Section 1 Product overview The Intellical MTC301 series ORP probes are digital, combination electrodes that measure the oxidation reduction potential (ORP/Redox) of wastewater, drinking water and general water samples. The probes are refillable and have a built-in temperature sensor and a ceramic pin reference junction.
207 mV at 25 °C Hydrogen Electrode Electrolyte filling solution 3 M KCl saturated with AgCl Storage solution Hach pH electrode storage solution or 3 M KCl Cable connection M12 digital output and connector Dimensions Diameter: 12 mm (0.47 in.) Length: 175 mm (6.9 in.) total;...
3.4 Product hazards C A U T I O N Chemical exposure hazard. Obey laboratory safety procedures and wear all of the personal protective equipment appropriate to the chemicals that are handled. Refer to the current safety data sheets (MSDS/SDS) for safety protocols. C A U T I O N Chemical exposure hazard.
Page 8
Figure 2 Remove the protective tape 6 English...
Figure 3 Probe position during use 1 Open filling hole 3 Level of calibration solution or sample 2 Level of electrolyte filling solution Section 5 Calibration The procedure that follows is applicable to meters that can connect to Intellical ORP (MTC) probes. Refer to the applicable meter documentation for meter operation and probe-specific settings.
5.2 Calibration procedure 1. Go to the 2. Pour the fresh 3. Open the filling 4. Rinse the probe calibrate menu. standard solution hole. with deionized water. Select the probe, if into a beaker. Dry the probe with a applicable. The lint-free cloth.
• Air bubbles below the sensor can cause a slow response or error in the measurement. Make sure to remove air bubbles before and during measurements. 6.2 Sample measurement procedure 1. Open the filling 2. If the samples 3. Rinse the probe 4.
7.1 Verification procedure 1. Go to the 2. Pour the fresh 3. Open the filling 4. Rinse the probe verification menu. standard solution hole. with deionized water. The display shows into a beaker. Dry the probe with a the standard solution lint-free cloth.
3. Rinse the probe with deionized water. Blot dry with a lint-free cloth. 4. To remove inorganic mineral deposits, soak the probe sensor and platinum disc in a solution of 0.1 M hydrochloric acid or 0.1 M nitric acid for a maximum of 15 minutes. 5.
N O T I C E Probes can become permanently damaged if kept in a storage solution that is not specified by the manufacturer. Use only the specified storage solution (Hach pH electrode storage solution or 3 M KCl). Do not store the probe in deionized water or in samples of low ionic strength. For short-term storage, the probe can stay in the sample for a maximum of 2 hours if the sample pH is not high.
Note: Product and Article numbers may vary for some selling regions. Contact the appropriate distributor or refer to the company website for contact information. Description Quantity Item no. Electrode filling solution, 3 M KCl solution saturated with 28 mL 2841700 AgCl Hach pH electrode storage solution 500 mL 2756549 Larger quantities are available. English 13...
Page 16
Section 10 Consumables (continued) Description Quantity Item no. 3 M KCl solution 50 mL 2756559 Electrode cleaning solution for regular maintenance 500 mL 2965249 Electrode cleaning solution for minerals/inorganic 500 mL 2975149 contamination Electrode cleaning solution for fats, oils and grease 500 mL 2964449 contamination...