S530 Parametric Test System Reference Manual

Operating environment conditions

To ensure operation within specifications, the S530 must be operated inside of the following
environmental conditions.
Temperature: 23 °C ±5 °C
Operating humidity: 30% to 60% relative humidity, noncondensing, after a two-hour warm up time
Vibration: High ambient vibration levels may require isolation pads or the repositioning of equipment
Air quality: The S530 system is compatible for use in a Class 10 clean room
Audible system noise: Decibel level is 65 dBA in optimal environmental conditions
Airflow: The S530 system is configured for top to bottom airflow
Altitude: Less than 2000 m (6,561 feet) above sea level
Noise interference: To prevent electrical noise from interfering with measurements, the ambient AC
magnetic field must not exceed 2 × 10
Avoid locating the S530 next to plasma etchers, large motors, magnets, RF transmitters,
equipment with flash lamps, and other potential sources of interference
Position equipment to avoid routing signal and power cables near sources of electrical noise

Floor plan

The following floor plan information is for the system cabinet only. Be sure to place the cabinet a
minimum distance of 15 cm (6 in.) up to a maximum distance of 122 cm (48 in.) from the prober.
Refer to the documentation for the prober or other test-fixture equipment to determine its floor space
requirements.
The system cabinet requires a floor space of approximately 1.2 m × 2.1 m (4 ft × 7 ft), plus additional
room for service personnel access from the front and back of the cabinet. The following figures show
a top view of the floor plan and the typical S530 system cabinet weight distribution and center of
gravity.
System cabinet size and weight
cabinet.
See the figure on the following page for floor plan specifications.