the cell, adjust the value through the universal knob, or touch the cell again to call up the virtual keypad for setting.
SDS2000X HD Series Digital Oscilloscope User Manual Display Bode plot display settings include amplitude, phase, cursors, trace visibility and waveform type. Amplitude Set the amplitude coordinate axis of the Bode plot. Touch Display > Amplitude to recall the amplitude setting dialog box: A.
Page 255
SDS2000X HD Series Digital Oscilloscope User Manual A. Set the reference position to center or top B. Set the scale of the phase coordinate axis C. Set the reference level, the max value of the phase coordinate axis D. Set the phase unit (Degree or Rad) E.
SDS2000X HD Series Digital Oscilloscope User Manual Data Analysis With the data list, cursor measurement, and automatic measurement functions, the Bode plot curve can be analyzed in detail. The data list provides the information of each scanning point. The cursors can be used to flexibly measure the change of each position of the curve.
Page 257
SDS2000X HD Series Digital Oscilloscope User Manual A. Turn on/off the data list B. Set the selected row in the list. Adjust the universal knob to set, or directly touch the list display area to select a specific line. C. Set the data source D.
Page 258
SDS2000X HD Series Digital Oscilloscope User Manual E. Clear the current measurement Close the measurement configuration window For example, to add UF measurements for C2 and GM measurements for C3, follow the steps below: Measure > Position > Source > UF Measure >...
SDS2000X HD Series Digital Oscilloscope User Manual 29 Display Display settings include display type of waveform, color, persistence, grid type, trace brightness, graticule brightness, etc. Touch the menu Display > Menu to recall the display dialog box. A. Switch the waveform display type to Vectors (line display) or Dots B.
Page 260
SDS2000X HD Series Digital Oscilloscope User Manual Vectors: The samples are connected by lines (i.e. interpolated) and displayed. The interpolation methods include linear interpolation and sin(x)/x interpolation. See the section of "Acquisition Setup" for details of interpolation. Dots: Displays the raw samples directly. Vector Display Dot Display 2 5 8...
Page 261
SDS2000X HD Series Digital Oscilloscope User Manual Note: In the Run state, due to the high waveform update rate of the oscilloscope, the waveform displayed is the superposition of multiple frames. Therefore, what is seen using the dots display is not the discrete sampling points, but the display effect similar to equivalent sampling.
Page 262
SDS2000X HD Series Digital Oscilloscope User Manual Color grade On Color grade Off Set Persist With persistence, the oscilloscope updates the display with new acquisitions but does not erase the results of previous acquisitions in the specified period. All previous acquisitions are displayed with reduced intensity.
Page 263
SDS2000X HD Series Digital Oscilloscope User Manual Infinite: Select “Infinite”, previous acquisitions will never be erased until a Clear Display or Clear Sweeps operation is performed. Set Grid Full Grid: Display 8*10 grid Light Grid: Display 2*2 grid ...
Page 264
SDS2000X HD Series Digital Oscilloscope User Manual No Grid Color Setting The color setting supports user-defined trace colors. Touch the color rectangle box to set in the pop- up palette page. Touch the Default area to restore the default color for all channels. Select a color directly on the palette and preview its hue, saturation, light, and RGB values or directly modify the color parameters to customize the color.
Page 265
SDS2000X HD Series Digital Oscilloscope User Manual Axis Label Setting Turn on the axis label, the horizontal axis label will be displayed at the bottom of the grid, and the vertical axis label will be displayed at the left of the grid. There are two display modes of axis: moving and fixed.
Page 266
SDS2000X HD Series Digital Oscilloscope User Manual Moving mode Fixed mode 2 6 4 i n t . s i g l e n t . c o m...
SDS2000X HD Series Digital Oscilloscope User Manual 30 Arbitrary Waveform Generator Overview The SDS2000X HD supports a built-in arbitrary waveform/function generation when activated using option SDS2000HD-FG. The AWG functions include: 6 basic waveforms: Sine, Square, Ramp, Pulse, Noise, and DC ...
SDS2000X HD Series Digital Oscilloscope User Manual Output For safety, pressing the Wave Gen button does not automatically turn on the output of AWG. To turn on the output, touch Output in the AWG dialog box after pressing Wave Gen . When the output is enabled, the Wave Gen button lights up.
