Connection and operation

Connection and operation

4.1 Connection

4.1
Connection

4.1.1 Installation

4.1.1
Installation
Never install the measuring instrument too close to hot equip‑
ment, like on top of a rotary vane pump, in order to avoid incor‑
rect measurements or damages to the sensor.
The gauge is intended for assembly directly to the apparatus
(application).
> Observe all specifications for installation, connection and
operation according to technical data,
 see chapter 9.1 Technical information
> Also observe rating plate data.
> Compare the permitted limits which are described in this
manual, with your actual application regarding operating
media, pressures, forces, moments, temperatures and volt‑
age.
Installation conditions
The gauge has acclimatized.
ƒ
Ambient conditions are observed and are within the limitation
ƒ
of use.
Limitation of use
Ambient temperature
Altitude, max.
Relative humidity
Protection type/Impact energy
Pollution degree
Avoid condensation or contamination by dust or liquids.
Connection and operation
9.1 Technical
information.
10–40 °C
2000 m
above sea level
30–85 %, non condensing
IP 54/5 J
2
(US)
50–104 °F
6.562 ft
above sea level
17