b.
Lowers its dispense probe into the rinse cup.
c.
Drains the diluent reservoir.
d.
Dries the aspiration line
When the SlideMaker completes this processing, READY appears on the SlideMaker
status line.
6.
Press
EXIT
7.
Press
RUN MODE
Draining the SlideMaker Vacuum Accumulator
Procedure
1.
Ensure:
SlideMaker is ready to make smears.
r
LH 700 Series analyzer is ready to process samples.
r
2.
Press
EXIT
3.
Press
ROUTINE FUNCTIONS
4.
Press
ROUTINE FLUIDICS
5.
Press
DRAINING VACUUM ACCUMULATOR
VACUUM ACCUMULATOR while the SlideMaker drains the vacuum accumulator into
the LH 700 Series analyzer waste container and rinses the vacuum accumulator.
When the SlideMaker completes this processing, READY appears on the SlideMaker
status line.
6.
Press
EXIT
7.
Press
RUN MODE
Priming/Backwashing SlideMaker
Procedure
1.
Ensure:
SlideMaker is ready to make smears.
r
LH 700 Series analyzer is ready to process samples.
r
2.
Press
EXIT
3.
Press
ROUTINE FUNCTIONS
4.
Press
ROUTINE FLUIDICS
5.
Press
PRIME SYSTEM
The SlideMaker status line displays BACKWASHING while the SlideMaker backwashes
the sample lines and rinses the rinse trough.
When the SlideMaker completes this processing, READY appears on the SlideMaker
status line.
6.
Press
EXIT
7.
Press
RUN MODE
PN 4277299DE
until you reach the MAIN MENU screen.
to return to the RUN MODE screen.
until you reach the MAIN MENU screen.
.
.
until you reach the MAIN MENU screen.
to return to the RUN MODE screen.
until you reach the MAIN MENU screen.
.
.
.
until you reach the MAIN MENU screen.
to return to the RUN MODE screen.
REAGENTS AND REAGENT RESERVOIRS
. The SlideMaker status line displays RINSING
7
TROUBLESHOOTING