Electrical Specification
Accuracy is given as +/- (% of reading digits + number of digits) or otherwise specified @ 23
5
o
C and less than 75% R.H.
DC Voltage
RANGE
60.00mV
600.0mV
6.000V, 60.00V, 600.0V
1000V
Input Impedance: 10M, 50 pF nominal
AC Voltage
RANGE
50Hz ~ 500Hz
60.00mV, 600.0mV
6.000V, 60.00V, 600.0V,
1000V
Input Impedance: 10M, 50 pF nominal
Ohm
RANGE
600.0,
6.000K, 60.00K
600.0K
6.000M
60.00M
Open Circuit Voltage: 0.45VDC typical
Audible Continuity Tester
Audible Threshold: Between 10 and 200
Response time: 32ms
Diode Tester
RANGE
1.000V
Test Current: 0.50mA typically
Open Circuit Voltage: < 1.6VDC typically
Accuracy
0.4%+3d
0.3%+3d
0.4%+3d
0.7%+3d
Accuracy
2.0% + 5d
2.2% + 5d
Accuracy
0.5%+6d
0.5%+3d
0.8%+4d
1.0%+5d
1.5%+5d
Accuracy
1.0% + 3d
Capacitance (Model 319 only)
RANGE
6.000F,
60.00F, 600.0F
2000F
Accuracies with film capacitor or better
DC Current
RANGE
Accuracy Burden Voltage
1)
0.7%+3d
600.0A
1)
0.5%+3d
6000A
60.00mA
1)
0.7%+3d
600.0mA
1)
0.5%+3d
6.000A
0.7%+3d
10.00A
2)
0.5%+3d
1)
Ranges for Model 319 only
2)
6A continuous, >6A to 10A for 30 sec. max
with 5 minutes cool down interval
AC Current
RANGE
Accuracy
50H
~ 500H
Z
Z
600.0A
1)
2.2%+5d
6000A
1)
2.0%+5d
60.00mA
1)
2.2%+5d
1)
600.0mA
2.0%+5d
6.000A
2.2%+5d
10.00A
2)
1.2%+5d
1)
Ranges for Model 319 only
2)
6A continuous, >6A to 10A for 30 sec. max
with 5 minutes cool down interval
13
o
C +/-
Accuracy
2.0%+5d
3.5%+5d
4.0%+5d
0.25 mV/uA
2.5 mV/mA
0.03V/A
Burden Voltage
0.25 mV/uA
2.5 mV/mA
0.03V/A

Advertisement