Three-Point Calibration for Enhanced Linearity with Extended-
Range And ppb PID Sensors
For better linearity at higher concentrations when a MultiRAE is equipped with a PID sensor, a
3-point calibration can be performed.
It is recommended to use a 3-point calibration using a high-range PID sensor in a multi-gas
monitor when a specific gas of interest concentration is required. For more information review
TN-114
IMPORTANT!
Three-point calibration is disabled by default, but can be enabled using Safety Suite Device
Configurator (SSDC) software on MultiRAE and MultiRAE Pro instruments with 10.6eV PID
sensors, including high-range ppm and ppb PID sensors. The MultiRAE Lite PID sensor does
not support three-point calibration.
Default calibration gas settings for MultiRAE PID sensors are as follows:
Sensor
MultiRAE Pro
parts-per-billion
(ppb) PID
MultiRAE
extended-range ppm
PID
MultiRAE Lite PID
MultiRAE or
MultiRAE Pro
equipped with 9.8eV
lamp
Honeywell MultiRAE Series™
Specifications
0 to 2,000 ppm
range, 10 ppb
resolution
0 to 5,000 ppm
range, 0.1 ppm
resolution
0 to 1,000 ppm
range, 1 ppm
resolution
0 to 2,000 ppm
range, 1 ppm
resolution
128
Zero
With a charcoal
filter or VOC
zeroing tube
Fresh air or
dry air
Fresh air or
dry air
Fresh air or dry
air
Span 2
(Third
Span
calibration
point, if
enabled)
10 ppm
100 ppm
Isobutylene
Isobutylene
100 ppm
1,000 ppm
Isobutylene
Isobutylene
100 ppm
Not
Isobutylene
supported
100 ppm
Not
Isobutylene 5
supported
ppm Benzene