Common, Math, Engine, Window, and Trigo. Stored waveforms are located in the Stored menu. Users can edit arbitrary waveforms using SIGLENT EasyWaveX PC software, send the stored waveforms to the instrument through the remote interface, or import the stored waveforms through a U disk.
Page 270
SDS2000X HD Series Digital Oscilloscope User Manual Sync Output When the Sync Output is turned on, the Aux In/Out port of AWG outputs a CMOS signal with the same frequency as the basic waveform (except Noise and DC) and arbitrary waveform. The maximum supported frequency is 10 MHz.
SDS2000X HD Series Digital Oscilloscope User Manual System Device Info – Includes Model, Serial Number, Firmware Version, and Hardware Version of the AWG module. Upgrade The firmware here refers to the firmware of the SAG1021I module. The SDS2000X HD supports firmware and configuration file upgrades for the SAG1021I via a U disk.
SDS2000X HD Series Digital Oscilloscope User Manual 31 Save/Recall The SDS2000X HD supports saving setups, reference waveforms, screenshots, and waveform data files to internal storage, external USB storage devices (e.g. U disk), or network storage. Saved setups and reference waveforms can be recalled as needed. For details of network storage, refer to the section “Network Storage”.
Page 273
SDS2000X HD Series Digital Oscilloscope User Manual CSV Data Saves the waveform data in ".csv" format. After selecting this type, you can touch the ParaSwitch to determine whether to include the scope configuration parameter information (horizontal timebase, vertical scale, etc.) or not. Touch the Save All Channel to save all displayed waveforms on the screen or select the Source to save.
SDS2000X HD Series Digital Oscilloscope User Manual Root directory Paste Save Rename Save as Delete Recall Move Save and Recall Instances Save the screenshot to the path “Udisk0\sds2khd\” in the format of BMP First, Insert the U disk. Secondly, set the parameters of the Save operation: ...
Page 276
In step 4, if Save is selected instead of Save As , the system saves the file as the default file name SDS2xx4XHD_BMP_n.bmp, where n is an integer incrementing from 1. The default save path is \SIGLENT\. Note: Touch Utility >...
Page 277
SDS2000X HD Series Digital Oscilloscope User Manual Note: A saved picture can be browsed on the oscilloscope by opening it in the File Manager. Recall the setup file "SDS2504XHD_Setup_2.xml" stored in path “Udisk0\Atom\” First, insert the U disk with the file "SDS2504XHD_Setup_2.xml" stored folder “\Atom\”. Second, set the parameters of the Recall operation: ...
Page 278
SDS2000X HD Series Digital Oscilloscope User Manual Third, select the “\Udisk0\Atom\” directory in the file manager, and then select the setup file "SDS2504XHD_Setup_2.xml" Fourth, touch the Recall icon and wait for the oscilloscope to finish recalling the setup. 2 7 6 i n t .
SDS2000X HD Series Digital Oscilloscope User Manual Utility System Information Operate Utility > Menu > System Info to check the system status. System information includes the contents shown below. System Setting 32.2.1 Language The SDS2000X HD supports multiple languages, including Simplified Chinese, Traditional Chinese, English, French, Japanese, German, Spanish, Russian, Italian, Portuguese, and so on.
SDS2000X HD Series Digital Oscilloscope User Manual 32.2.2 Screen Saver The screen saver will be activated if the oscilloscope has not been operated for a period. At this time the backlight of the display is cut off to save power consumption. Operate Utility >...
SDS2000X HD Series Digital Oscilloscope User Manual Touch the text box of Hour, Minute, Second, Day, Month, and Year to edit these fields. Touch the Modify Date/Time button to perform the change. Touch the text box of Time Zone to select the time zone.
Page 282
SDS2000X HD Series Digital Oscilloscope User Manual Vertical Ref Fixed Offset: When the vertical scale is changed, the vertical offset remains fixed. As the vertical scale is changed, the waveform expands/contracts around the main X-axis of the display Fixed Position: When the vertical scale is changed, the vertical offset remains fixed to the grid position on the display.
Page 283
SDS2000X HD Series Digital Oscilloscope User Manual Fixed position, timebase is changed to 5 ns/div, the grid number of delay (2div) remains fixed, while the horizontal delay changes to -10 ns Fixed delay, timebase is changed to 5 ns/div, the horizontal delay value remains fixed, while the grid number of delay changes to 4div i n t .
