Resistive material moisture and temperature measuring instrument
Table of Contents

Advertisement

Quick Links

H60.0.14.6C-17
ab Version 2.2
Temperature Measuring Instrument
Please carefully read these
instructions before use!
Please consider the safety instructions!
Please keep for future reference!
GHM Messtechnik GmbH - Standort Greisinger
Hans-Sachs-Str. 26 - 93128 Regenstauf GERMANY
 +49 9402 / 9383-0  +49 9402 / 9383-33 
Resistive Material Moisture And
GMH 3831
Operating Manual
WEEE-Reg.-Nr. DE 93889386
info@greisinger.de

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the GMH 3831 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for GREISINGER GMH 3831

  • Page 1  Please consider the safety instructions!  WEEE-Reg.-Nr. DE 93889386 Please keep for future reference! GHM Messtechnik GmbH - Standort Greisinger Hans-Sachs-Str. 26 - 93128 Regenstauf GERMANY  +49 9402 / 9383-0  +49 9402 / 9383-33  info@greisinger.de...
  • Page 2: Table Of Contents

    H60.0.14.6C-17 Operating Manual GMH 3831 Page 2 of 20 _____________________________________________________ _____________________________________________________________________________ Index GENERAL NOTE ..............................3 SAFETY ..................................3 ............................... 3 NTENDED USE ..........................3 AFETY SIGNS AND SYMBOLS ............................3 AFETY INSTRUCTIONS PRODUCT DESCRIPTION ............................4 ..............................4 COPE OF SUPPLY ...........................
  • Page 3: General Note

    H60.0.14.6C-17 Operating Manual GMH 3831 Page 3 of 20 _____________________________________________________ _____________________________________________________________________________ 1 General note Read through this document attentively and make yourself familiar to the operation of the device before you use it. Keep this document in a ready-to-hand way in order to be able to look up in the case of doubt.
  • Page 4: Product Description

    H60.0.14.6C-17 Operating Manual GMH 3831 Page 4 of 20 _____________________________________________________ _____________________________________________________________________________ This device must not be used at potentially explosive areas! The usage of this device at potentially explosive areas increases danger of deflagration, explosion or fire due to sparking.
  • Page 5: Connections

    H60.0.14.6C-17 Operating Manual GMH 3831 Page 5 of 20 _____________________________________________________ _____________________________________________________________________________ 3.4 Connections 1: Output: Operation as interface: Connect to optically isolated interface adapter (accessory: GRS 3100 or USB 3100 N) Operation as analogue output: Connection via suitable cable. Attention: The output mode has to be configured (p.r.t chapter 7)
  • Page 6: Device Configuration

    H60.0.14.6C-17 Operating Manual GMH 3831 Page 6 of 20 _____________________________________________________ _____________________________________________________________________________ 4 Device configuration For configuration of the device press "Menu"-key (key 4) for 2 seconds, the first menu will be shown. Choose between the individual values that can be set by pressing the "Menu"-key (key 4) again.
  • Page 7: Some Basics Of Precision Material Moisture Measuring

    H60.0.14.6C-17 Operating Manual GMH 3831 Page 7 of 20 _____________________________________________________ _____________________________________________________________________________ 5 Some basics of precision material moisture measuring 5.1 Measuring method The electrical resistance depends on the material moisture in many cases. Therefore the device measures the (possibly extremely high) values of resistance and converts them to the displayed value by means of integrated characteristic curves.
  • Page 8: Measuring In Wood: Measuring With Two Measuring-Needles

    H60.0.14.6C-17 Operating Manual GMH 3831 Page 8 of 20 _____________________________________________________ _____________________________________________________________________________ 5.6 Measuring in wood: Measuring with two measuring-needles Normally wood is measured with measuring-needles. Used electrodes: impact-electrode GSE91 or GSG91, reciprocating piston electrode GHE91. For measuring wood, punch in the measuring-needles across to the wood-grain, having a good contact between the needles and the wood (measuring along wood-grain deviates minimal) Select correct wood-sort (refer to Appendix A).
  • Page 9: Measuring

    H60.0.14.6C-17 Operating Manual GMH 3831 Page 9 of 20 _____________________________________________________ _____________________________________________________________________________ 5.7.3 Measuring Split the log: Use an suitable axe for splitting. Fast working splitting machines can produce heat which may falsify the measuring. Preparation of core temperature measuring of log: Drive in the impact electrode, put temperature probe into the generated hole.
  • Page 10: Soft' Materials (Polystyrene Or Similar): Measuring With Measuring-Rods Or -Pins (Gms 300/91)

