calibration mass on the pan. Press Yes to accept the weight or No to select an alternate weight.
[--C--] will be displayed while the reading is stored. The display will show [done] if the calibration
was successful. The scale returns to the previous application mode and is ready for use.
1
Note:
Pound masses are used when calibrating in the lb unit.
Span:
Linearity:
Geographic
Adjustment:
End Calibration:
*Bold always represents factory default value.
Span [SpaN]
Initiates a span calibration procedure (zero and span)
Lin [LIN]
Initiates a linearity calibration procedure (zero, mid-point and span).
GEO [GEO]
Geographical Adjustment Factor (GEO) is used to adjust the calibration based on the current
location. Settings from 0 to 31 are available with 12 being the default.
Refer to table 4-2 to determine the GEO factor that corresponds to your location.
End Cal [End]
Advance to the next menu or return to the top of the current menu.

4.3 Setup Menu

Enter this menu to set scale parameters.
Reset:
Power on unit:
Auto Tare:
Zero Range:
Beeper Signal:
Beeper Key:
Accumulation:
End Setup:
Required Span Calibration Mass (sold separately)
Max
Mass
3000g
3kg / 6lb
6000g
6kg / 15lb
Perform
Perform
Set 0...12...31*
Exit menu
no, yes
auto, kg, g, lb, oz, lb:oz
off, on, on-acc
2%, 10%
off, accept, under, over, under-over
off, on
off, auto, manual
Exit menu
TABLE 4-1
1
Max
Mass
15000g
30000g
1
15kg / 30lb
30kg / 60lb