Honeywell Midas-M User Manual
Hide thumbs Also See for Midas-M:

Advertisement

Quick Links

MIDAS-M MULTIGAS TRANSMITTER
Fixed Single Point Extracted
Multi Gas Transmitter
3016M5001 Rev A
ECO HAA190080

Advertisement

Table of Contents
loading

Summary of Contents for Honeywell Midas-M

  • Page 1 MIDAS-M MULTIGAS TRANSMITTER Fixed Single Point Extracted Multi Gas Transmitter 3016M5001 Rev A ECO HAA190080...
  • Page 3: Table Of Contents

    Table Of Contents CHAPTER 1 — Introduction About Us Product Description Overview User Interface Default Configuration Installation CHAPTER 2 — Mount the Transmitter Device Mount the Relay Module Sample and Exhausting Tubing Calculations In-Line Filters Local Detector Option Electrical Installation Electrical Connection Wiring The Relay Module Mount The Sensor Cartridge...
  • Page 4 Test Mode Calibration Mode Calibration & Bump Test Bump Test Internal Web Server Physical Network Components Internet Settings Running the Web Server CHAPTER 4 — Maintenance System Leak Check CHAPTER 5 — Additional Information EU Directive 2012/19/EU: Waste Electrical and Electronic Equipment (WEEE) Troubleshooting and Fault Diagnosis Specifications...
  • Page 5 Modbus TCP DOS attack Install Web Server Certificate in Google Chrome Install Web Server Certificate in MS Browsers Warranty Sensor Cartridge Warranty Contact Us...
  • Page 7: Chapter 1 - Introduction

    While you build a better world, we’re building smarter safety. For nearly 50 years, Honeywell Analytics has been protecting people who put their lives on the line every day. It takes a special level of drive and dedication to work in hostile environments, and we believe your workers deserve the same kind of commitment from us when it comes to securing their health and safety.
  • Page 8: Overview

    Power over Ethernet (PoE) connection that enables a single Ethernet connection to be made for all power, control and communication requirements. Overview The Midas-M gas detector comprises of 4 parts: the main module, the mounting bracket assembly, the sensor cartridge and the pump module. Midas-M...
  • Page 9: User Interface

    User Interface Detector's User Interface overview. Midas-M User Manual...
  • Page 10: Default Configuration

    NOTE: Oxygen levels are 20.9% v/v in a normal atmosphere, equivalent to 17.3 mA. Use caution when integrating an oxygen Midas-M unit using the 4-20 mA output since fault, inhibit, and no power conditions are below 4 mA and, by default, an oxygen depletion alarm is triggered on a falling alarm (default 19.5% v/v).
  • Page 11: Chapter 2 - Installation

    Otherwise, Midas-M can have flow leakage or malfunction. Mount the Transmitter Device The Midas-M gas detector has an integral mounting bracket assembly that is easily mounted to a suitable vertical surface such as a wall, tool housing, mounting plate on a pole etc.
  • Page 12 NOTE Notice the space required to install the instrument, open the door, push in/pull out modules, and tubing. Mounting procedure Midas-M User Manual...
  • Page 13 1. the door. 2. Un-tight the two screws. 3. Carefully pull the main module forward. 4. Drill two holes 2.2 in (55.9 mm) Use 2 x M4 Screws or equivalent for mounting (head size 7-11 mm). Midas-M User Manual...
  • Page 14 Tighten the screws to secure the module to the mounting bracket assembly. Remove the internal packing card securing the pump. Failure to remove this packing will result in damage to the Midas-M detector. Midas-M User Manual...
  • Page 15: Mount The Relay Module

    5. Push the mounting bracket assembly horizontally and engage the Relay Module connector. a. Align the screw holes and the slot-boundaries, and tight the machine screws (3016D0355) provided in packaging. b. Connect the mounting bracket assembly to the Relay Module. Midas-M User Manual...
  • Page 16 Midas-M User Manual...
  • Page 17: Sample And Exhausting Tubing Calculations

    Tubing ID, mm (in) 3.18 (0.125) Honeywell Analytics recommends thick-wall tubing (1/8” ID) for best speed of response. Due to its lower surface area, thick- wall tubing may require less conditioning than thin-wall tubing. The flow rate is electronically maintained at approximately 600 cc/min and may vary within acceptable tolerances.
  • Page 18: In-Line Filters

