C O M P O N E N T T E S T I N G , R E P A I R & R E P L A C E M E N T
The majority of all the servicing and repairs can be done without removing the front bib. However, to gain access to the components
behind the bib, it may be loosened or removed.
To loosen the bib (Figure 19):
1.
Remove the two screws (located directly above the hour meter) that hold the bib to the unit's internal structure. This will allow
access to the components behind the bib.
Cabinet screws
Figure 18
Bib handle
Screw
Figure 19
26
Rear cover
Cabinet screw