Record Length
Scope Record Mode.......... 27500 points on each input
Scope Normal Mode............ 1000 points on each input
Scope Glitch Capture Mode... 500 points on each input
Glitch Detection
5 µ s to 120 s/div ....... displays glitches as fast as 50 ns
Waveform Display ...............A, B, A+B, A-B, A*B, A vs B
Normal, Average (2,4,8,64 x), Persistence
Time Base Accuracy ...................................... ±100 ppm
Trigger and Delay
Trigger Modes..................................... Automatic, Edge,
Trigger Delay ............................... up to +1000 divisions
Pre Trigger View ...........................one full screen length
Max. Delay................................................... 10 seconds
Automatic Connect-and-View Trigger
Source ........................................................... A, B, EXT
Slope ................................................. Positive, Negative
Edge Trigger
Screen Update .......... Free Run, On Trigger, Single Shot
Source ........................................................... A, B, EXT
Slope ................................................. Positive, Negative
External, Video, Pulse Width
Trigger Level Control Range......................... ± 4 divisions
Trigger Sensitivity A and B
DC to 5 MHz at >5 mV/div ........................0.5 divisions
DC to 5 MHz at 5 mV/div .............................. 1 division
200 MHz (FLUKE 199) ................................. 1 division
250 MHz (FLUKE 199) ............................... 2 divisions
100 MHz (FLUKE 196) ................................. 1 division
150 MHz (FLUKE 196) ............................... 2 divisions
60 MHz (FLUKE 192) ................................... 1 division
100 MHz (FLUKE 192) ............................... 2 divisions
Isolated External Trigger
Bandwidth ...........................................................10 kHz
Modes.................................................. Automatic, Edge
Trigger Levels (DC to 10 kHz) ................. 120 mV, 1.2 V
Video Trigger
Standards ........................... PAL, PAL+, NTSC, SECAM
Modes.....................Lines, Line Select, Field 1 or Field 2
Source ........................................................................ A
Polarity............................................... Positive, Negative
Sensitivity..................................... 0.7 division sync level
Specifications
9
Dual Input Oscilloscope
89