    3 months. Refer to the Gases Table for specific gases. Local Detector Option The Midas-M gas detector can also be used to monitor for gas at the location of the detector. To do this, an inline filter is simply connected to the sensor cartridge gas inlet port. The external dust filter part number is 780248 for normal gases and 1830-0055 or 1991-0147 for corrosive gases.
  • Page 19: Electrical Installation

    If Power over Ethernet (PoE) is used to power the device, then 24 VDC power must not also be connected to the device, (or conversely, if 24 VDC is used to power the Midas-M, then electrical power via the Ethernet port must not be applied). Failure to observe this requirement may cause damage to the gas detection system and will not be covered by the standard warranty.
  • Page 20 Instrument grounding is required to ensure stable performance and to limit the effects of radiofrequency interference before installation. The following diagram shows the terminal module layout and terminal identification as well as the jumper locations. Transmitter Terminals Midas-M User Manual...
  • Page 21 - Channel 4 mA output - Channel 4 mA output - Channel 4 Relay 1 Normally Closed Relay 1 Common Relay 1 Normally Open Relay 2 Normally Closed Relay 2 Common Relay 2 Normally Open Relay 3 Normally Closed Midas-M User Manual...
  • Page 22 Relay 2 Relay 3 INSTRUMENT F Instrument fault Instrument Alarm 1 Alarm 2 ONLY only fault Separate fault Maintenance Instrument SEPARATE F relays alarm fault fault Combined fault COMBINED F Alarm 1 Alarm 2 Any fault relay Midas-M User Manual...
  • Page 23: Electrical Connection

    Electrical Connection Midas-M can be powered by either 24 VDC via traditional discrete wiring or by approximately 48 VDC delivered through the Ethernet cable from a PoE source. In either case, the 4-20 mA analog output can be used. The analog output can be configured for fully isolated operation. With 24 VDC power the 4-20 mA output can be configured for sink, source or isolated output operations.
  • Page 24 Generic example Midas-M 4-Wire Isolated output Generic example Midas-M 3-wire Sinking output Midas-M User Manual...
  • Page 25 Generic example Midas-M Isolated 4-20mA output w/PoE power Midas-M User Manual...
  • Page 26: Wiring The Relay Module

    Eight gas alarm relays which indicate level 1 and level 2 alarms of each channel. A common fault relay which indicate maintenance fault or instrument fault. Specifications No separated power supply required Supplied from the Midas-M Power Supply transmitter Power <...
  • Page 27 Default configuration All alarm relays Normally de-energized, energized on alarm Common fault relay Normally energized, de-energized on fault Midas-M User Manual...
  • Page 28: Mount The Sensor Cartridge

    Mount The Sensor Cartridge The Midas-M Sensor Cartridge enables the detection of a wide range of asphyxiant, toxic, flammable, pyrophoric, corrosive, and oxidizer (including Oxygen) gases. The plug & play Sensor Cartridge is located inside the detector’s chassis. NOTE The Midas-M sensor cartridge is supplied separately.
  • Page 29: Chapter 3 - Device & Web Operations

    CHAPTER Device & Web Operations Learn what you can do with your Honeywell Midas®-M Detector, from commissioning to Maintenance Comissioning Comissioning should be followed carefully and only performed by suitably trained personnel. WARNING Before carrying out any work, ensure local and site procedures are followed. Ensure that the associated control panel is inhibited to prevent false alarms.
  • Page 30: Monitoring Mode

    Monitoring Mode After commissioning, the Midas-M detector is set in the Monitoring Mode, which is the starting point for all of the device's operations. The examples in the following table are for a linear 4-20 mA output over a full scale range of 2 ppm.
  • Page 31 Fault relay de- activated Alarm relay 1 de- activated Green Yellow flash of the flash Instrument Alarm relay 2 de- channel or all fault activated channels dependent Yellow on fault type flash Fault relay activated Midas-M User Manual...
  • Page 32: Review Information

