Connection and operation
IMPORTANT!
IMPORTANT!
18
4.1.2 Vacuum connection
4.1.2
Vacuum connection
> Maximum admissable pressure at vacuum sensor: 1.5 bar/
750 Torr (absolute).
> Pollution and damages, especially at the flange, could affect
the measurement.
> If possible assemble the vacuum gauge vertically; this pre‑
vents condensate accumulation.
Connection via small flange
Required connection material: Clamping ring, centering or cen‑
tering ring for KF DN 16.
1. Remove dust cap.
2. Put the gauge with the centering onto the connection of the
apparatus small flange KF DN16.
3. Fix the vacuum gauge with a clamping ring.
Connection via hose nozzle
Required connection material: Hose nozzle DN 6/10 mm G1/4"
with O ring and compatible hose clip (option).
1. Remove dust cap.
2. Screw the hose nozzle plus the o ring hand‑tight into the
inside thread.
3. Use the hose nozzle to assemble the gauge to a vacuum
hose or directly at the apparatus.
4. Fix the vacuum hose, e. g., with a hose nozzle.
5. Fix the vacuum gauge.
> Use a stable vacuum hose that is suitable for the required
vacuum range. For fine vacuum range, flexible rubber hoses
for example are not useful because of possible gas emis‑
sion.
> Connect hose tubes as short as possible.