Advertisement

USER'S MANUAL
TBM811, TBM812, TBM829
TBM811XEX, TBM812XEX, TBM525

Advertisement

Table of Contents
loading

Summary of Contents for Brymen TBM811

  • Page 1 USER'S MANUAL TBM811, TBM812, TBM829 TBM811XEX, TBM812XEX, TBM525...
  • Page 2 The meter protection rating, against the users, is double insulation per IEC61010-1 2nd Ed., EN61010-1 2nd Ed., UL61010-1 2nd Ed. and CAN/CSA C22.2 No. 61010.1-0.92 to Category IV 1000 Volts AC & DC. TBM811, TBM812 & TBM829 Terminals (to COM) measurement category: Category IV 1000 Volts AC & DC mAA :...
  • Page 3 WARNING To reduce the risk of fire or electric shock, do not expose this product to rain or moisture. To avoid electrical shock hazard, observe the proper safety precautions when working with voltages above 60 VDC or 30 VAC rms. These voltage levels pose a potential shock hazard to the user.
  • Page 4 3) PRODUCT DESCRIPTION Note: Top of the line model is used as representative for illustration purposes. Please refer to your respective model for function availability. 1) 4 digits 9999 counts dual displays 2) Push-buttons for special functions & features 3) Selector to turn the Power On or Off and Select a function 4) Input Jack for 10A (20A for 30sec) current function...
  • Page 5 Average sensing RMS calibrated RMS (Root-Mean-Square) is the term used to describe the effective or equivalent DC value of an AC signal. Most digital multimeters use average sensing RMS calibrated technique to measure RMS values of AC signals. This technique is to obtain the average value by rectifying and filtering the AC signal.
  • Page 6 magnitude. Note: *Range-Lock and Function-Lock Feature: When a measurement reading is being mode, press the RANGE or SELECT button momentarily 1 time displayed in AutoCheck can lock the range or function it was in. Press the button momentarily repeatedly to step through the ranges or functions.
  • Page 7 in such low impedance mode. Peak initial load current, while probing 1000VAC for example, can be up to 471mA (1000V x 1.414 / 3kΩ), decreasing abruptly to approx. 3.1mA (1000V x 1.414 / 460kΩ) within a fraction of a second. Do not use AutoCheck mode on circuits that could be damaged by such low input impedance.
  • Page 8 selection will be saved as power up default for repeat measurement convenience. TBM525 & TBM829: DCmV, DCmV , DC+ACmV , Logic-Level Hz & Duty% +ACmV +ACmV Press the SELECT button momentarily to select the subject functions in sequence. Last selection will be saved as power up default for repeat measurement convenience. TBM822(XEX) &...
  • Page 9 ACmV , dBm (TBM829 only), Hz functions +ACmV Press the SELECT button momentarily to select the subject functions in sequence. Last selection will be saved as power up default for repeat measurement convenience. only),  Resistance, nS Conductance (TBM525 & TBM829 Continuity functions Press the SELECT button momentarily to select the subject functions in sequence for models 525, 521, 829 and 827.
  • Page 10 Continuity function is convenient for checking wiring connections and operation of switches. A continuous beep tone indicates a complete wire. CAUTION Using resistance and continuity function in a live circuit will produce false results and may damage the instrument. In many cases the suspected component must be disconnected from the circuit to obtain an accurate reading Temperature functions (TBM525 &...
  • Page 11 Normal forward voltage drop (forward biased) for a good silicon diode is between 0.400V to 0.900V. A reading higher than that indicates a leaky diode (defective). A zero reading indicates a shorted diode (defective). An OL indicates an open diode (defective). Reverse the test leads connections (reverse biased) across the diode.
  • Page 12 to-phase voltage which is significantly higher than the phase-to-earth voltage. To avoid exceeding the voltage rating of the protection fuse(s) accidentally, always consider the phase-to-phase voltage as the working voltage for the protection fuse(s). Electric Field EF-Detection (TBM829 only) At any function, press the EF button for one second or more to toggle to EF-Detection feature.
