EN - User's manual
CA 10101E
pH-meter

Advertisement

Table of Contents
loading

Summary of Contents for Chauvin Arnoux CA 10101E

  • Page 1 EN - User’s manual CA 10101E pH-meter...
  • Page 2 The product is declared recyclable following an analysis of the life cycle in accordance with standard ISO 14040. Chauvin Arnoux has adopted an Eco-Design approach in order to design this appliance. Analysis of the complete life- cycle has enabled us to control and optimize the effects of the product on the environment. In particular this appliance exceeds regulation requirements with respect to recycling and reuse.
  • Page 3: Table Of Contents

    1.2. Accessories ..................................4 1.3. Replacement parts ................................. 4 1.4. Inserting the batteries ..............................5 2. PRESENTATION OF THE INSTRUMENT ..........................6 2.1. CA 10101E ..................................6 2.2. Terminal strip .................................. 6 2.3. Functions of the instrument ............................7 2.4. Keypad ................................... 7 2.5.
  • Page 4: First Use

    1. FIRST USE 1.1. DELIVERY CONDITION The CA 10101E pH-meter is delivered in a cardboard with: ■ one protective sheath fitted on the instrument, ■ four AA or LR6 alkaline batteries, ■ one USB to micro USB cable, ■ one mains-USB adapter, ■...
  • Page 5: Inserting The Batteries

    1.4. INSERTING THE BATTERIES ■ Turn the instrument over. ■ Press the locking tab and lift off the battery compartment cover. ■ Remove the rubber plug. ■ Insert the 4 batteries provided, with the polarities as shown. ■ Put the rubber plug back in place. ■...
  • Page 6: Presentation Of The Instrument

    2. PRESENTATION OF THE INSTRUMENT 2.1. CA 10101E Terminal strip. CA 10101E pH Meter Backlit LCD display unit. Keypad with 6 keys. pH/mV °C/°F Protecting sheath. On/Off key. Type B micro USB connector protected by a cap. 2.2. TERMINAL STRIP One BNC socket for the measuring probe.
  • Page 7: Functions Of The Instrument

    2.3. FUNCTIONS OF THE INSTRUMENT The CA 10101E is a pH-meter designed for use in the laboratory. It can be used to measure pH, temperature, and oxidation reduction potential (ORP), depending on which electrode is connected. It also enables the pH or ORP measurement value to be provided in the form of a voltage on the analogue output.
  • Page 8: Display

    2.5. DISPLAY MEM FULL Display of the measurement. OFFSET Display of the unit. CAL SET Display of the temperature. Display of the time. Indicates the battery voltage level. When the symbol is empty, the batteries must be replaced. Indicates that the instrument is connected to a PC via the USB cable or that the power adapter is connected. Indicates that auto-off is disabled and the instrument is in permanent mode.
  • Page 9: Prop

    2.7. PROP To make reading easier, the instrument can be set on its prop.
  • Page 10: Use In Stand-Alone Mode

    3. USE IN STAND-ALONE MODE The instrument can operate in two modes: ■ the stand-alone mode described in this section, ■ the record mode, in which it is controlled by a PC. This mode is described in the next section. In order to ensure proper operation of the instrument, always leave the electrode connected and the cap on the USB con- nector closed.
  • Page 11 ■ Plunge the end of the electrode in the first buffer solution. Take care to completely submerge the glass ball and the salt bridge in the solution. Minimum level of immersion. To measure the temperature, you have two choices: ■ ATC mode: Connect the temperature sensor to the jack ■...
  • Page 12 ■ The instrument displays the pH and temperature meas- ■ Press the CAL key. urements and the time. The instrument briefly displays the current calibration val- ATC = Automatic Temperature Compensation ues (the slope in % and the offset in mV). ...
  • Page 13 If you want to abort the calibration of the electrode, long-press the END, key before the end of the measurement. Otherwise, when the measurement has stabilized the first calibration point is applied. The instrument then informs you that you can calibrate a second point. (N°...
  • Page 14 ■ Wait for the measurement to stabilize. Once again, the instrument proposes terminating the calibration (by pressing the END key) or continuing at a third point (by pressing the CAL key). CAL SET CAL SET ■ If two calibration points are enough, long-press the END key. The slope and offset of the electrode will be calculated and displayed. However, you can continue the calibration on a third and last point by pressing the CAL key again.
  • Page 15 3.1.3. MEASUREMENTS Once the calibration is over, the electrode is ready to make measurements. Between measurements, the electrode must be rinsed and dried. When you have finished using it, put it back in its storage bottle. For each measurement, wait until the measurement has stabilized and the temperature is correctly established. symbol indicates the quality of the electrode just calibrated.
  • Page 16: Oxidation Reduction Potential Measurement (Orp)

    3.2. OXIDATION REDUCTION POTENTIAL MEASUREMENT (ORP) ORP = Oxidation Reduction Potential 3.2.1. CALIBRATION Calibration is done at a single point. ■ Connect the ORP electrode to the measuring instrument’s BNC socket. ■ Remove the storage bottle from the electrode. ■ Rinse the electrode in de-ionized water, then dry it. ■...
  • Page 17 ■ Plunge the end of the electrode into the calibration solution. Take care to completely submerge the salt bridge in the solution. Minimum immersion level. To measure the temperature, you can: ■ Connect the temperature sensor to the jack socket on the ■...
  • Page 18 It then proposes selecting the calibration set. Two sets are available. ORP at 25°C (mV) Calibration set 1 Calibration set 2 If you have your own calibration solutions, you can change these values in the ORP_Set.txt file (see §4.3). ■ Choose the calibration set by long presses on the ▲ and ■...
  • Page 19: Recording The Measurements

