MultiRAE-series instruments use instrument firmware version 1.52/sensor firmware version 1.52.
MultiRAE series device uses instrument FW 1.31/ sensor FW 1.08A
MultiRAE lite with Wi-Fi option uses instrument FW 1.36/ sensor FW 1.36
Note: Fws 1.36 are not compatible with  MultiRAE series device not using Wi-Fi module (and vice
and versa).
IMPORTANT! BUMP TEST THE MONITOR BEFORE EACH DAY'S USE
Prior to each day's use, every gas detection monitor should be bump tested to confirm the
response of all sensors and activation of all alarms by exposing the monitor to a
concentration of target gas that exceeds the low alarm set point. A bump test is also
recommended if the monitor has been subjected to physical impact, liquid immersion, an
Over Limit alarm event, or custody changes, or anytime the monitor's performance is in
doubt.
To ensure greatest accuracy and safety, only bump test and calibrate in a fresh air
environment.
The monitor should be calibrated every time it does not pass a bump test, but no less
frequently than every six months, depending on use and exposure to gas and contamination,
and its operational mode.
• Calibration intervals and bump test procedures may vary due to national legislation.
• Honeywell recommends using calibration gas cylinders containing the gas that is
appropriate to the sensor you are using, and in the correct concentration.
MultiRAE series device uses instrument FW 1.31/ sensor FW 1.08A
MultiRAE lite with Wi-Fi option uses instrument FW 1.36/ sensor FW 1.36