C O M P O N E N T T E S T I N G , R E P A I R & R E P L A C E M E N T
Component Testing, Repair & Replacement

Proper Repair Procedures

The DeVilbiss 5 Liter Oxygen Concentrator is designed for ease of service. To aid service personnel, a Service Kit (part #444-501) is
available which contains the necessary gauges, tools, and testing instruments to properly service the oxygen concentrator. On parts
that are sold separately, the part number is indicated in parenthesis.
The following parts are included in the Service Kit.
Service Kit
(1) 6-8" Slotted bit
(1) #1 Phillips bit
(1) #2 Phillips bit
(1) 7/16" Socket 1/4" Drive
(1) Crescent wrench
(1) Channel lock pliers
(1) 8" Duckbill pliers
(1) 5/32" Allen bit
(1) 9/64" Allen bit
(1) 7/64" Allen bit
(1) T10 bit
(2) 0-30 PSI 0-30 in. Hg gauge and hose (part #PVO2D-601)
(1) Tool box
(2) Test fittings (part #303DZ-637)
(1) Torx screwdriver w/bits (part #MC44D-712)
(1) AC/DC test light (part #PVO2D-603)
(1) 1/4" Ratchet wrench
(1) 3 mm Hexbit
(1) T-15 Torx "L" wrench
(1) 10 mm Socket 1/4" Drive
(1) 1/4" Drive extension
(1) Plastic storage case
(1) Plastic error indicator tool (part #303DZ-635)
In addition to the Service Kit, an oxygen analyzer (part #O2ANA) is needed to periodically check oxygen concentration levels. A
volt meter will be needed for more accurate voltage testing.
NOTE – Be sure to read all of the steps involved before beginning any of the procedures in this manual.
NOTE – After repairing or replacing a component, run the unit for 20 minutes and check the oxygen concentration.
CAUTION – When servicing the DeVilbiss 5 Liter Oxygen Concentrator, be absolutely certain that the correct tools
are used and that the parts are free of oil and grease or any material not compatible with oxygen. Teflon
recommended and must be applied to the male threads omitting the first thread to eliminate the possibility of tape
particles entering the oxygen system. LOX-8™ sealant may be used in place of Teflon tape.