Setup Menu - OHAUS RC41M15-M Instruction Manual

Table of Contents

Advertisement

Ranger™ 4000 Count Series
Zero2:
Perform
Span1:
Perform
Span2:
Perform
Linearity1:
Perform
Linearity2:
Perform
Geographic
Adjustment:
Set 0...12...31*
End Calibration: Exit menu
*Bold always represents factory default value.
Zero2 [ZErO]
nd
Initiates a 2
scale zero calibration.
Span1 [SpaN]
Initiates a span calibration procedure (zero and span). A span calibration is important when
initially setting up the scale.
Span2 [SpaN2]
nd
Initiates a 2
scale span calibration procedure.
Lin1 [LIN]
Initiates a linearity calibration procedure (zero, mid-point and span).
Lin2 [LIN2]
nd
Initiates a 2
scale linearity calibration procedure.
GEO [GEO]
Geographical Adjustment Factor (GEO) is used to adjust the calibration based on the current
location. Settings from 0 to 31 are available with 12 being the default. Refer to table 4-2 to
determine the GEO factor that corresponds to your location.
End Cal [End]
Advance to the next menu or return to the top of the current menu.

4.3 Setup Menu

Enter this menu to set scale parameters.
Reset:
Dual Scale:
Cap2:
Grad2:
Power on unit:
Zero Range:
Auto Opt:
Auto Tare:
Beeper Opt:
Beeper Signal:
Beeper Key:
Accumulation:
End Setup:
no, yes
off, on
1...9999
0.00005...0.5
auto, kg, g, lb, oz, lb:oz
2%, 10%
off, on
on, off, on-acc
off, on
off, accept, under, over, under-over
off, on
off, auto, manual
Exit menu
EN 13