    Information you can review from selected option FW Revision Part Number Serial Number Transmitter FW checksum Web version FW Revision Boot loader revision Part Number Cartridge Serial Number Sensor parameter revision of each channel Gas type of each channel Alarms Midas-M User Manual...
  • Page 33 Calibration interval of each channel Calibration due days of each channel Calibration Sensor expired days of each channel Date/Time Date and Time IP obtain method IP address Subnet Mask Network Gateway MAC Address Event Log Event logs Advance Channel ON/OFF Midas-M User Manual...
  • Page 34: Setup Operations

    Alarm 2 type and level of each channel Alarms Deadband level of each channel Alarm latching/non-latching of each channel Alarm delay time of each channel 4-20mA Concentration range of 4-20mA output of each channel. Internal Relay configuration (Instrument Fault Only, Separate Fault Relay, Midas-M User Manual...
  • Page 35 Calibration interval of each channel Date Date format (YYYY MM DD, DD MM YYYY, MM DD YYYY) Date/Time Time IP obtain method (Manual, Auto) IP address Network Subnet Mask Gateway Advance Channel ON/OFF Security ON/OFF Security Password Midas-M User Manual...
  • Page 36: Test Mode

    Operations you can Test from the selected option Bump Gas bump test Alarm 1 of each channel Alarm 2 of each channel Simulation Maintenance Fault Instrument Fault 4-20mA Force 4-20mA output of each channel Inhibit ON/OFF Inhibit Midas-M User Manual...
  • Page 37 Option Operations you can Test from the selected option Inhibit time Midas-M User Manual...
  • Page 38: Calibration Mode

    From the Calibration Mode panel, you can configure any of the following options: Option Operations you can calibrate from the selected option Zero Zero gas calibration of each channel Span Span gas calibration of each channel Flow Pump flow calibration 4-20mA 4-20mA output calibration of each channel Midas-M User Manual...
  • Page 39: Calibration & Bump Test

    ISO 9000 quality controlled procedure. The Midas-M sensor cartridge design is very robust and resistant to long term drift; in fact in a normal operating environment it is possible to extend the calibration interval with Midas-M up to 24 months (subject to local requirements concerning calibrations).
  • Page 40: Bump Test

    This is particularly important when the sensor is operating outside of normal environmental conditions (20°C, 50% rH). The Gas Table indicates the recommended calibration or bump test gas for the Midas-M sensor cartridges: Consult with your Honeywell Analytics Service Representative for approved methods and materials before commencing calibrations and bump tests.
  • Page 41: Internal Web Server

    The Midas-M gas detector unit utilizes an Ethernet port with the TCP/IP protocol as standard. The Midas-M can function as a HTML web page server and these web pages can be viewed on external computer equipment (PC, PocketPC etc.) by using a standard software program, i.e.;...
  • Page 42: Running The Web Server