    3.2.2. RESTORE THE INITIAL CALIBRATION Press the CAL key, then the MEM key.    If you do not want to restore the initial calibration, choose no before pressing the CAL key. Otherwise, choose YES and press the CAL key. The offset returns to 0.0 mV. 3.2.3.
  • Page 20: Analogue Outputs

    3.4. ANALOGUE OUTPUTS The CA 10101E is equipped with one analogue output (two 4 mm banana socket, one red and one black) which supply continuously a voltage proportional to the pH or ORP value. Connect a voltmeter with an input impedance greater than 1 MΩ between the Vout and GND terminals of the instrument.
  • Page 21 Use a text editor to open the SetAnalogOutput_pH.txt file to see its default contents: SET OUTPUT_0V SET OUTPUT_5V Which corresponds to the output voltage curve (V) according to the pH measured: 10 11 12 You can modify the values, which will then modify the curve. The pH values must be between -2 and 16.
  • Page 22: Errors

    Er.20: Calibration error. The file defining the set of calibration solutions is missing. Download it from our web site: www.chauvin-arnoux.com Click on the “Support” tab, then search on the name of your instrument “CA 10101E”. Copy the file to the memory of the instrument connected to the PC via USB.
  • Page 23: Use In Recording Mode

    4. USE IN RECORDING MODE The instrument can operate in two modes: ■ the stand-alone mode described in the previous section, ■ the record mode, in which it is controlled by a PC. This mode is described below. 4.1. CONNECTION The instrument communicates by a USB link, using the USB to micro USB cord provided.
  • Page 24 In the contents, you will find the pH_Set.txt and ORP_Set.txt files. You can open and modify these files using a text editor: ■ add or remove a calibration set ■ modify a calibration set by adding, removing, or modifying buffer solutions. Do not change the structure of the file.
  • Page 25: Data Logger Transfer Software

    For the ORP_Set.txt file Number of calibration sets. Calibration set number. Number of buffer solutions in the calibration set (only one per set). ORP of the buffer solutions in mV. 4.4. DATA LOGGER TRANSFER SOFTWARE Once the instrument is connected to the PC, open the Data Logger Transfer software. For context-sensitive information about the use of the Data Logger Transfer software, refer to the Help menu.
  • Page 26 4.4.2. DATE AND HOUR In the Instrument menu, the icon lets you set your instrument’s date and time. These cannot be changed while recording or when a recording session has been scheduled. By clicking , you can choose the date and time display formats. 4.4.3.
  • Page 27: Other Software

    In this case, all of the data will be lost. 4.5. OTHER SOFTWARE The CA 10101E can be used with the Regressi software via the Heito P310 instrument’s communication protocol and with the Graph2D software via the Heito MPC350 instrument’s communication protocol.
  • Page 28: Technical Characteristics

    5. TECHNICAL CHARACTERISTICS 5.1. REFERENCE CONDITIONS Quantity of influence Reference values Temperature 23 ± 3°C Relative humidity 45 to 60%RH Battery supply voltage 4 to 6.4V USB supply voltage 5V ± 5% Electric field < 1V/m Magnetic field < 40A/m The intrinsic uncertainty is the error specified for the reference conditions.
  • Page 29: Memory

    5.2.4. ANALOGUE OUTPUT Generation range 0 to 5,000 mV Resolution 1 mV Intrinsic uncertainty ± 0.5% ± 2 mV Input impedance > 1 MΩ Analogue output refreshed every second. Protection up to 30 V and protection against short-circuits. 5.2.5. INFLUENCE OF TEMPERATURE Influence of temperature (from -10°C to 55°C at 50% RH) on the instrument.
  • Page 30: Usb

    5.4. USB Protocol: USB Mass Storage Maximum transmission speed: 12 Mbit/s Type B micro-USB connector B 5.5. POWER SUPPLY The instrument is supplied by three 1.5V LR6 or AA alkaline batteries. It is possible to replace the batteries by rechargeable NiMH batteries of the same size.
  • Page 31: Environmental Conditions

    5.6. ENVIRONMENTAL CONDITIONS Instrument for indoor and outdoor use. Altitude < 2000m, and 10,000m in storage. Pollution degree % RH °C -30 -20 1 = Range of reference. 2 = Range of use. 3 = Storage range (with neither primary nor rechargeable batteries). 5.7.
  • Page 32: Maintenance

    6. MAINTENANCE Except for the batteries, the instrument contains no parts that can be replaced by personnel who have not been specially trained and accredited. Any unauthorized repair or replacement of a part by an “equivalent” may gravely impair safety. 6.1.
  • Page 33: Calibration Log

    See you on our site: www.chauvin-arnoux.com Then go to “Support”, then “Download our software”, then “CA 10101E”. Updating the embedded software may reset the configuration and cause the loss of the recorded data. As a precaution, save the data in memory to a PC before updating the embedded software.
  • Page 34 ■ Connect the instrument to your PC using the USB cord provided. ■ Copy the .bin file to the instrument as if were a USB key. ■ When the copying is done, press the MEM key and the instrument displays LOAD, indicating that the software is being installed. ■...
  • Page 35: Warranty

    7. WARRANTY Except as otherwise stated, our warranty is valid for 24 months starting from the date on which the equipment was sold. The extract from our General Conditions of Sale is available on our website. www.chauvin-arnoux.com/en/general-terms-of-sale The warranty does not apply in the following cases: ■...
  • Page 36 FRANCE INTERNATIONAL Chauvin Arnoux Chauvin Arnoux 12-16 rue Sarah Bernhardt Tél : +33 1 44 85 44 38 92600 Asnières-sur-Seine Fax : +33 1 46 27 95 69 Tél : +33 1 44 85 44 85 Our international contacts Fax : +33 1 46 27 73 89 info@chauvin-arnoux.com...