TA Instruments DHR Series Getting Started Manual page 26

Small angle light-scattering (sals) accessory
Hide thumbs Also See for DHR Series:
Table of Contents

Advertisement

4
Enter the desired values and click OK.
The averaged image should resemble that shown below. A similar process can be used to obtain a
background image for water.
Figure 16 Averaged Mie scattering image in ImageJ.
Once you have both averaged images, the background is subtracted using the Image Calculator window
(shown below) selected from the Process menu.
Figure 17 Image Calculator window in ImageJ.
5
When the background correction has been made, click the Point selections icon
6
Place the crosswires as close to the center of the beamstop as possible, and click to fix (the position can
be adjusted later, if it is not correct).
DHR/AR Series SALS Accessory Getting Started Guide
on the toolbar.
Page 26

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ar series

Table of Contents