    For example, if the IP address of the Midas-M was set to 169.254.60.47 (the factory default) then an appropriate IP address for the PC would be 169.254.60.42. IP Address setting in Windows 7 Running the Web Server In a web browser, set the URL to “http://xxx.xxx.xxx.xxx” where the “xxx” fields are replaced with the IP address of the Midas-M.
  • Page 43 The gas information for each channel is displayed. The color bar under each box Channels shows the status of the channel. General Device's general information Summary A print-ready summary of device's information. & Print Reset If alarms are occurring, reset the alarms. Alarm Midas-M User Manual...
  • Page 44 Log Review On this page, you can view information about records in text format and as a detailed graph. From the home page, select General > Event History. Midas-M User Manual...
  • Page 45 The following graph shows detailed information about the selected log. The graphic displays the gas concentration before and after a specific time from when the event occurred. Midas-M User Manual...
  • Page 46 Get the calibration information implemented on a per-channels basis when the device was manufactured. On this page, you can view information from the producer, device, and name of the calibrator. From the home page, select General > Calibration Certificate. Midas-M User Manual...
  • Page 47 Get Fault and Contact information To get fault's list and Contact info, select General > Contact Info/Help. Midas-M User Manual...
  • Page 48 6. Enter the Alarm delay. When an alarm condition is reached, an alarm occurs at a delayed time. 7. Set up Latching in an alarm situation. 8. Set the time for calibrating the progress alarm. 9. Click Accept. Midas-M User Manual...
  • Page 49 On the home page, select Set-up > Relays. You can set the relays as follows: Relay configuration: Set the operation conditions for Relay 1,2,3. Alarm relay: Set whether to generate alarm relay. Fault relay: Set whether to generate a fault relay. Midas-M User Manual...
  • Page 50 On the home page, select Set-up > 4-20 mA. You can set the 4-20MA as follows: Channel: Select the channel to set. Output 4 mA: Set the gas concentration for 4mA. Output 20 mA: Set the gas concentration for 20mA. Midas-M User Manual...
  • Page 51 Latching faults: Set whether to perform latching when faults occur. Maintenance: Set whether to generate a fault for maintenance Temperature: Set whether to generate a fault for Temperature. This option is enabled if the Maintenance setting is allowed too. Midas-M User Manual...
  • Page 52 You can set the network as follows: Obtain IP: You can select whether Manual or Automatic. Ip address, Subnet, and Default gateway are enabled only when the Obtain IP option is set to Manual. Hostname: Hostname setting. Midas-M User Manual...
  • Page 53 Set up the Time and Date Set the time and date. On the home page, select Set-up > Time/Date. You can set the Time and Date in a 24 hour format and synchronize it with the computer. Midas-M User Manual...
  • Page 54 On the home page, select Set-up > Advance. Zero Calibration Perform Zero calibration for a specific channel or all at once. 1. On the home page, select Calibration > Zero Calibration. 2. Select the channel or channels to zero calibrate, and then click Start. Midas-M User Manual...
  • Page 55 3. Wait until the gas reading is stable, and then click Accept. Midas-M User Manual...
  • Page 56 Perform an span calibration for each channel of the device. 1. On the home page, select Calibration > Span Calibration. 2. On the Span calibration page, you can select the gas, the gas type, and the target concentration. 3. Click Start and then follow screen instructions. Midas-M User Manual...
  • Page 57 Flow Calibration Calibrate the amount of gas entering the device. On the home page, select Calibration > Flow Calibration, and then follw screen instructions. You can adjust setpoints,and sync setpoints. Midas-M User Manual...
  • Page 58 Perform a mA calibration for each channel of the device. 1. On the home page, select Calibration > 4-20mA Calibration. 2. Select a channel, and then click Start. 3. Increase or decrease the mA value, and then click Next to Zero calibration (4 mA). Midas-M User Manual...
  • Page 59 Perform a Bump Test Perform a Bump Test on diverse channels. 1. On the home page, select Test > Bump Test. 2. Click Enter Bump Test Mode. 3. Optional Step. You can click Stop Bump Test to cancel the procedure. Midas-M User Manual...
  • Page 60 Simulate function for testing alarm faults. 1. On the home page, select Test > Alarm/Fault Simulation. 2. Select a Channel. 3. Select whether the simulation should be on an Alarm or a Instrument fault. 4. Click Accept. Midas-M User Manual...
  • Page 61 Perform Inhibit Test Enable or disable the inhibit state. 1. On the home page, select Test > Inhibit. 2. From the Inhibit level option, select No inhibit or Inhibit. 3. Select the Inhibit timeout in minutes. 4. Click Accept. Midas-M User Manual...
  • Page 62 Note: The 4-20 MA test may cause unexpected alam activation. Only authorized operators should perform this task. 1. On the home page, select Test > 4-20 mA Test. 2. Select a Channel. 3. Click Accept. 4. Increase or decrease the mA value. 5. Click Close. Midas-M User Manual...
  • Page 63 Midas-M User Manual...
  • Page 65: Chapter 4 - Maintenance

    CHAPTER Maintenance Midas-M is a fully serviceable product designed with modular components that can be readily replaced by trained service personnel so as to minimize the time that the gas detector is not available. External in-line air filters should be replaced every three months or more frequently if the system is sampling in environments that have high levels of particulate matter or very acidic / wet atmospheres.
  • Page 66: System Leak Check