  • Page 13 1ms CREST capture mode (TBM525 & TBM829 only) Press CREST button momentarily to activate CREST (Instantaneous Peak-Hold) mode to capture voltage or current signal duration as short as 1ms. The LCD “C” & “MAX” turn on. The meter beeps when new MAX (maximum) or MIN (minimum) reading is updated. Press the button momentarily to read the MAX , MIN, and MAX-MIN (Vp-p) readings in sequence.
  • Page 14 OL  readings. In other words, the meter will intelligently avoid entering the APO mode when it is under normal measurements.. To wake up the meter from APO, press the SELECT, RANGE, RELATIVE or HOLD button momentarily or turn the rotary switch OFF and then back on.
  • Page 15 ●Press the Yes button momentarily to confirm starting a new logging session directly without erasing the formerly logged one(s). ●Or you can press the Erase button momentarily to erase ALL of the formerly logged session-page(s), and start a new logging session from the very first session-page (P.001) with maximum meter memory.
  • Page 16 3) Recall logged data ●Press the (Up-arrow) and (Down-arrow) buttons simultaneously to enter the Recall mode. The last session-page number shows up for 0.5 second before displaying the last logged data item. “R” & “C” annunicators turn on. ●Press the (Up-arrow) or (Down-arrow) button momentarily to review the logged data one-at-a-time in sequence.
  • Page 17 Battery use: Single 9V battery; NEDA1604G, JIS006P IEC6F22, NEDA1604A, JIS6AM6 or IEC6LF22 Fuses use: TBM811, TBM812 & TBM829 models: Fuse (FS1) for AmA current input: 1A/600Vac, IR 10kA or better, F fuse; Fuse (FS2) for A current input: 10A/600Vac, IR 100kA or better, F fuse TBM811XEX, TBM812XEX &...
  • Page 18 access door and thus the battery compartment up. Replace the battery. Re-fasten the screws. Fuse replacement: Loosen the 4 screws from the case bottom. Lift the end of the case bottom nearest the input jacks until it unsnaps from the case top. Replace the blown fuse(s). Replace the case bottom, and ensure that all the gaskets are properly seated and the two snaps on the case top (near the LCD side) are engaged.
  • Page 19 Safety: Double insulation per IEC61010-1 2nd Ed., EN61010-1 2nd Ed., UL61010-1 2nd Ed. & CAN/CSA C22.2 No. 61010.1-0.92 to Category IV 1000Vac & Vdc. Transient protection: 12kV (1.2/50s surge) TBM811, TBM812 & TBM829 Terminals (to COM) Measurement Category: Category IV 1000Vac & Vdc mAA :...
  • Page 20 or IEC6LF22 Power Consumption: 5 mA typical APO Timing: Idle for 30 minutes APO Consumption: 50A typical Dimension: L208mm X W103mm X H64.5mm with holster Weight: 635 gm with holster Accessories: Test lead pair; battery installed; user’s manual; BKP60 banana plug type-K thermocouple (TBM829 &...
  • Page 21 16, 32, 50, 75, 93, 110, 125, 135, 150, 200, 250, 300, 500, 600, 800, 900, 1000, 1200 AutoCheck (ACV) AutoCheck (DCV) RANGE Accuracy RANGE Accuracy 50Hz ~ 60Hz 9.999V, 99.99V, 999.9V 0.5%+3d 9.999V, 99.99V, 999.9V 1.0%+4d Lo-Z DCV Threshold: Lo-Z ACV Threshold: >...
  • Page 22 30 second max with 5 minutes cool down Response time < 100s interval Capacitance RANGE Accuracy 60.00nF, 600.0nF 0.8% + 3d DC Current 1.0% + 3d 6.000F Burden voltage 2.0% + 3d 60.00F Model Model 3.5% + 5d 600.0F RANGE Accuracy 811XEX 6.000mF 5.0% + 5d...
  • Page 23 Logic Level Frequency ( Hz) & Duty Diode Tester Cycle (D%) RANGE Accuracy @ DCmV Range Accuracy 2.000V 1.0%+1d Function Test Current (Typically): 0.4mA 5.00Hz ~ Open Circuit Voltage: < 3.5 VDC Frequency 0.04%+4d 1.000MHz Duty Cycle 0.00% ~ 100.0% 3d/kHz+2d Temperature Sensitivity: 2.5Vp (Square wave) for 3V &...
  • Page 24 BRYMEN TECHNOLOGY CORPORATION. BRYMEN assumes no risk for damage in transit. BRYMEN will, at its option, repair or replace the defective product free of charge. However, if BRYMEN determines that the failure was caused by misused, altered, neglected, or damaged by accident or abnormal conditions of operation or handling, you will be billed for the repair.