SDS2000X HD Series Digital Oscilloscope User Manual Save Button Push the Save button on the front panel to perform the specified save action. The default is to save a screenshot as a picture (.bmp/.png/.jpg). Touch Utility > Save Button to set the file type and path.
SDS2000X HD Series Digital Oscilloscope User Manual I/O Setting 32.5.1 This oscilloscope complies with LXI version 1.5, defined in LXI device specification 2016, which can be used to quickly build automated test systems. Touch LXI to recall the LXI status box as below: Touch LAN Reset to restore the LAN defaults to your oscilloscope.
D. When is not checked, set the DNS IP address manually E. When accessing more than two SIGLENT instruments through the web browser, it is necessary to set a different VNC port number for each instrument. The range is from 5900 to 5999.
SDS2000X HD Series Digital Oscilloscope User Manual A. Select drive. The default is I: B. Input the network storage directory in the text box. Check “Remember” to remember it C. Check“Anonymous”to access the network directory in guest mode (Username: Guest, no password). This needs the server to permit anonymous access. D.
SDS2000X HD Series Digital Oscilloscope User Manual Install Options The SDS2000X HD provides a few options to enhance its functionality. Contact your local SIGLENT sales representative or SIGLENT technical support to get the corresponding option key. Perform the following steps to install the option: Utility >...
SDS2000X HD Series Digital Oscilloscope User Manual Maintenance 32.7.1 Upgrade The firmware is upgradeable through external USB memory devices/U disk. Make sure the U disk contains the correct upgrade file (*.ads) and is connected to the oscilloscope before performing the upgrade.
SDS2000X HD Series Digital Oscilloscope User Manual Click Upgrade in the upgrade dialog box to start the upgrade. The oscilloscope first copies the upgrade file (*.ads) to the local memory and parses it. If the parse succeeds, it will show the following dialog.
Page 291
SDS2000X HD Series Digital Oscilloscope User Manual Press the "Single" button to switch to red and green display mode. Observe whether there is any color distortion, bad pixels, or scratches on the screen. Press the "Run/Stop" button to exit the screen test mode.
Page 292
SDS2000X HD Series Digital Oscilloscope User Manual Knob test: Turn each knob clockwise, counter-clockwise, and press down. Observe whether the value on the corresponding knob (default 0) on the screen increases or decreases in real-time and whether the knob lights up when it is pressed. Button test: Press each button and check whether the corresponding button icon on the screen lights up in real-time.
SDS2000X HD Series Digital Oscilloscope User Manual 32.7.3 Self-Calibration The self-calibration program can quickly calibrate the oscilloscope to reach the best working state and the most precise measurement. It is recommended to perform a self-calibration if the change of ambient temperature is more than 5 ℃. Note: Make sure the oscilloscope has been warmed up or operated for more than 30 minutes before the self-calibration.
SDS2000X HD Series Digital Oscilloscope User Manual 32.7.4 Developer Options This function is used for internal development by SIGLENT. 2 9 2 i n t . s i g l e n t . c o m...
SDS2000X HD Series Digital Oscilloscope User Manual 33 Remote Control The SDS2000X HD provides a LAN port and a USB Device port which can be used for remote control in multiple ways. Web Browser A built-in web server provides an approach to interact with the oscilloscope by a web browser. It doesn’t require any additional software to be installed on the computer.
SDS2000X HD Series Digital Oscilloscope User Manual A. Display and control area of the instrument. The display in this area is a copy of the instrument display. Using the mouse to operate in the area is equivalent to directly operating the touch screen display of the instrument. B.
Check whether the power switch is on. Check whether the fuse is burned out. If the fuse needs to be changed, please contact SIGLENT timely and return the instrument to the factory for replacement by the maintenance personnel authorized by SIGLENT.
Page 298
Make sure that the U disk system format is FAT32. Restart the instrument and then insert the USB storage device to check it. If the USB storage device still cannot be used normally, please contact SIGLENT. 2 9 6 i n t . s i g l e n t . c o m...
Page 299
DC power supplies, electronic loads and other general purpose test instrumentation. Since its first oscilloscope was launched in 2005, SIGLENT has become the fastest growing manufacturer of digital oscilloscopes. We firmly believe that today SIGLENT is the best value in electronic test & measurement. Headquarters: SIGLENT Technologies Co., Ltd...