    ” (Flow Fail) – approximate time to fault is 24 seconds. 5. Remove the plug. 6. Allow 15 seconds or so to allow the Midas-M flow to stabilize then clear the fault by pressing and holding ‘X’ 7. Plug the Outlet Exhaust line.
  • Page 67 11. Allow 15 seconds or so to allow the Midas-M flow and concentrations to stabilize then clear the fault by pressing and holding ‘X’ 12. Return the unit to the Monitoring Mode. Midas-M User Manual...
  • Page 69: Chapter 5 - Additional Information

    CHAPTER Additional Information Learn from about strategic information related to the Honeywell Midas®-M Detector. EU Directive 2012/19/EU: Waste Electrical and Electronic Equipment (WEEE) This symbol indicates that the product must not be disposed of as general industrial or domestic waste. This product should be disposed of through suitable WEEE disposal facilities.
  • Page 70: Troubleshooting And Fault Diagnosis

    Troubleshooting and Fault Diagnosis General troubleshooting guide and specific fault code table for the Midas-M gas detector. Fault code descriptions Number Date Data Field Condition Recovery Simulated M User enters simulation Reset simulated fault fault mode Check for background gas concentration,...
  • Page 71 IR Replace IR cell Failure cell Cell memory has no Cell memory response or corrupted Replace cell error data Cell Memory Checksum error. Replace cell Invalid Cell IR Cell IR voltage is out of Replace cell Midas-M User Manual...
  • Page 72 Checksum error. Reboot invalid Transmitter memory Communication failed Reboot comm. Error Transmitter Replace transmitter or app memory Checksum error. update the latest invalid version Transmitter Transmitter 3.3V is out Voltage Replace Transmitter 3.3V error of range Midas-M User Manual...
  • Page 73 Number Date Data Field Condition Recovery Transmitter Transmitter 24V is out of Voltage Replace Transmitter 24V error range Transmitter Transmitter 5V is out of Voltage Replace Transmitter 5V error range Midas-M User Manual...
  • Page 74: Specifications

    LEDs: Power (Green), Alarm 1 (Red), Alarm 2 (Red), Fault (Yellow)" "Alarm1, Alarm2, Fault Relays (3) rated 1.0 A @ 30Vdc or 0.5A @ Relays in 125Vac Max 10uA @ 10mV minimum, configurable as normally open Transmitter or closed, latched or unlatched" Midas-M User Manual...
  • Page 75 2 wire, 14 AWG maximum Digital CAT5 Cable or equivalent: RJ45 connector Gas Concentration Display and Interface "4-digit alphanumeric display with separate units, concentration bar Instrument graph for each channel and other icon driven indicators 4 button interface keypad" Midas-M User Manual...
  • Page 76 1 year Sensor cartridge 2 years Normal condition: (1) No gas alarm, (2) Without tube and pressure/vacuum Full load condition: (1) All 4 channels gas alarm are on, (2) Maximum tubing length and pressure/vacuum on the inlet/exhaust line Midas-M User Manual...
  • Page 77: Ordering

    Ordering This section contains details of how to order complete Midas-M detector and sensor cartridge kits, separate transmitters and sensor cartridges as well as spares and accessories. Transmitter Part Description Number MMT-01 Midas-M 4-20mA Extractive Transmitter MMT-02 Midas-M TCP/IP Modbus Extractive Transmitter...
  • Page 78: Smart Sensor

    Description number MMS-00 Midas-M Dummy Sensor MMS-E2 Midas-M Hydrogen Chloride (HCI) 0 - 8ppm, 2 years MMS-X2 Midas-M Silane (SiH4) 0 -20 ppm, 2 years MMS-G2 Midas-M Hydrogen (H2) 0 - 1000ppm, 2 years MMS-L2 Midas-M Chlorine (CI2) 0 - 2ppm, 2 years...
  • Page 79 MM-FRT Midas-M Front Module MM-B01 Midas-M 4-20mA Base Module MM-B02 Midas-M TCP/IP Base Module MM-M01 Midas-M 4-20mA Main Module MM-M02 Midas-M TCP/IP Main Module MM-CAT Midas-M Cartridge wo Smart Sensors Mounting Accessories Part number Description MIDAS-A-036 MIDAS DIN Rail Mounting Kit 0235-0128 Flexible Conduit 21"...
  • Page 80 Tubing Teflon® FEP 1/4" OD x 1/8" ID (Sample Inlet) (100' continuous length) 100440 Tubing Polypropylene 1/4" OD X 3/16" ID (Exhaust) (Price per Foot) 1991-0137 Tubing Polypropylene 1/4" OD X 3/16" ID (Exhaust) (100' continuous length) Midas-M User Manual...
  • Page 82: Gas Table

