Controller, Gauge And Vacuu·bus - vacuubrand VACUU-VIEW Instructions For Use Manual

Table of Contents

Advertisement

Meaning
VACUU·BUS
Address assignment
IMPORTANT!
20901330_EN_VVIEW (extd)_V1.11_230223
4.1.4 Controller, gauge and VACUU·BUS
4.1.4
Controller, gauge and VACUU·BUS
VACUU·BUS
is a communication system for peripheral acces‑
®
sories that are connected to VACUUBRAND controller or gauge.
VACUU·BUS
components will be automatically detected by
®
VACUU·SELECT/CVC 3000/DCP 3000. By uniform connectors
and Y adapter the bus‑system can be extended with up to 32
peripherals.
Connenct VACUU·VIEW as VACUU·BUS
> Plug the mains connector of VACUU·BUS
the rear side of VACUU·SELECT, DCP 3000 or CVC 3000.
5 Mains supply via controller or gauge.
Plug connectors of the newest series have a guide tongue
for proper connection. Slide the connection into the guiding
groove on the rear side of the controller or gauge.
Feature
When using the gauge as VACUU·BUS
connected to a CVC 3000, the controller detects the gauge auto‑
matically as vacuum sensor.
First perform address assignment at CVC or DCP, when working
with several VACUU·VIEW gauges of the same type.
For detailed descriptions about address assignment
 see online manual 999151 (CVC 3000)
Address assignment
Vacuubus component
VACUU·VIEW
VACUU·VIEW extended
Reference sensor
Connection and operation
999151 (CVC
3000).
Address-N°
1–4
1–4
1–4
® ®
component (client)
®
into the port on
®
component, e. g., when
®
Abbreviations in
CVC/DCP
VACUU-
SELECT
VSK _
VSK _
VS-C _
VS-C _
VSP _
VSP _
VS-P _
VS-P _
Ref._
Ref._
VS-REF_
VS-REF_
21

Advertisement

Table of Contents
loading

This manual is also suitable for:

Vacuu-view extended

Table of Contents