1 Sensor membrane specifications
2 General information
on page 3
3 What you have received
Section 1 Sensor membrane specifications
Specifications are subject to change without notice.
Specification
Recommended applications
Material
Thickness [μm]
Calibration gas
Dissolved measurement range
1
Accuracy
Expected current in air @ 1 bar 25°C
[μA]
Temperature compensation range
Temperature measuring range
2
Response time
Recommended minimum liquid flow
3
rate
[mL/min]
Recommended minimum linear flow
3
rate
[cm/sec]
Section 2 General information
In no event will the manufacturer be liable for damages resulting from any improper use of product or
failure to comply with the instructions in the manual. The manufacturer reserves the right to make
changes in this manual and the products it describes at any time, without notice or obligation.
Revised editions are found on the manufacturer's website.
2.1 Safety information
The manufacturer is not responsible for any damages due to misapplication or misuse of this product
including, without limitation, direct, incidental and consequential damages, and disclaims such
damages to the full extent permitted under applicable law. The user is solely responsible to identify
critical application risks and install appropriate mechanisms to protect processes during a possible
equipment malfunction.
1
At room temperature (25 °C).
2
Response time at 25°C for a 90% signal change
3
Liquid flow through an ORBISPHERE 32001 flow chamber, with protection cap and no grille
4
These flow rates take into account the decomposition of ozone in the tubing between the line
and the flow chamber (theoretical flow rates in the absence of decomposition would be 10 times
less)
on page 3
on page 5
Membrane 2956A
Trace measurement
0 ppb to 50 ppm
The greater of ±1% of reading
(± 5% for sensors calibrated in air)
4 Installation
on page 5
5 Maintenance
on page 8
6 Troubleshooting
on page 11
PFA
25
Span gas or air
or ± 0.4 ppb,
or ±1 Pa
25.3
– 5 to 45° C
– 5 to 100° C
25 sec.
4
350
30
Membrane 29552A
High concentration (> 1 mg/l)
PTFE
50
0 ppb to 200 ppm
The greater of ±1% of reading
(± 5% for sensors calibrated in air)
or ± 20 ppb,
or ± 4 Pa
6.5
6 min.
4
100
10

English 3