5240
5100-5800
5500
5785
E E M M S S C C o o m m p p l l i i a a n n c c e e T T a a b b l l e e — — I I n n p p u u t t A A C C p p o o w w e e r r P P o o r r t t
P P h h e e n n o o m m e e n n o o n n
Electrical fast
transients/burst
Surges Line-to-line
Conducted
disturbances induced
by RF fields
Voltage dips

Voltage interruptions

E E M M S S C C o o m m p p l l i i a a n n c c e e T T a a b b l l e e — — S S i i g g n n a a l l i i n n p p u u t t / / o o u u t t p p u u t t p p a a r r t t s s P P o o r r t t
P P h h e e n n o o m m e e n n o o n n
Conducted
disturbances induced
by RF fields
Pulse modulation 217 Hz, 9 V/m
B B a a s s i i c c E E M M C C
I I m m m m u u n n i i t t y y t t e e s s t t l l e e v v e e l l s s
s s t t a a n n d d a a r r d d
H H o o m m e e h h e e a a l l t t h h c c a a r r e e e e n n v v i i r r o o n n m m e e n n t t
EN 61000-4-4:
±2 kV
100 kHz repetition frequency
EN 61000-4-5:
±0,5 kV, ±1 kV
EN 61000-4-6:
3 V, 0.15 MHz - 80 MHz
6 V in ISM bands and amateur radio bands between 0.15 MHz
and 80 MHz
80% AM at 1 kHz
IEC 61000-4-
0 % UT; 0.5 cycle
11:
At 0º, 45º, 90º, 135º, 180º, 225º, 270º and 315º
0 % UT; 1 cycle
and
70 % UT; 25/30 cycles
Single phase: at 0º
IEC 61000-4-
0 % UT; 250/300 cycles
11:
B B a a s s i i c c E E M M C C
I I m m m m u u n n i i t t y y t t e e s s t t l l e e v v e e l l s s
s s t t a a n n d d a a r r d d
H H o o m m e e h h e e a a l l t t h h c c a a r r e e e e n n v v i i r r o o n n m m e e n n t t
IEC 61000-4-6:
3V, 0.15 MHz - 80 MHz
6 V in ISM bands and amateur radio bands between 0.15 MHz
and 80 MHz
80% AM at 1 kHz