    Gas Table Maximum Calibration Bump Test Sensor Flow Recommended Default Default Sensor Range Resolution Part Rate Maximum Gas Name Formula Alarm Alarm Warm-up Output Recommended Conc End of line filter Number (cc/min) Calibration Conc. Sample Line Time(minutes) Equiv. Time Bump Test (ppm) Time Length(m)
  • Page 83 Maximum Calibration Bump Test Sensor Flow Recommended Default Default Sensor Range Resolution Part Rate Maximum Gas Name Formula Alarm Alarm Warm-up Output Recommended Conc End of line filter Number (cc/min) Calibration Conc. Sample Line Time(minutes) Equiv. Time Bump Test (ppm) Time Length(m) (ppm)
  • Page 84: Combination Index

    Combination Index SiH4 HF H (MMS- (MMS- (MMS- (MMS- (MMS- (MMS- (MMS- (MMS- (MMS- (MMS- (MMS- (MMS- SiH4 (MMS- (MMS- (MMS- (MMS- (MMS- (MMS- (MMS- (MMS- (MMS- HF H (MMS- Midas-M User Manual...
  • Page 85: Reflex

    REFLEX sampling interval in order to establish the actual viability of the cell. Within a relatively short time, Midas-M will be able to alert the user via fault codes that the electrochemical cell is likely to be requiring replacement and is possibly unable to correctly detect gas.
  • Page 86: A Modbus/Tcp Interface

    This applies to registers Gas Concentration and Alarm Threshold. The Midas-M fills 122 Modbus/TCP holding registers as listed in Table 1. The Monitoring Status register is a concise summary of Midas-M status. The other registers provide more complete information.
  • Page 87 0 : Ch1 fault active bit 1 : Ch2 fault active bit 2 : Ch3 fault active bit 3 : Ch4 fault active bit 4 : Device fault active bit 5~15 : for future expansion 40005 Top Fault Number of most important active fault Midas-M User Manual...
  • Page 88 Value * 1/10 DECP 40019 to Value in fixed Conversion equation: ppm value = 40020 Concentration point Value * 1/10 DECP 40021 to Value in fixed Conversion equation: ppm value = 40022 Concentration point Value * 1/10 DECP Midas-M User Manual...
  • Page 89 If any fault exists this will take a value in each bit. if any maintenance fault exists, this will take a value zero. The meaning of bit will be described as fault code in more detail Midas-M User Manual...
  • Page 90 : External Relay 7 energized (Ch3 alarm2) bit 12 : External Relay 8 energized (Ch4 alarm1) bit 13 : External Relay 9 energized (Ch4 alarm2) bit 14~15 : for future expansion 40030 Temperature Temperature in Celsius (oC) Midas-M User Manual...
  • Page 91 10 : Cell Failure IR Comm. (F51) bit 11 : Cell EEPROM Read Error (F52) bit 12 : Cell EEPROM Corrupted (F53) bit 13 : Cell Failure IR Volt. (F54) bit 14~15 : : for future expansion Midas-M User Manual...
  • Page 92 10 : Cell Failure IR Comm. (F51) bit 11 : Cell EEPROM Read Error (F52) bit 12 : Cell EEPROM Corrupted (F53) bit 13 : Cell Failure IR Volt. (F54) bit 14~15 : : for future expansion Midas-M User Manual...
  • Page 93 10 : Cell Failure IR Comm. (F51) bit 11 : Cell EEPROM Read Error (F52) bit 12 : Cell EEPROM Corrupted (F53) bit 13 : Cell Failure IR Volt. (F54) bit 14~15 : : for future expansion Midas-M User Manual...
  • Page 94 10 : Cell Failure IR Comm. (F51) bit 11 : Cell EEPROM Read Error (F52) bit 12 : Cell EEPROM Corrupted (F53) bit 13 : Cell Failure IR Volt. (F54) bit 14~15 : : for future expansion Midas-M User Manual...
  • Page 95 Write Command (Write command1 and command2 to execute) 40122 Write 0x015e : Reset Alarm & Fault 0x025e Command 1 : Enter Inhibit (default 30min) 0x035e : Exit Inhibit 40123 Write Fixed value as 0x3626 Command 2 Midas-M User Manual...
  • Page 96 19 AsH3 AsH3 20 F2 21 SiF4 SiF4 22 O2 23 Si2H6 Si2H6 24 SO2 25 CO2 26 SiH2Cl2 SiH2Cl2 27 NO 28 N2O 29 NO2 30 Br2 31 H2S H2S High 32 POCl3 H2S Low 33 3MS Midas-M User Manual...
  • Page 97 M21 - Cell Expires Soon M40 - Temperature Near Limit M41 - Flow Low F01 - LED indicator error F02 - Simulated F Fault F10 - Baseline Fault F20 - Calibration Expired too long F21 - Cell Expired Midas-M User Manual...
  • Page 98 F80 - Transmitter NOR flash error F81 - Transmitter memory invalid F82 - Transmitter memory comm. Error F83 - Transmitter app memory invalid F84 - Transmitter 3.3V error F85 - Transmitter 24V error F86 - Transmitter 5V error Midas-M User Manual...
  • Page 99: Security

