Adjustment Under Vacuum - vacuubrand DVR 2pro Instructions For Use Manual

Vacuum gauge
Table of Contents

Advertisement

Operation
+
38

6.2.3 Adjustment under vacuum

NOTICE
The adjustment at vacuum always occurs to the final
measured value with 0 mbar.
Pump down to an ultimate vacuum as low as possible.
>
>
Check the accuracy of the ultimate vacuum with a calibrated
>
>
reference vacuum gauge.
Adjustment under vacuum
1. Connect the gauge to a vacuum pump which pumps to a pre‑
cise vacuum, e. g., a rotary vane pump down to < 0,5 mbar.
2. Press and hold the
and then press ON/OFF.
5 Display adjustment mode ‑ only active for approx. 20
seconds.
Display only at vacuum less than < 20 mbar (15 Torr).
3. Press key
UP/DOWN
Changing pointer direction with
ƒ
ƒ
= decrease value /
5 Value for vacuum < 0,5 mbar.
UP/DOWN
key at the switched‑off gauge
repeatedly until 0 mbar is displayed.
MODE key
N°: 999341_EN_DVR 2pro/12.07.2017
= increase value

Advertisement

Table of Contents
loading

This manual is also suitable for:

Dvr 3pro