A “Declaration of Conformity” in accordance with the directive has been made and is on file. European Communities should contact the Authorized Representative listed below regarding any Medical Device Directive (MDD) inquiries. Contact Name: Parker Hannifin Manufacturing Sarl Mailing Address: Via privata Archimede 1, 20094 Corsico, Italy Phone: +44 (0) 1271-313131 The Quality System for Porter Instrument is certified to ISO 13485.
Page 2
IMPORTANT: READ MANUAL COMPLETELY BE F O R E O P E R AT I N G T H I S D E V I C E Basic delivery technique is described. Also, this manual contains instructions on periodically required checks to be performed by the user.
Page 3
FLOWMETER FEATURES FIGURE 1 Oxygen Flowmeter Tube Emergency Air Valve indicates the automatically flow of O in L/min ± 5%. provides the patient wit h ambient air if gas flow is interrupted. (Check Monthly) Nitrous Oxide Flowmeter Tube Non-Rebreathing Valve indicates flow of N O in L/min ±...
INTENDED USE: SAFETY FEATURE CHECKS Flowmeter for Analgesia IMPORTANT: Delivery System These are safety features, which you should routinely check to assure proper function. This device is intended for patient use by an attending If any of these safety features are not functioning physician, dentist or licensed professional properly properly, contact your authorized distributor or Porter trained in its use.
Page 5
NON-REBREATHING VALVE CONCENTRATION CONTROL KNOB VALVE – CHECK MONTHLY CHECK MONTHLY With unit turned OFF, disconnect the corrugated rubber Set the Concentration Control Knob to 50%, and Flow tubing from the Rubber Goods and breathe into the Control Knob to 2 to 3 L/min. The ball indicators will corrugated tubing connected to the bag tee.
D I R E C T I O N S F O R U S E 7. When the procedure is nearing completion, NOTE: These directions detail a basic delivery technique. amounts of N O should be decreased. Terminate However, this is not a comprehensive description and not a the flow of N O and deliver 100% O to begin a...
Page 7
NOTE: If your unit does not operate as described CLEANING METHODS in Steps 1 through 10 under “Directions for Use”, We recommend the use of an approved disinfectant please contact your authorized distributor Porter for the healthcare environment for cleaning the outside Instrument.
Page 8
O, Blue Bands on flow tube. 70% Max Concentration N Similar to basic Model MXR design, except, different color-coding for Oxygen – white C3000 bands on flow tube and white buttons. Flowmeter body is painted white and has a failsafe block cover that is also painted white. The overall length of meter body and tube assemblies is reduced in length by approximately 1.23”.