5.3.4 Adjustment under vacuum

Adjustment under vacuum
NOTICE
NOTICE
The adjustment at vacuum for VACUU·VIEW extended
always occurs to the final measured value with 0 mbar.
An adjustment to a reference vacuum value is not possible.
Pump down to an ultimate vacuum as low as possible.
>
VACUU·VIEW (extended) adjusted under vacuum.
1. Connect the gauge
um pump which pumps to a precise end vacuum lower than
< 10
mbar
‑3
or
connect the gauge
which pumps to a precise vacuum lower than < 0,1 mbar.
> Check the accuracy of the ultimate vacuum with a calibrated
reference vacuum gauge.
2. Please wait until the high vacuum pump has reached ultimate
vacuum and until the gauge has completed warm‑up.
3. Call up the menu Settings/Adjustment
menu
play should be 0
5
4. Press Enter
to confirm entry.
Enter
5. Subsequently confirm safety prompt by pressing Enter
5 VACUU·VIEW (extended) adjusted under vacuum.
VACUU·VIEW extended
VACUU·VIEW
Settings/Adjustment. The value on the dis‑
.
0
0
20901330_EN_VVIEW (extd)_V1.11_230223
to a high vacu‑
to a high vacuum pump
Enter.