Diode / Continuity Testing

Press G to toggle between the continuity and diode test functions for the primary
display. (These functions cannot be selected for the secondary display.)
To perform a continuity test:
1. If required, press G to select the continuity test function.
2. Connect the test leads between the Meter and the circuit under test as shown in
Figure 3-10.
The beeper emits a continuous tone if the input is below 20 e.
To perform a diode or transistor junction test:
1. If required, press G to select the diode test function.
2. Connect the test leads between the Meter and the diode or transistor junction as
shown in Figure 3-11.
The forward voltage of the semiconductor junction (or junctions) is measured.
Readings are displayed in the 2 V range at the fast measurement rate. The Meter
displays 0L if the input is above +2 V.
Figure 3-10. Continuity Test
Operating the Meter from the Front Panel
Selecting a Measurement Function
INPUT
V
2W/4W
HI
1000 V
1000V
MAX
600 V
750V
1V
LO
500 V pk
200 mA
MAX
10 A
MAX
mA
3
SENSE
4 W
HI
CAT I
300V
CAT II
LO
10A
eue15.eps
3-15