Operating The Ss7 Sc-Hst - Hach DOC026.52.00769 User Manual

Turbidimeter
Table of Contents

Advertisement

Operation

5.7 Operating the SS7 sc-HST

38
If condensation forms in the enclosure, increase the air
pressure (and flow) by increasing the air pressure setting of the
pressure regulator for the flow multiplier.
Make sure the bubble trap is working. Bubbles on the surface
of the liquid will cause incorrect readings.
If deposits accumulate inside the unit, wash the inside with
warm water spray.
The vent cover at the top of the turbidimeter body
page
14, item 5) can be removed for cleaning if necessary.
Loosen the light source before removing the cover. Check the
alignment of the light after reinstalling the cover using the new
alignment plate included in the kit
Make sure the cover sits flat on top of the slant tube
when installed.
Calibrate the instrument using the calibration cup and formazin
as described in
section 5.5.2 on page
Note: Do not operate the instrument without the cover. Do not operate the
instrument if the flow multiplier is not working.
(Figure 6 on
(section 6.4.1 on page
31.
40).

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ss7 scSurface scatter 7 sc

Table of Contents