Triggering On External Waveforms - Fluke 196 User Manual

Scopemeter
Hide thumbs Also See for 196:
Table of Contents

Advertisement

Fluke 192/196/199
Users Manual

Triggering on External Waveforms

Use external triggering when you want to display
waveforms on inputs A and B while triggering on a third
signal. You can choose external triggering with automatic
triggering or with edge triggering.
1
Supply a signal to the red and black 4-mm
banana jack inputs. See Figure 29.
In this example you continue from the Trigger on Edges
example. To choose the external signal as trigger source,
continue as follows:
2
Display the
key labels.
3
Select Ext (external) edge trigger.
Observe that the key labels at the bottom of the screen
have been adapted to allow selection of two different
external trigger levels: 0.12 V and 1.2 V:
50
(On Edges)
TRIGGER
Figure 29. External Triggering
4
Select 1.2V under the Ext LEVEL
label.
From this point the trigger level is fixed and is compatible
with logic signals.