    Security Midas-M is shipped from the factory with security disabled. It is recommended to enable the security function for safe use of the device. The security function can only be enabled through the Set up > Security menu of the local UI and requires at least 8 of passwords. If the security function is enabled, the web server can be accessed through HTTPS.
  • Page 100 Midas-M User Manual...
  • Page 101: Install Web Server Certificate In Google Chrome

    2. Click Certificate to download the certificate and follow onscreen instructions. 3. On the Chrome main page, click the three dots (Advanced) button and select Settings > Manage certificates and then follow onscreen instructions as follows to set the certificate. Midas-M User Manual...
  • Page 102 4. Close and re-open the Chrome Web browser, and then access to your device's address https://xxx.xxx.xxx.xxx . A " " message is displayed. The certificate should Connection is secure be reinstalled when the IP address is changed Midas-M User Manual...
  • Page 103: Install Web Server Certificate In Ms Browsers

    3. On your computer locate and run Certificate Manager certmgr.msc On the Certificate Manager window, expand Trusted Root Certification Authorities and right click on Certificates. Select All tasks > Import... Follow the Certificate Import Wizard onscreen instructions. Midas-M User Manual...
  • Page 104 The certificate should be reinstalled when the IP address is changed Midas-M User Manual...
  • Page 105: Warranty

    Honeywell Analytics under a Quality Management System that is certified to ISO 9001. As such, the Midas-M Transmitter (including the pump) is warranted under proper use, to the original end- user purchaser, against any defects in materials or workmanship related failures for a period of 12 months from the date of first turn-on or 18 months from delivery from Honeywell Analytics to the customer, whichever is less.
  • Page 106: Sensor Cartridge Warranty

    Honeywell Analytics. The initial warranty period is not extended by virtue of any works carried out there under. Honeywell Analytics reserves the right to change this policy at any time. Contact Honeywell Analytics for the most current warranty information.
  • Page 107: Contact Us

    Switzerland Tel: +41 (0)44 943 4300 Fax: +41 (0)44 943 4398 gasdetection@honeywell.com Asia Pacific, India. Honeywell Analytics Asia Pacific, Co., Ltd. 7F SangAm IT Tower, 434 Worldcup Buk-ro, Mapo-gu, Seoul 03922, South Korea Tel: +82 (0) 2 6909 0300 Fax: +82 (0) 2 2025 0388 India Tel: +91 124 4752 700 Analytics.ap@honeywell.com...
  • Page 108 Hi-Tech Park Pudong New Area Shanghai 201203, China Tel: 021-80386800 Fax: 021-60246070 gaschina@honeywell.com Taiwan Honeywell Taiwan Ltd 6F-2, No.8, ZiQiang S. Road, Jubei City, 30264 Taiwan Tel: +886-3-5169284 Fax: +886-3-5169339 analytics.tw@honeywell.com Midas-M User Manual...
  • Page 109 © 2019...

Table of Contents