Triggering On External Waveforms (Models 190-Xx2) - Fluke ScopeMeter 190-502 User Manual

Manual
Hide thumbs Also See for ScopeMeter 190-502:
Table of Contents

Advertisement

Fluke 190 series II
Users Manual
Triggering on External Waveforms
(models 190-xx2)
Use external triggering when you want to display
waveforms on inputs A and B while triggering on a third
signal. You can choose external triggering with automatic
triggering or with edge triggering.
1
Supply a signal to the red and black 4-mm
banana jack inputs.
In this example you continue from the Trigger on Edges
example. To choose the external signal as trigger source,
continue as follows:
2
Display the
TRIGGER
key labels.
3
Select Ext (external) edge trigger.
F1
64
(On Edges)
TRIGGER
www.
Shop for Fluke products online at:
Observe that the key labels at the bottom of the screen
have been adapted to allow selection of two different
external trigger levels: 0.12 V and 1.2 V:
4
F3
From this point the trigger level is fixed and is compatible
with logic signals.
MyFlukeStore
1.877.766.5412
.com
Select 1.2V under the Ext LEVEL
label.