Suggested Patient Setup

1.
Position the unit near an electrical outlet in the room
where the patient spends most of his or her time.
NOTE – Do not connect to an electrical outlet controlled by
a wall switch.The outlet should be independent of other appli-
ances.
2.
Locate the unit a minimum of five feet (1.5 meters) from
fireplaces, radiators, heaters, and hot-air registers.
3.
Position the unit at least 6 inches (15 centimeters) from
walls, draperies, or any other objects that might prevent
the proper flow of air in and out of the oxygen
concentrator.
WARNING—Oxygen promotes rapid burning. Do
not smoke when using this unit or when near a person
receiving oxygen therapy. Do not operate the oxygen
concentrator within a minimum of five feet (1.5 meters)
from hot, sparking, or burning objects or naked flames.
Do not use in rooms heated by paraffin or portable gas
heaters.
WARNING—Electric Shock Hazard. Only qualified
Sunrise Medical homecare providers may remove the
cabinet.
4.
Attach the appropriate oxygen accessories (oxygen tubing
or humidifier) to the oxygen outlet.
NOTE – A maximum of 50 feet (15 meters) of tubing (plus 3
feet (1meter) cannula) is allowed between the concentrator and
the patient.
Oxygen Tubing Only Connection (Figure 11):
a)
Thread the cannula fitting (part #CN100 or CN200)
onto the oxygen outlet port.
b)
Attach the cannula oxygen tubing (part #OST07,
OST15, OST25, or OST50).
Outlet
connector
cannula
fitting
Oxygen
tubing
Figure 11
O P E R A T I O N / I N S T A L L A T I O N
Oxygen Tubing With Humidification Connection (Figure 12):
If the physician has prescribed an oxygen humidifier as part of
the patient's therapy, follow these steps:
Bubble humidifier
NOTE—For optimum performance, the DeVilbiss 5 Liter
Oxygen Concentrator has a preset nominal output pressure of
8.5 PSI (60 kPa). Use only "bubble-type" humidifiers. Do not use
"jet-type" humidifiers.
NOTE—Condensation from the humidifier may occur in
longer lengths of tubing or if the tubing is laying on a cold floor.
This can be reduced by using a removable humidifier stand
(part #MC44DM-509).
To use the stand:
5.
6.
12
a)
Fill the humidifier bottle (part #HUM16) with distilled
water. Do not overfill. (If using a prefill, go to letter c.)
b)
Thread the wing nut located on the top of the
humidifier bottle to the oxygen outlet so that it is
suspended. Make sure it is securely tightened.
c)
Attach the oxygen tubing (part #OST07, OST15,
OST25, or OST50), not to exceed 50 feet (15
meters), directly to the humidifier bottle outlet fitting.
Oxygen
outlet port
Oxygen
tubing
Figure 12
Attach a straight humidifier adapter fitting (part #444-506)
to the bottle by turning the wing nut on the humidifier
until it is tight on the fitting.
Secure the bottle in the strap.
Attach one end of the oxygen tubing to the oxygen outlet
on the unit and the other end to the plastic adapter fitting
on the humidifier. Locate the humidifier near the patient.
Attach the nasal cannula (part #CANOO), catheter, or face
mask to the oxygen tubing (per the manufacturer's
directions).
Follow the Operating Instructions.