    H60.0.14.6C-17 Operating Manual GMH 3831 Page 10 of 20 _____________________________________________________ _____________________________________________________________________________ 5.8.2 ‘Soft‘ Materials (polystyrene or similar): Measuring with Measuring-rods or -pins (GMS 300/91) Useable electrodes: impact electrode GSE91 or GSG91, reciprocating piston electrode GHE91. Procedure as described in chapter measuring in wood.
  • Page 11: Output

    H60.0.14.6C-17 Operating Manual GMH 3831 Page 11 of 20 _____________________________________________________ _____________________________________________________________________________ 7 Output The output can be used as serial interface or as analogue output (0-1V). If none of both is needed, we suggest to switch the output off, battery life then is extended.
  • Page 12: Fault And System Messages

    H60.0.14.6C-17 Operating Manual GMH 3831 Page 12 of 20 _____________________________________________________ _____________________________________________________________________________ 9 Fault and system messages Display Meaning Remedy Blinking curve display: Displayed value is out of Limited measuring precision! specified range (Wood: 8..40%u) The display value is only usable as indication,...
  • Page 13: Specification

    H60.0.14.6C-17 Operating Manual GMH 3831 Page 13 of 20 _____________________________________________________ _____________________________________________________________________________ 13 Specification Measuring Channel 1 Channel 2 Principle Resistive material-moisture-measuring Temperature-measuring thermocouple type K matching DIN EN 13183-2: 2002 or internal temperature-measuring or according VDI 4206 sheet 4 Char. curve...
  • Page 14: Appendix A: Sorts Of Wood

    H60.0.14.6C-17 Operating Manual GMH 3831 Page 14 of 20 _____________________________________________________ _____________________________________________________________________________ Directives / standards: The instruments confirm to following European Directives: 2014/30/EU EMC Directive 2011/65/EU RoHS Applied harmonized standards: EN 61326-1 : 2013 emissions level: class B emi immunity according to table 3 and A.1 Additional fault: <1 %...
  • Page 15 H60.0.14.6C-17 Operating Manual GMH 3831 Page 15 of 20 _____________________________________________________ _____________________________________________________________________________ Beech, New Zeeland Red Carabeen, Yellow h.99 Sloanea woollsii 6..85% h.44 Nothofagus fusca 5..100% (sapwood untreated) Cathormion Cathormion, New Guinea h.100 4..68% Beech, Silky h.45 umbellatum Citronella moorei 8..85% Cedar , Amercan h.102...
  • Page 16 H60.0.14.6C-17 Operating Manual GMH 3831 Page 16 of 20 _____________________________________________________ _____________________________________________________________________________ Chisocheton Guarea, black h.68 Guarea cedrata 7..100% Kiso h.218 6..65% schumannii Guarea, white h.69 Guarea cedrata 9..85% Lacewood, Yellow h.219 Polyalthia oblongifolia 5..87% Guariuba Clarisia racemosa h.150 8..70% Anthocephalus Laran h.223...
  • Page 17 H60.0.14.6C-17 Operating Manual GMH 3831 Page 17 of 20 _____________________________________________________ _____________________________________________________________________________ Meranti, Nemesu from 1999 h.274 Pine / Pine, Stone h.345 Shorea pauciflora 4..100% Pinus pinea 6..100% Meranti, Seraya from 1999 h.275 Pine, Aleppo h.324 Shura curtisii 5..78% Pinus halepensis 8..98%...
  • Page 18 H60.0.14.6C-17 Operating Manual GMH 3831 Page 18 of 20 _____________________________________________________ _____________________________________________________________________________ Dacrydium Tatajuba h.30 Bagassa guianesis 7..50% Rimu (truewood untreated) h.368 8..50% cupresinum Taun Maleisien h.195 Pometia pinnata 0..100% Robinia Robinia pseudoacacia h.369 2..92% Taun New Guinea h.196 Pometia pinnata 6..100%...
  • Page 19: Appendix B: Additional Materials

    H60.0.14.6C-17 Operating Manual GMH 3831 Page 19 of 20 _____________________________________________________ _____________________________________________________________________________ 15 Appendix B: Additional materials Select material you want to measure, enter number on the device, e.g. concrete b25 = b. 6 15.1 Measuring of building materials Material Number Range Concrete Concrete 200kg/m³...
  • Page 20 H60.0.14.6C-17 Operating Manual GMH 3831 Page 20 of 20 _____________________________________________________ _____________________________________________________________________________...