Download Print this page

Power Meter Specifications - Kingfisher KI 7340C Series Quick Start Manual

Premium two way optical loss test set with optical return loss (orl)

Advertisement

KI 7340C Series -
This is the industry fastest and easiest bi-directional loss test set. From start
of test to acceptance report takes one mouse click and 4 seconds per
wavelength.
The real-time loss and ORL display on both instruments means that cable
certification and rectification use the same procedure, which simplifies
training and operating procedures.
Autotest is available on both Test and Meter ports and is compatible with all
other Autotest instruments. When a pair of identical models are used, the
HOLD functions provide handy communication method between two users,
so they can synchronize while working through successive fibers.
High availability is the result of zero warm up, >190 hour battery life, patented
interchangeable optical connectors for both ports, 3 year calibration cycle
and superior reliability.
The instrument is also a standalone traceable power meter, multi- light source
and Optical Return Loss Tester.

POWER METER SPECIFICATIONS

Damage
Response 
level
(Nm)
(dBm)
Calibration  (nm)
InGaAs detector
600 ~ 1700
+15
780, 820, 850, 980
1270, 1290, 1300, 1310,
1330, 1350, 1370, 1390,
1410, 1430, 1450, 1470,
1490, 1510, 1530, 1550,
1570, 1590, 1610, 1625,
H5 (InGaAs) detector
800 ~ 1700
+25
820, 850, 980
1270, 1290, 1300, 1310,
1330, 1350, 1370, 1390,
1410, 1430, 1450, 1470,
1490, 1510, 1530, 1550,
1570, 1590, 1610, 1625, 1650
H3B (InGaAs) detector
800 ~ 1700
+30
4
820, 850, 980
1270, 1290, 1300, 1310,
1330, 1350, 1370, 1390,
1410, 1430, 1450, 1470,
1490, 1510, 1530, 1550,
1570, 1590, 1610, 1625, 1650
Ge detector
600 ~ 1650
+25
780, 820, 850, 980
1270, 1290, 1300, 1310,
1330, 1350, 1370, 1390,
1410, 1430, 1450, 1470,
1490, 1510, 1530, 1550,
1570, 1590, 1610, 1625, 1650
Si detector
350 ~ 1100
+5
635, 650, 660, 780, 850,
Note 1: Midrange linearity excludes top 5 dB and bottom 10 dB of range.
Note 2: Calibration condition: non coherent light, -35±5 dBm, 23±1°C, ±1 nm, 10±3 nm FWHM, PC ceramic connector, 100 µm fiber.
Note 3: Includes contributions of: varying optical connector types, calibration uncertainty, linearity over temperature & range, and fiber core diameter up to 200 µm.
Note 4: H3B can sustain the damage level for 2 minutes.
Note 5: At calibration wavelengths in bold type.
PREMIUM TWO WAY OPTICAL LOSS TEST SET WITH OPTICAL RETURN LOSS (ORL)
Power range
(dBm)
+5 ~ -60
+5 ~ -70
1650
+15 ~ -50
+15 ~ -60
+27 ~ -40
+27 ~ -50
+15 ~ -60
+15 ~ -60
+0 ~ -70
980
The ORL Zero function compensates for residual reflections, and provides
extended measurement range with improved linearity.
The ORL User Calibration Mode compensates for stray losses in a test set-up,
which improves overall accuracy.
Multimode LED sources feature standards compliant beam geometry &
modal distribution across the fiber core, results in greatly improved
measurement accuracy.
The new InGaAs detectors have wider wavelength response range from 600
~ 1700 nm. It provides good response for all common wavelengths.
Flexible KITS™ PC software is a real-time measurement, Pass/Fail assessment
and reporting solution. Easily customized for any language and reporting
format, it also supports memory download, data logging, label printing,
legacy instruments and enterprise level data management.
Tone &
Autotest
Midrange
Calibration
Min
linearity
1
Accuracy
2
(dBm)
(d)
(%)
-45
0.02
1 %
-50
(0.06 dB)
-35
0.02
1 %
-40
(0.06 dB)
-25
0.02
1 %
-30
(0.06 dB)
-45
0.04
1 %
-50
(0.06 dB)
-50
0.02
1 %
(0.06 dB)
typical
Polarization
Total
Sensitivity
Uncertainty
3
 Sensitivity ± 30
(dB)
(dB)
nm
< 0.005
0.3
< 0.005
0.3
< 0.005
0.35
< 0.005
0.5
< 0.005
0.3
typical
max
kingfisherfiber.com
5
(dB)
0.03
0.03
0.03
0.03
0.03
typical

Advertisement