Setting Pump Stall Threshold Values - Static Method - Honeywell MultiRAE Series User Manual

Portable multi-gas monitor
Table of Contents

Advertisement

Dynamic pump stall calibration is complete. You can now exit Diagnostic Mode.
Important! When you are done setting thresholds, exit Diagnostic Mode and test the instrument
before actual use.
Setting Pump Stall Threshold Values – Static Method
Use the following values for reference when using the Static method to adjust the pump stall
threshold values:
Low Speed
Vacuum
≤ -2.5in Hg
Flow
>200 cc/min
rate
Idle (I)
100 to 200 counts
(Block-Idle) > 100
Block
counts
Stall
(Idle + Block)/2
Setting
Honeywell MultiRAE Series™
High Speed
≤ -10in Hg
>300 cc/min
150 to 250 counts
(Block-idle) > 100
counts
(Idle + Block)/2
162
User Manual